Merck-Hitachi D-7000

Seriennummer: 020042
Status: Archived Product?The unit is no longer available.
Verzending:
Beschrijving:
Merck-Hitachi HPLC-system. Existing of Interface D-7000. DAD L-7450. Column oven L-7360 with heating- and cooling function. Conductivity detector L-7470. Isocratic pump L-7110. Degasser L-7612, 4 channels. Quaternary pump L-7100. Autosampler L-7200, 100 µL sample loop, six-way valve, tray with 80 positions. Computer. Software preinstalled. 100-240 V. 50/60 Hz. Total power about 700 W. GPIB. CE marked. All units with front operating field and display (except for DAD and interface)
siehe Storeview
Meer informatie
Objektnummer B00020042
Seriennummer 020042
Object Naam Merck-Hitachi D-7000
Status Archiv

Status, leverings- en betalingsvoorwaarden

Apparatuurcontrole

De gebruikte apparatuur wordt voorafgaand aan levering gecontroleerd door Labexchange Service GmbH. U ontvangt volledig functionerende apparatuur.

 

Verzending

De vermelde verzendtijden zijn telkens de kortste voor een artikel. In bepaalde gevallen kunnen de daadwerkelijke verzendtijden daarvan afwijken. De uiteindelijke verzendtijden worden aangegeven in de opdrachtbevestiging.

In de regel bieden we combinatieleveringen aan. Levertijden zijn afhankelijk van het artikel met de langste levertijd. Deelleveringen zijn mogelijk tegen een toeslag.

 

Verzendmethoden

Koeriersdiensten, transportbedrijven, zelf afhalen, levering door Labexchange wagenpark

 

Informatie levering

De prijzen zijn exclusief verzendkosten. De genoemde verzendkosten zijn de te verwachten kosten. Afwijkingen zijn mogelijk. In het geval geen kosten voor verzending zijn gespecificeerd, vraag die dan afzonderlijk aan.

De opgegeven vracht- en verpakkingskosten hebben betrekking op de goedkoopste transportroute en zijn onder voorbehoud van onvoorziene kostenstijgingen. Door onvoorziene gebeurtenissen kunnen vrachttarieven en levertijden op elk moment veranderen en moeten ze worden aangepast aan de huidige situatie. Incoterm coderingen volgens Incoterms 2010: Bij afhalen EXW, CFR voor zendingen over zee, CPT per luchtfracht, andere zendingen DAP. Opmerking: We geven geen preferentieel certificaat/EUR1 af. Bij zelf afhalen/af fabriek (EXW) uit derde landen en de EU wordt 16% btw als borg ingehouden, tot we de ontvangstbevestiging/het leveringscertificaat van de koper hebben ontvangen.

 

Betalingsvoorwaarden

Wij accepteren geen betalingen Letter of Credit, PayPal etc. Het factuurbedrag is volledig verschuldigd. Er zijn geen betalingskortingen. De goederen blijven tot volledige betaling ons eigendom.

Land

Mogelijke betaalmethoden

Opmerking

Duitsland, Oostenrijk, Zwitserland

Betaling via factuur, vooruitbetaling, per creditkaart

Betaling via factuur is mogelijk voor ondernemingsklanten.

Nederland, België en Luxemburg

Betaling via factuur, vooruitbetaling, per creditkaart

Betaling via factuur is mogelijk voor ondernemingsklanten.

Andere landen

vooruitbetaling, per creditkaart

 

 

Onze Algemene Voorwaarden voor Verkoop, Levering en Betaling zijn hierop van toepassing. Deze voorwaarden zijn hier te downloaden.

Tussenverkoop is ons voorbehouden.

 

Beschrijving status:

Alle artikelen zijn gebruikte artikelen, tenzij bij een artikel uitdrukkelijk wordt vermeld dat het om een nieuw apparaat gaat.

Status

Conditie

Opmerking

Onmiddellijk verkrijgbaar

gebruikt Het artikel is al gecontroleerd en is in perfecte staat. Het kan onmiddellijk aan u worden verstuurd.
Op voorraad

gebruikt

We hebben het artikel in ons magazijn. Onze technici zullen het artikel voorafgaand aan levering controleren. U ontvangt volledig functionerende artikel.

