JavaScript lijkt te zijn uitgeschakeld in uw browser. Voor de beste gebruikerservaring, zorg ervoor dat javascript ingeschakeld is voor uw browser.
Usually, laboratory and analysis devices are very exigent system with a very own specifications and requirements. Therefore, the startup of a laboratory device often rises the question how to assemble, adjust and to launch it.
We would be pleased to take over the professional installation of your laboratory devices!Please contact us.
Benefit from the installation by us in more than one way:
Garantieservices
Optionele garantieverlenging tot een jaar
Optionele installatie van apparatuur voor onmiddellijk en betrouwbaar gebruik
Aantrekkelijke huur- en leasingaanbiedingen om uw budget te sparen
Vragen over dit apparaat en gratis terugbellen
Als je dit artikel tweedehands koopt in plaats van nieuw, bespaar je evenveel CO2 als wanneer je 3610 km rijdt met een benzineauto.
Garantie:Alle gebruikte apparatuur wordt vóór levering aan een test en/of revisie onderworpen en wordt alleen dan geleverd als die perfect functioneert. De test en daaruit gebleken goede functionering worden bevestigd in een door een onafhankelijk servicebedrijf opgesteld servicecertificaat. De Labexchange Fullguarantee bedraagt 60 dagen. (Geldig voor landen binnen de EU en CH)Functioneringscertificaat:Zowel de test als het goede functioneren van het apparaat worden, behalve in het servicecerapport, gedocumenteerd in het functioneringscertificaat.24 maanden waarborg:De waarborgtermijn voor onze apparaten bedraagt 24 maanden vanaf leveringsdatum.
Indien gewenst nemen we de mogelijkheid voor verlenging van de garantietermijn van uw apparaat in overweging. Als het mogelijk is, bedraagt de garantietermijn 6 maanden voor een meerprijs van 10% over de catalogusprijs en 1 jaar bij een meerprijs van 20%. (Geldig voor DE, AT, CH, FR, NL, BE, LU, GB en IR)
Speel op zeker! Het doet ons genoegen u de installatie van uw laboratoriumapparatuur aan te kunnen bieden. Onze gekwalificeerde technici installeren het apparaat op de door u gewenste datum in het laboratorium. U kunt zo snel en veilig aan het werk met uw nieuwe laboratoriumapparatuur.
De installatie van apparatuur door Labexchange brengt ook andere voordelen met zich mee. U krijgt op die manier een garantieverlenging tot 3 maanden en heeft bovendien recht op de Labexchange Terugkoopgarantie.
Kwetsbare laboratoriumapparatuur en analysesystemen, evenals veel van de beschikbare laboratoriumkappen, zijn al inclusief installatie.
* Verplichte velden
De gebruikte apparatuur wordt voorafgaand aan levering gecontroleerd door Labexchange Service GmbH. U ontvangt volledig functionerende apparatuur.
De vermelde verzendtijden zijn telkens de kortste voor een artikel. In bepaalde gevallen kunnen de daadwerkelijke verzendtijden daarvan afwijken. De uiteindelijke verzendtijden worden aangegeven in de opdrachtbevestiging.
In de regel bieden we combinatieleveringen aan. Levertijden zijn afhankelijk van het artikel met de langste levertijd. Deelleveringen zijn mogelijk tegen een toeslag.
Koeriersdiensten, transportbedrijven, zelf afhalen, levering door Labexchange wagenpark
De prijzen zijn exclusief verzendkosten. De genoemde verzendkosten zijn de te verwachten kosten. Afwijkingen zijn mogelijk. In het geval geen kosten voor verzending zijn gespecificeerd, vraag die dan afzonderlijk aan.
Voor levering in landen buiten de EU wordt voor export een toeslag van € 70,00 in rekening gebracht. Opmerking: We geven geen preferentieel certificaat/EUR1 af. Bij zelf afhalen/af fabriek (EXW) uit derde landen en de EU wordt 16% btw als borg ingehouden, tot we de ontvangstbevestiging/het leveringscertificaat van de koper hebben ontvangen.
Wij accepteren geen betalingen per creditkaart, Letter of Credit, PayPal etc. Het factuurbedrag is volledig verschuldigd. Er zijn geen betalingskortingen. De goederen blijven tot volledige betaling ons eigendom.
