JavaScript lijkt te zijn uitgeschakeld in uw browser. Voor de beste gebruikerservaring, zorg ervoor dat javascript ingeschakeld is voor uw browser.
Usually, laboratory and analysis devices are very exigent system with a very own specifications and requirements. Therefore, the startup of a laboratory device often rises the question how to assemble, adjust and to launch it.
We would be pleased to take over the professional installation of your laboratory devices!Please contact us.
Benefit from the installation by us in more than one way:
Garantieservices
Optionele garantieverlenging tot een jaar
Optionele installatie van apparatuur voor onmiddellijk en betrouwbaar gebruik
Aantrekkelijke huur- en leasingaanbiedingen om uw budget te sparen
Vragen over dit apparaat en gratis terugbellen
Garantie:Alle gebruikte apparatuur wordt vóór levering aan een test en/of revisie onderworpen en wordt alleen dan geleverd als die perfect functioneert. De test en daaruit gebleken goede functionering worden bevestigd in een door een onafhankelijk servicebedrijf opgesteld servicecertificaat. De Labexchange Fullguarantee bedraagt 60 dagen. (Geldig voor landen binnen de EU en CH)Functioneringscertificaat:Zowel de test als het goede functioneren van het apparaat worden, behalve in het servicecerapport, gedocumenteerd in het functioneringscertificaat.24 maanden waarborg:De waarborgtermijn voor onze apparaten bedraagt 24 maanden vanaf leveringsdatum.
Indien gewenst nemen we de mogelijkheid voor verlenging van de garantietermijn van uw apparaat in overweging. Als het mogelijk is, bedraagt de garantietermijn 6 maanden voor een meerprijs van 10% over de catalogusprijs en 1 jaar bij een meerprijs van 20%. (Geldig voor DE, AT, CH, FR, NL, BE, LU, GB en IR)
Speel op zeker! Het doet ons genoegen u de installatie van uw laboratoriumapparatuur aan te kunnen bieden. Onze gekwalificeerde technici installeren het apparaat op de door u gewenste datum in het laboratorium. U kunt zo snel en veilig aan het werk met uw nieuwe laboratoriumapparatuur.
De installatie van apparatuur door Labexchange brengt ook andere voordelen met zich mee. U krijgt op die manier een garantieverlenging tot 3 maanden en heeft bovendien recht op de Labexchange Terugkoopgarantie.
Kwetsbare laboratoriumapparatuur en analysesystemen, evenals veel van de beschikbare laboratoriumkappen, zijn al inclusief installatie.
* Verplichte velden
De gebruikte apparatuur wordt voorafgaand aan levering gecontroleerd door Labexchange Service GmbH. U ontvangt volledig functionerende apparatuur.
De vermelde verzendtijden zijn telkens de kortste voor een artikel. In bepaalde gevallen kunnen de daadwerkelijke verzendtijden daarvan afwijken. De uiteindelijke verzendtijden worden aangegeven in de opdrachtbevestiging.
In de regel bieden we combinatieleveringen aan. Levertijden zijn afhankelijk van het artikel met de langste levertijd. Deelleveringen zijn mogelijk tegen een toeslag.
Koeriersdiensten, transportbedrijven, zelf afhalen, levering door Labexchange wagenpark
De prijzen zijn exclusief verzendkosten. De genoemde verzendkosten zijn de te verwachten kosten. Afwijkingen zijn mogelijk. In het geval geen kosten voor verzending zijn gespecificeerd, vraag die dan afzonderlijk aan.
Voor levering in landen buiten de EU wordt voor export een toeslag van € 70,00 in rekening gebracht. Opmerking: We geven geen preferentieel certificaat/EUR1 af. Bij zelf afhalen/af fabriek (EXW) uit derde landen en de EU wordt 16% btw als borg ingehouden, tot we de ontvangstbevestiging/het leveringscertificaat van de koper hebben ontvangen.
Wij accepteren geen betalingen per creditkaart, Letter of Credit, PayPal etc. Het factuurbedrag is volledig verschuldigd. Er zijn geen betalingskortingen. De goederen blijven tot volledige betaling ons eigendom.
Land
Mogelijke betaalmethoden
Opmerking
Duitsland, Oostenrijk, Zwitserland
Betaling via factuur, vooruitbetaling
Betaling via factuur is mogelijk voor ondernemingsklanten.