Gepubliceerd

gebruikt

Het artikel is nog bij de aanbieder. Na uw bestelling wordt het door ons aangekocht, gecontroleerd en aan u verstuurd. Een functioneringscertificaat en een servicerapport zijn bij de levering inbegrepen.

Nieuw

nieuw

Dit is een artikel dat nieuw van de fabriek komt. Hiervoor gelden de garantiebepalingen en wettelijke garantietermijn van de fabrikant.

Labprocure

gebruikt

Labprocure GmbH is verantwoordelijk voor de inhoud van dit apparaataanbod. Labprocure aanvaardt aansprakelijkheid voor de hier geadverteerde aanbiedingen en voor de daarin opgenomen foto's en aanbiedingsteksten. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

manufacturer :       Merck-Hitachi

model :                   HPLC

annotation :           Dokumente engl.

 

 

The systems are modular designed.

Please finde the modules you are looking for.

 

L-7100 HPLC-Pump

 

Precision and Reliability

The L-7100 has been designed along the lines of the proven serial dual­piston principle, characterised by an extremely constant flow rate, low pulsation and reliable delivery of all commonly used solvents. Moreover, the L-7100 is able to form solvent gradients with high precision and accuracy. Each individual component and the finished instrument are subject to stringent quality control measures. A validation certificate and the manufacturer's test report docu­ments that all functions and specifi­cations have been fulfilled.

 

Versatile in Application

User-friendly programming via key­board and LCD dialogue enable the L-7100 to be used as

n Isocratic HPLC pump

n Quaternary low-pressure gradient pump in conjunction with a gra­dient accessory

n High-pressure gradient pump in conjunction with one or two further L-7100 pumps and as

n Column packing device at con­stant pressure.

The L-7100 pump can store

n 9 time programmes, in which mobile phase composition, flow rate and four events can be pro­grammed and varied over a period of time.

n Via a Link function, programmes for multi-method applications can be linked.

 

A shut-down routine can be pro­grammed to be carried out at the end of an analysis series by

n Delayed pump stop and a

n System flush programme.

 

Due to minimised internal volume, a highly constant flow-rate and excel­lent gradient reproducibility at low flow-rates, the L-7100 pump is

n Optimally suited for narrow-bore

applications.

 

For applications involving sensitive materials requiring operation at low temperatures

n The cold room compatibility of the L-7100 is advantageous.

 

Premature wear caused by salt­containing eluents is prevented by the fitting of a

n Piston cleaning facility, whereby the piston wash procedure can be carried out automatically in combination with the LaChrom autosamplers.

 

GLP Functions

In view of the requirements of qua­lity assurance systems, all modules of the LaChrom system have been fitted with comprehensive facilities for status display, reliability monito­ring and the provision of correspon­ding documentation and traceability.

n All parameters, programmes and gradient curves of the L-7100 pump can be printed out by the La Chrom data systems.

n The pressure signal of the

L-7100 pump is constantly moni­tored as part of the reliability check. Any deviations from the normal pressure profile, caused for example by air bubbles or malfunction of pump valves, are registered during the course of the analysis.

n The Pump pressure profile can be printed out in the analysis report and hence serve in the assessment of the quality of the analytical results.

 

Maintenance is facilitated by the

n Maintenance logbook which auto­matically registers the total volume delivered. This provides an indica­tion as to when the next replace­ment of the pump seal or the next cleaning of the pump valves

should take place.

 

Controller Function

In stand-alone operation, the system control function of the

L-7100 enables external instruments to be controlled by providing

n Four event and other control signals.

 

 

Technical Data L-7100

Delivery system                 Series configured double piston pump with electronic compensation of solvent compressibility and elimination of residual pulsation

Maximum pressure            400 bar

Flow rate                            0.001 –9.999 ml/min

Flow stability                      ≤ 0,2% RSD

Gradient modes                 Quaternary low-pressure gradient, binary or ternary high-pressure gradient

Special functions               Programme link, column pack programme, delayed pump stop,

                                          system flush programme

GLP functions                    Pump logbook, display of pressure profile deviations, key-board lock

Documentation                  Print-out of pump parameters, gradient programmes, gradient curves, pump logbook via LaChrom data systems

Outputs                              4 time-progammable event contacts, analogue pressure signal,

                                          busy/ready , D-line

I nputs                                 Start, stop, pump on, D-line

Materials                             Stainless steel, ruby, sapphire, ärconium oxide-ceramic, PTFE, carbon


L-7400 Programmable UV-Detektor

 

The detectors is equipped with a high-quality reference beam system which compensates for fluctuations in the light beam instability and ageing of the lamp. Newly devel­oped deuterium lamps guarantee longer life and increase the already proven operational reliability of the detectors. The measuring cells are so designed that any refractive index effects brought about by alteration of chromatographic parameters are minimised. This results in the highest degree of detection sensitivity throughout the entire measuring range of the detectors.