Land
Mogelijke betaalmethoden
Opmerking
Duitsland, Oostenrijk, Zwitserland
Betaling via factuur, vooruitbetaling
Betaling via factuur is mogelijk voor ondernemingsklanten.
Nederland, België en Luxemburg
Andere landen
vooruitbetaling
Onze Algemene Voorwaarden voor Verkoop, Levering en Betaling zijn hierop van toepassing. Deze voorwaarden zijn hier te downloaden.
Tussenverkoop is ons voorbehouden.
Alle artikelen zijn gebruikte artikelen, tenzij bij een artikel uitdrukkelijk wordt vermeld dat het om een nieuw apparaat gaat.
Status
Conditie
Onmiddellijk verkrijgbaar
gebruikt
We hebben het artikel in ons magazijn. Onze technici zullen het artikel voorafgaand aan levering controleren. U ontvangt volledig functionerende artikel.
Het artikel is nog bij de aanbieder. Na uw bestelling wordt het door ons aangekocht, gecontroleerd en aan u verstuurd. Een functioneringscertificaat en een servicerapport zijn bij de levering inbegrepen.
Nieuw
nieuw
Dit is een artikel dat nieuw van de fabriek komt. Hiervoor gelden de garantiebepalingen en wettelijke garantietermijn van de fabrikant.
manufacturer :
model :
annotation : Dokumente engl .
The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
Overview of the LCT Premier
The LCT Premier TM time-of-flight (TOF) mass spectrometer features a ZSpray source with electrospray ionization (ESI) and modular LockSpray TM interface.
The LCT Premier uses ion guide technology for optimum transfer of ions from the source to the orthogonal acceleration time of flight ( oa -TOF) mass analyzer. Ion detection is by dual microchannel plate detector assembly. Data is acquired by a 4.0 GHz time-to-digital converter (TDC) and histogrammed in an embedded PC. Instrument control and data acquisition is by the MassLynx TM software system.
lonization Techniques
Using the Zspray atmospheric pressure ionization (API) source, three techniques are available.
ElectroSpray lonization
Atmospheric Pressure Chemical lonization
Atmospheric Pressure Photoionization
Performance Specifications
Resolution in Positive Ion (V Mode)
Infuse mellitin at a concentration of 5 pmol /µL, in 50:50 acetonitrile :water + 0.2% formic acid, using a flow rate of 5 µL/min. Ensure that the intensity of the peak at 712 Da is <300 counts/s.
The peak at 712 Da should be resolved such that the mass divided by the peak width at half height is greater than 5000 when using a scan time of 1 s and an inter-scan delay of 0.1 s. At least 30 s of data should be summed.
Resolution in Positive Ion (W Mode)
Infuse mellitin at a concentration of 5 pmol /µL in 50:50 acetonitrile :water + 0.2% formic acid, at a flow rate of 5 µL/min. Ensure that the intensity of the peak at 712 Da is <300 counts/s.
The peak at 712 Da should be resolved such that the mass divided by the peak width at half height is greater than 10,000 when using a scan time of 1 s and an inter-scan delay of 0.1 s. At least 30 s of data should be summed.
Resolution in Negative Ion (V Mode)
Infuse raffinose at a concentration of 500 pg/µL in 50:50 acetonitrile :water , at a flow rate of 10 µL/min. Ensure that the intensity of the peak at 503 Da is <300 counts/s.
The peak at 503 Da should be resolved such that the mass divided by the peak width at half height is greater than 5,000 when using a scan time of 1 s and an inter-scan delay of 0.1 s. At least 30 s of data should be summed.
Resolution in Negative Ion (W Mode)
Infuse raffinose at a concentration of 500 pg/µL in 50:50 acetonitrile :water at a flow rate of 10 µL/min. Ensure that the intensity of the peak at 503 Da is < 300 counts/s.
The peak at 503 Da should be resolved such that the mass divided by the peak width at half height is greater than 10,000 when using a scan time of 1 s and an inter-scan delay of 0.1 s. At least 30 s of data should be summed.