Nederland, België en Luxemburg
Andere landen
vooruitbetaling
Onze Algemene Voorwaarden voor Verkoop, Levering en Betaling zijn hierop van toepassing. Deze voorwaarden zijn hier te downloaden.
Tussenverkoop is ons voorbehouden.
Alle artikelen zijn gebruikte artikelen, tenzij bij een artikel uitdrukkelijk wordt vermeld dat het om een nieuw apparaat gaat.
Status
Conditie
Onmiddellijk verkrijgbaar
gebruikt
We hebben het artikel in ons magazijn. Onze technici zullen het artikel voorafgaand aan levering controleren. U ontvangt volledig functionerende artikel.
Het artikel is nog bij de aanbieder. Na uw bestelling wordt het door ons aangekocht, gecontroleerd en aan u verstuurd. Een functioneringscertificaat en een servicerapport zijn bij de levering inbegrepen.
Nieuw
nieuw
Dit is een artikel dat nieuw van de fabriek komt. Hiervoor gelden de garantiebepalingen en wettelijke garantietermijn van de fabrikant.
Modell: IC-System
Kommentar: Dokumente engl. (für Id. 14628)
The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.
The GP40 Gradient Pump is an integral part of a DX 500 chromatography system. It is a microprocessor-based, dual-piston, variable-Speed, gradient delivery system designed to blend and pump mixtures of up to four different mobile phases at precisely controlled flow rates. The pump can deliver the selected mobile Phase composition isocratically, or as a multistep linear or curved gradient, A Digital Signal Processor (DSP) provides high Speed control of pump flow and Pressure.
The GP40 can operate as a stand-alone product or with other Dionex modules as part of a complete chromatography system. It can also be used with non-Dionex modules that meet Interface requirements for software, TTL, or relay control.
The GP40 can be controlled locally, using the front panel keypad and display, or from a remote host computer with a Dionex
DX LANTM interface installed and PeakNet software installed an the host computer. Limited remote control is also available from any device capable of providing compatible TTL signals to control the pump.
The pump's two basic modes of control, Direct control and Method control, enable it to operate with or without reference to time-based events.
Specifications
Main Power Requiremtns 85 to 270 Vac, 47163 Hz, 2.5 amps. The GP40 power supply is main voltalte auto-sensing and requires no manual adjustment.
Fuse Requirements Two 3.15 amp fast-blow 1EC 127 fuses (P/N 954745)
Operating Temperature 10 °C to 50 °C (50 °F to 104 °F)
Operating Humidity 5 to 95% relative humidity (non-condensing) Humidity
Dimensions 33.5 cm high x 22.5 cm wide x 42 cm deep (13.1 inx8.9inx 16.8 in)
6 cm clearance required in back of the module
Weight 19 kg (42 Ibs)
Display Liquid crystal with .adjustable backlighting.
Keypad 26-button pad for entering commands and numerical values for
screen parameters.
Eluent Selection Four different mobile phase components; each can be
proportioned from 0-100%
Pump Dual-piston, variable speed, 100 iaL (standard bore), 25 µL (microbore) stroke, user-selectable constant pressure or constant flow feedback control
Flow Rate Standard bore pump head: 0.04 to 10.0 mL/min, linearly variable in increments of 0.01 mL/min
Microbore pump head: 0.01 to 2.50 mL/min, linearly variable in increments of 0.01 mL/min
Operating Pressure 35 MPa (5000 psi) maximum
Pressure Resolution 0.07 MPa (10 psi)
High Pressure Limit 0 to 35 MPa (0-5000 psi) in increments of
0.05 MPa (7.25 psi); trips instantaneously
Low Pressure Limit 0. to 35 MPa (0 to 5000 psi) in increments of
0.05 MPa (7.25 psi); trips after a time-out of 0.4 mL for the standard bore GP40 or 0.1 mL for the microbore GP40
Delay Volume Approximately 0.92 mL for standard bore pump heads or 0.72 mL for microbore pump heads, measured from the proportioning valve up to, and including, the 61-cm (24-in) pressure transducer outlet line.
Methods Stores up to 100 separate gradient methods (00 through 99), each of which may contain up to 50 separate steps. The actual number of stored methods depends an available memory.
Control The keypad is used to select the pump's operating parameters and to program methods.
Storage Non-volatile memory protects against the loss of programs during Power-down or in the event of a Power failure.
Remote Limited remote operation via TTL-input logic level, and TTL-output and relay contact closures, or full remote programming and control through DX LAN interface.