 

Textfeld: Technical Data 
Optics	Double beam system
Light source	Deuterium lamp, L-7420: + Tungsten lamp
Wavelength range	190–600 nm, L-7420: 190–900 nm
Accuracy of wavelength setting ± 1 nm
Optical band width	10 nm
Time constants	Selectable, 0.1 s (FAST), 0.5 s (MED), 2.0 s (SLOW) 
Noise	<- 0.8 x 10–5 AU (250 nm)
Drift	<- 3 x 10 –4 AU/h (250 nm)
Detector range	0.001 to 2.5 AUFS
Time-programmable parameters	Wavelength, auto zero, hold for baseline
Working parameters	Time constants, recorder range, offset, L-7420: + polarity
Digital communication	D-Line
Analogue output	Data processor 1 V/AU, recorder 10 mV/ AU
Inputs	Marker in, spectrum start in, start in, lamp off in, autozero in
Diagnostic functions	Lamp energy, lamp light-on time, lamp switching, wavelength
accuracy
Optional accessories	Semi-micro, micro and preparative flow cell
Wavelength Programming and Spectral Recording

Detection at the absorption maxi­mum of an analyte improves not only the detection sensitivity but also the linearity and reproducibility of an analysis. The wavelength-pro­grammable detectors are capable of adjusting automatically to the opti­mal wavelength prior to each peak. This occurs so quickly that each component of a sample is detected at its maximum sensitivity. During wavelength switching, the autozero function can be activated; this pre­vents baseline steps occurring and increases reliability when integrating peaks. In order to enable a particu­lar sample component to be specifi­cally identified, a complete spectrum can be stored and printed out on a recorder or integrator when the chromatogram has been completed.

 

High Operational Reliability and Ease of Use

For GLP work, information on lamp energy, lamp light-on time and an automatic test of wavelength accu­racy can be recalled. The easy-to­read display and simple operation enables the detectors to be opti­mally set. Numerous programming possibilities are available for wave­length, baseline parameters and spectral files. A total of nine time programmes can be entered using a dialogue process. Once entered, programmes remain intact even after the instrument has been swit­ched off.

 

Digital Communication and System Advantages

The new LaChrom detectors have been fitted with an analogue and a digital data output. They can be used either as stand-alone instru­ments or be incorporated into the LaChrom system. When combined with other LaChrom modules, digi­tal data communication ensures maximum data security and inter­ference-free data processing.

The new high-performance data communication network D-Line also facilitates the intelligent and interference-free transfer of digital information between the LaChrom modules as well as the control and display of all parameters, files and spectra via an integrator or PC. Functions such as the "event logbook", the "reliability report" and the automatically compiled service report facilitate GLP work and simplify both maintenance and checking of the instruments.

 

 

Technical Data

Optics                                              Double beam system

Light source                                     Deuterium lamp

UV cell path length                           10 mm

Wavelength range                            190–600 nm

Wavelength accuracy                     ± 1 nm

Spectral band width                         10 nm

Time constants                                Selectable, 0.1 s (FAST), 0.5 s (MED), 2.0 s (SLOW)

Noise                                                ≤ 0.8 x 10 –5 AU (250 nm)

Drift                                                  ≤ 3 x 10 –4 AU/h (250 nm)

Detector range                                 0.001 to 2.5 AUFS

Time-programmable parameters    Wavelength, auto zero, hold for baseline

Working parameters                        Time constants, recorder range, offset

Digital communication                     D-Line

Analogue output                               Data processor 1 V/AU, recorder 10 mV/ AU

Inputs                                               Marker in, spectrum start in, start in,

                                                        lamp off in, autozero in

Diagnostic functions                        Lamp energy, lamp light-on time, lamp switching, wavelength accuracy

Optional accessories                       Semi-micro, micro and preparative flow cell

 