JavaScript seems to be disabled in your browser. For the best experience on our site, be sure to turn on Javascript in your browser.
24 months legal warranty
60 days Labexchange Fullguarantee
Certificate of operativeness
The Labexchange right of return
Optional installation of equipment for immediate and safe use of equipment
Questions regarding this device
If you buy this item used instead of new, you save as much CO2 as driving 1130 km with a gasoline car.
The legal warranty period for our devices is 24 months from delivery date.
All used instruments of Labexchange are subject to a technical inspection before delivery and will be supplied only, when they function perfectly well. The verification and functionability is documented by a service report of an independent service company. The Labexchange Fullguarantee is 60 days from delivery date. Malfunctions occurring within the warranty period will be corrected by us free of charge. (Valid for countries within the EU and CH)
Our certificate of operativeness ensures a technically perfect condition.
You are on the safe side with our used laboratory and analysis systems! For even more security during the purchase, we are granting the Labexchange right of return.
Return your laboratory or analysis system simply without explanation.
Our special terms and conditions are to be applied:Labexchange right of return(https://www.labexchange.com/fileadmin/user_upload/pdf/downloads/en/Rueckgaberecht_08_18_E.pdf)
On demand, we verify the option of an extension of warranty for your device. If possible, the period of warranty is 6 months for 10 % extra charge and 1 year for 20 % extra charge on the list price. (Valid for DE, AT, CH, FR, NL, BE, LU, GB and IR)
Play it safe! We are pleased to offer you the installation of your laboratory equipment. Our qualified technicians install the device on your desired date in the laboratory, so that you can work quickly and safely with your new laboratory equipment.
The installation of the equipment by Labexchange brings further advantages. You thereby obtain a guarantee extension to 6 months and you have also the right to the Labexchange Repurchase Guarantee. You receive sensible laboratory equipment and analytical systems as well as much of the available laboratory hoods, already incl. installation.
* Required Fields
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
For deliveries to countries outside the EU, an additional fee of 70,00€ for export is charged. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
We do not accept payment by credit card, letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country
Possible payment methods
Comment
DE, AT, CH
Payment by invoice, payment in advance
Payment by invoice is only possible for corporate clients.
NL, BE, LU
Payment by invoice is only possible for corporate clients
Other countries
Payment in advance
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
All articles are used articles, except an article is listed especially as a new device.
Status
Condition
Immediately available
Used
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.
Published
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.
New device
new
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.
Manufacturer Shimadzu
Model: 5000A
The following illustrations and descriptions refer to the instrument model and are drawn from brochures.
They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
This analyzer measures total carbon (TC), inorganic carbon (IC) and total organic carbon (TOC) in water. Its operation is based on the combustion/non-dispersive infrared gas analysis method widely employed for TOC measurement.
Measuring object
TC, IC, TOC (TC-IC), NPOC*
Method
Combustion/non-dispersive infrared gas analysis method
Combustion temperature
680°C
Range
4 ppb to 4,000 ppm (4 ppb to 5,000 ppm for IC)
Analysis time
Normally 2 to 3 minutes for TC or IC (At the maximum,
4.5 minutes for TC and 3.5 minutes for IC)
Repeatability
Less than 2,000 ppm (Less than 2,500 ppm for IC):
Standard deviation is within 1% of the full-scale range.
2,000 to 4,000 ppm (2,500 to 5,000 ppm for IC):
Standard deviation is within 2% of the full-scale range.
Sample introduction method
Introduction with automatic injector
Sample injection volume
Variable between 50 and 2,000 µl (with 2,500 µl microliter syringe)
Variable between 4 and 250 µl. (with 250 µl microliter syringe)
Pretreatment method for IC
With automatic sparging
Carrier gas
High purity air (filled in a cylinder)
Pressure : Approx. 6 kgf/cm2(Max. 7 kgf/cm2)
Flow rate: 150 ml/min. (300 ml/min. including sparging gas)
CPU function
Signal processing
Data processing
Control
Others
Linearization, baseline correction, peak detection/ area calculation
Concentration calculation by one to four-point calibration curve methods, automatic selection of optimum calibration curve from three calibration curves, filing of up to 18 calibration curves, automatic deletion of abnormal values and automatic additional measurement and calculation thereof, SD/CV calculation, and repeated measurement wich automatic change in the range and Sample injection volume
Sample injector movements, TC furnace temperature, and on/off of carrier gas and Spargelgas
Automatic selection of optimum measuring condition, maintenance/check functions, abnormality detection, and timer reservation for re-Start
Display
LCD (white liquid crystal with backlight)
640 x400 dots, 88 characters x 22 lines
Printer
240 dots for graphic, 40 characters/line, Chart width 110 mm, Thermosensitive type
Ambient temperature
5 ~ 35°C
Power requirements
AC 100, 110, 115, 120, 127, 220, 230 or 240 V as ordered,
±10%, 5 A (Ordinary power consumption: About 300 VA),
50- 60Hz
Outside dimensions
About 480Wx520D x 480H (mm) (Excluding protrusions)
Weight
About 46 kg
* NPOC (Non-Purgeable Organic Carbon): TOC measured after sparging