manufacturer: Waters
model: Micromass Quattro Micro
The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
Specifications
Dimensions
|
Height
|
572mm (23”)
|
|
Length
|
880mm (34.6”)
|
|
Width
|
390mm (15.4”)
|
Weights
The bench must be able to support the total weight of the system:
Instrument
115 kg (253 lb)
Data system (computer, monitor, printer)
50 kg (110 lb)
Total weight
165 kg (363 lb)
Operating temperature
The ambient temperature should be in the range 15 - 28°C (59 - 82°F) with a short-term variation (1.5 hr) of no more than 2°C (4°F). The ambient temperature in which the rotary pump is to operate must be in the range 15 - 40°C (59 - 104°F).
The maximum overall heat dissipation into the room is 3.0 kW. This figure does not take into account ancillary equipment such as LC systems. Air conditioning may have to be installed or upgraded to accommodate the additional heat load into the room.
Operating humidity
The relative humidity should not exceed 70%. It is recommended that the instrument be situated in a draft free position in an air conditioned laboratory free from excessive amounts of dust.
Shipping and storage temperature
-20 to 60°C (-4 to 140°F)
Electrical specifications
Quattro micro API (including rotary vacuum pump)
|
Voltage & Frequency
|
230V, 50/60 Hz (+8%, -14%)
|
|
Minimum Current
|
13A
|
|
Maximum Current
|
16A
|
|
Power Consumption
|
2.0 kW
|
|
Power Cord Supplied
|
2.5 m (8 ft.) fitted with country plugs as shown below.
|
Canada:
The user must supply a fused supply rated at 15A.
UK:
The cable is fused at 13A.
Rest of World:
The user must supply a fused supply rated between 13-16A .
Japan:
A single phase, 3 wire 200V 5% phase/neutral supply rated between 13-16A must be supplied. Additionally, a transformer must be ordered from Waters to boost the input voltage.
Australia & New Zealand:
The equipment has been designed in compliance with the international safety standard IEC1010-1 (EN61010-1). To be fully effective, the building installation must comply with AS 3000: Electrical Installations for Australia / New Zealand.
Line frequency
50Hz, 47 to 53Hz
60Hz, 57 to 63Hz
Fuse rating
10A, 250V ac
Electrical safety
The Quattro micro complies with the European Directive on Electrical Safety as defined by IEC 1010 and the International standard IEC 1010-1 (EN61010-1).
IEC 1010 Part 1 Ammendment 2. The instrument is categorised as Pollution Category 1 and Installation Category 2.
For Australia and New Zealand: the building installation must comply with AS 3000.
Electrical Safety 195