Channels 4-channel membrane vacuum degas
Materials wetted materials, PEEK, PTFE
Performance >69% degas ratio @ 1 or 3 mL/min with 8.5 ppm 02 at input 0.200 Kg F/cm2 flow resistance @ 10 mL/min
LC 20 Chromatography Enclosure
The LC20 Chromatography Enclosure is an integral part of the Dionex DX 500 Chromatography Systems. It can house the following system components, which must be ordered separately or as part of another module: Separator columns, optional guard columns, SRSTM (Self-Regenerating Suppressor), DS3 Detection Stabilizer, and detector cell(s) for the ED40 Electrochemical Detector and CD20 Conductivity Detector.
The LC20 is available in two configurations, depending an the type of injection valve installed:
LC30, PEEK automated injection valve (P/N 044088)
LC30, stainless steel automated injection valve (P/N 044171)
Eluent from the gradient pump enters the Rheodyne injection valve port inside the enclosure. From the injection valve, eluent and sample flow through the guard column (if used), the separator column, the suppressor or other post-column device, and finally through the detector cells.
The LC20 may be configured for optional dual-channel operation. It will then include a second Rheodyne injection valve and component panel. There is enough tubing in the LC20 Ship Kit (P/N 046300) to plumb two systems.
The LC20 does not require Power. Internal components are operated by the GP40/IP20 pump.
Ambient 4 °C to 75 °C
Temperature: (40 °F to 167 °F)
Humidity: 5-95% relative (non-condensing)
Air Pressure: 400 to 800 KPa (60 to 120) psi for the Rheodyne injection valve and column select valve;
Use laboratory-quality air or regulated compressed air, nitrogen, or helium.
Operating Pressure 35 MPa (5000 psi) maximum liquid path (tubing, valves, columns, etc.)
Dimensions: 50 cm high x 22.5 cm wide x 49 cm deep (20.8 in x 8.9 in x 20.4 in)
Weight: 12 kg (25lbs)
Rheodyne Injection 35 MPa (5000 psi) metal-free or stainless steel rotary valve
Valve for low-volume injections.
Column Select Valve 28 MPa (4000 psi) metal-free column select valve to direct
eluent and sample to one of two columns. The column select valve option is installed at the factory but is not plumbed into the system. Tubing to plumb the LC20 for column select applications is included in the LC20 Ship Kit (P/N 046300).
Insulation
Dampening Dampens outside temperature variations by a factor of
- 5 to 10 over minutes
- 2 to 4 over 60 minutes
The CD20 is a sensitive, accurate, and versatile conductivity detector for detecting and quantifying ionic analytes in liquid and ion chromatography. It is especially useful for those analytes which lack UV chromophores and cannot be determined with adequate sensitivity by UV absorbance. This includes a wide variety of both organic and inorganic molecules; cations as well as anions. The major organic analytes are carboxylic, sulfonic, and phosphonic acids; and primary, secondary, tertiary, and quaternary amines. Inorganic analytes include strong acid anions such as the halides, sulfate, nitrate, and phosphate; alkali metal and alkaline earth cations. Conductivity detection, especially when combined with chemical eluent suppression, provides excellent sensitivity and selectivity for numerous ionic species, both organic and inorganic.
The CD20 Conductivity Detector is housed in a Dionex DX500 enclosure. The detector is designed to be stacked on top of other DX500 units to a recommended maximum height of four units.
Main Power 85 to 270 Vac, 47/63 Hz; 40 W Max, 25 W typical. The CD20 power supply is auto-sensing and no voltage adjustment is required.
Fuse Two 3.15 amp slow-blow IEC 127 fuses (P/N 954745).
Analog Output User-selectable full-scale Output of 10, 100, or 1000 mV.
Operating Temperature 4 °C to 40 °C (40 °F to 104 °F)
Operating Humidity 5 to 95% relative humidity (non-condensing)
Dimensions 22.5 cm W x 17.0 cm H x 42.0 cm D
(8.8 in W x 6.6 in H x 16.4 in D)
6 cm (2.4 in) clearance required behind detector.
Weight 8.2 kg (18 lb)
Display A Power-Up screen and seven basic User-Interface screens for method generation and controlo The CD20 Diagnostics Program adds eight diagnostic screens. The display is liquid crystal with adjustable backlighting
Keypad 26-key keypad for entering commands and numerical values for screen parameters.