Sensitivity in Positive Ion
In V mode, infuse 50 pg/µL of leucine enkephalin in 50:50 acetonitrile :water + 0.1% formic acid at a flow rate of 5 µL/min. Using a scan rate of 1 s, an inter-scan delay of 0.1 s and a mass range of 100 to 1000 Da , check that the resolution at 556 Da is greater than 5000 and the intensity is greater than 440 counts/s.
Switch the instrument to W mode. Check that the resolution is greater than 10,000 and the intensity is greater than 110 counts/s .
Sensitivity in Negative Ion
In V mode, infuse 500 pg/µL. of raffinose in 50:50 acetonitrile :water at a flow rate of 10 µL/min. Using a scan rate of 1 s, an inter-scan delay of 0.05 s and a mass range of 100 to 1000 Da , check the resolution at 503 Da is greater than 5000 and the intensity is greater than 240 counts/s.
Switch the instrument to W mode. Check that the resolution is greater than 10,000 and the intensity is greater than 60 counts/s .
Mass Calibration Accuracy
Infuse 50 pg/µL of sodium formate at a flow rate of 5 µL/min into the analyte probe.In V mode, ensure that the intensity is less than 300 counts/s and that the resolution is greater
than 5000. Run a calibration using the parameters listed in Table E-1, with no smoothing and no background subtraction, and using a centroid top value of 80%.
Table E-1 Mass Calibration Accuracy Parameters
Scan Type Continuum
--------------------------------------------------------------
Scan Range 150-900 Da
Scan Time 5 s
Inter-Scan Delay 0.1 s
Acquisition Time 1 min
Save the calibration and then perform an acquisition for 1 minute using the same parameters. Combine and center the data, using 4 channels and a centroid top value of 80%.Measure at least 6 peaks and calculate the mass error ( ppm ) for each peak. Using the formula below, calculate the RMS error. The RMS error should be < 5 ppm .
Mass Measurement Accuracy
Infuse 50 pg/µL of sodium formate at 5 µL/min into the analyte probe . In W mode, ensure that the intensity is less than 300 counts/s and that the resolution is greater than 10,000. Run a calibration using the parameters listed in Table E-1. Create a calibration with no smoothing and no background subtraction, using a centroid top value of 80%.Save the calibration and then perform the following acquisition.
Infuse 50 pg/µL of leucine enkephalin into the reference probe and 500 pg/µL of raffinose into the ESI probe. Ensure that both probes give an intensity of <200 counts/s and that the resolution is greater than 10,000.
Acquire data in LockSpray mode for 25 minutes, using the following parameters:
Scan Range 100-1000 Da
Scan Time 1 s
Acquisition Time 25 min
Reference Scan Frequency 5
Reference Cone Voltage 100 V
Combine 12 scans at each of the following times: 2, 4, 6, 8, 10, 12, 14, 16, 18, 20 minutes.
Using the Mass Measure feature an each of the spectrums, create a centered spectrum using lockmass 556.2771 Da . Calculate the RMS error using the formula below. The m/z value of raffinose is 527.1588 Da .
The calculated RMS error should be <3 ppm .
Chromatographic Signal to Noise
Note: This test should only be performed if there is already a suitable HPLC system installed.
Infuse 50 pg/µL of leucine enkephalin at a flow rate of 5 µL/min. Tune the ion beam in V mode and check that the resolution is greater than 5000.
Purge the HPLC using 75:25 methanol :water and 5 mM ammonium acetate. Connect the 10-µ.L sample loop, injection port, and waste tube to the Rheodyne injector. Connect a Waters Atlantis dC18 column (2.1 x 30 mm; 3 µM) between the injector and the probe. Condition the column for approximately 20 minutes.
Set up the following sample list acquisition parameters:
Table E-2 Sample List Acquisition Parameters
---------------------------------------------------------------
Scan Time 2 s
Enter-Scan Delay 0.I s
Acquisition Time 30 min
Make five injections of 1 pg/µL reserpine in 75:25 methanol :water and 5 mM ammonitun acetate, at 5 minute intervals.
Using the 609 Da mass chromatogram, calculate the peak to peak, 1 SD signal to noise ratio for each peak. The average signal-to-noise value should be >100.