Range 0.01 µS to 3000 µS, full scale
Temperature Compensation 0.0 to 3.0% per °C
Cell Drive Variable 8 kHz square wave
Local Operation Front panel controls and display status of all functions
Remote Operation All functions controlled by PeakNet software an a Computer. Connection is via the DX LAN. Control of 4 of 7 functions is possible via TTL or Relay contacts.
Cell Body PEEK
Active Volume 1.0 µL
Maximum Pressure 2.0 mPa (300 psi)
Electrodes 316 stainless steel
Suppiy Current 0, 100, 300, 500 mA @ 1.5 to 7.5 V.
Over-Voltage Alarm 8.5 V
Over-Temperature Alarm 40 °C
Operating Temperature 25 °C to 40 °C (50 °F to 104 °F)
Warm-up Time 10 minutes (typical)
Under/Over Temperature Not Ready message displayed when not at set temperature
Alarms
The AD20 Absorbance Detector is a dual-beam, variable wavelength photometer. Full spectral capability is provided by two light sources: a deuterium lamp for ultraviolet detection and a tungsten lamp for visible wavelength operation. The AD20 features programmable wavelength selection and automatic calibration for precision, reproducibility, and ease-of use. The AD20 is designed for use with a variety of standard bore and microbore flow cells.
The AD20 can be controlled locally from the front panel, or from a remote host computer via the optional Dionex DX LAN interface and PeakNet Software.
Main Power 90 to 265 Vac, 47/63 Hz; 150 W Max, 100 W typical. The power supply is auto-sensing and requires no voltage adjustment.
Fuse Two 3.15 amp fast-blow IEC127 fuses (1)/N 954745)
Analog Output User-selectable full-scale output of 10, 100, or 1000 mV
Operating Temperature 10 °C to 40 °C (50 °F to 104 °F)
Dimensions 22.5 cm W x 33.5 cm H x 42.0 cm D
(8.8 in W x 13.1 in H x 16.4 in D)
6 cm (2.4 in) clearance required behind the detector
Weight 13.6 kg (30 lbs)
Display Liquid crystal display with adjustable backlighting
Keypad 26 buttons for entering commands and numerical values for screen parameters
Optical System Dual-beam, fiber-optic beamsplitter and concave holographic diffraction grating
Light Sources Deuterium lamp (30 W) for ultraviolet spectrum analysis; tungsten lamp (10 W) for visible spectrum analysis
Wavelength Range 190 to 800 nm variable; continuous in 1 nm steps
Wavelength Accuracy ± 1 nm
Bandwidth 8 nm
Absorbance Discrete settings of 0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1,
Range 0.2, 0.5, 1.0, and 2.0 AU
Autozero Capability 2 AU
Autozero Accuracy ± 100 µAU
Noise < 10 µAU peak-to-peak at 254 nm, 2-second rise time, flowing deionized water
Drift < 100 µAU/hour (after warm-up)
< 100 µAUPC (alter warm-up)
Filter Rise Time Discrete settings of 0.05, 0.1, 0.2, 0.5, 1.0, 2.0, 5.0, or 10.0 seconds
Analog Output Range Discrete settings of 0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1.0, and 2.0 AUFS
Remote Operation Four of the seven AD20 functions can be controlled via TTL or Relay contacts
DX LAN Operation All functions controlled by PeakNet Software an a PC, when
(Optional) connected to the AD20 via the Dionex DX LAN interface.
Volume and Optical Path Standard bore Gell, used with 4 mm (0.016 in) columns: 9 µL
Length volume, 10 mm (0.002 in) optical path length
Narrow bore cell, used with 4 mm (0.016 in) and 2 mm (0.008 in) columns: 7.5 µL volume, 6 mm (0.002 in) optical path length
Microbore cell, used with 2 mm (0.008 in) columns: 2.5 µL volume, 3 mm (0.01 in) optical path length
Maximum Operating 1.38 mPa (200 psi) with a heat exchanger; 3.45 mPa (500 psi)
Pressure without a heat exchanger
Chemical Compatibility The use of a base stronger than 0.1 M is not recommended, as
it will etch the silica windows.
Coil for 2 mm Columns Volume of 20 µL; dispersion of 2.2 µL at 1.0 mL/min
Coil for 4 mm Columns Volume of 60 µL; dispersion of 5.3 µL at 1.0 mL/min
Onze Labexchange-team zullen u graag helpen:
Christian Schmid Laboratorium en analytiek, Laboratoriuminrichting, Life science
Hubert Sauter Spectroscopie, Chromatografie