HPLC detectoren

Gilson 155

€ 1.100,00
Excl. 19% VAT , excl. Shipping Cost
€ 1.309,00 incl. Vat
Seriennummer: 39278
Status: Stock unit?The device is in our stock, however, has to be inspected by our technician, in order to be able to guarantee complete function.
Verzending: 10-14 werkdageniFor incoming orders until 12:00 (German local time). If customs handling is required, shipment will be delayed.
Beschrijving:
Gilson UV/Vis detector 155. Folie bedieningspaneel met LCD display. 230 V. 50/60 Hz.
Installation
De installatie van dit laboratoriumapparaat voeren wij vakkundig uit. Wij denken graag met u mee en sturen u graag een vrijblijvende offerte.

Usually, laboratory and analysis devices are very exigent system with a very own specifications and requirements. Therefore, the startup of a laboratory device often rises the question how to assemble, adjust and to launch it.


We would be pleased to take over the professional installation of your laboratory devices!

Please contact us.


Benefit from the installation by us in more than one way:

  • Immediate and smooth startup of your laboratory device
  • A professional installation extends the lifetime of your laboratory device
  • Guarantee extends to 3 months
  • Freight charges do not apply*
  • You are entitled to benefit from the Labexchange Repurchase Guarantee.

* products of the product groups laboratory furniture and laminar flow are excluded. In individual cases, other product groups can be excluded.


Toevoegen aan offertelijst
Meer informatie
Objektnummer B00039278
Seriennummer 39278
Object Naam Gilson 155
Status Lagergerät
Fabrikant Gilson
Verzending 10-14 werkdagen

Status, leverings- en betalingsvoorwaarden

Apparatuurcontrole

De gebruikte apparatuur wordt voorafgaand aan levering gecontroleerd door Labexchange Service GmbH. U ontvangt volledig functionerende apparatuur.

 

Verzending

De vermelde verzendtijden zijn telkens de kortste voor een artikel. In bepaalde gevallen kunnen de daadwerkelijke verzendtijden daarvan afwijken. De uiteindelijke verzendtijden worden aangegeven in de opdrachtbevestiging.

In de regel bieden we combinatieleveringen aan. Levertijden zijn afhankelijk van het artikel met de langste levertijd. Deelleveringen zijn mogelijk tegen een toeslag.

 

Verzendmethoden

Koeriersdiensten, transportbedrijven, zelf afhalen, levering door Labexchange wagenpark

 

Informatie levering

De prijzen zijn exclusief verzendkosten. De genoemde verzendkosten zijn de te verwachten kosten. Afwijkingen zijn mogelijk. In het geval geen kosten voor verzending zijn gespecificeerd, vraag die dan afzonderlijk aan.

Voor levering in landen buiten de EU wordt voor export een toeslag van € 70,00 in rekening gebracht. Opmerking: We geven geen preferentieel certificaat/EUR1 af. Bij zelf afhalen/af fabriek (EXW) uit derde landen en de EU wordt 16% btw als borg ingehouden, tot we de ontvangstbevestiging/het leveringscertificaat van de koper hebben ontvangen.

 

Betalingsvoorwaarden

Wij accepteren geen betalingen per creditkaart, Letter of Credit, PayPal etc. Het factuurbedrag is volledig verschuldigd. Er zijn geen betalingskortingen. De goederen blijven tot volledige betaling ons eigendom.

Land

Mogelijke betaalmethoden

Opmerking

Duitsland, Oostenrijk, Zwitserland

Betaling via factuur, vooruitbetaling

Betaling via factuur is mogelijk voor ondernemingsklanten.

Nederland, België en Luxemburg

Betaling via factuur, vooruitbetaling

Betaling via factuur is mogelijk voor ondernemingsklanten.

Andere landen

vooruitbetaling

 

 

Onze Algemene Voorwaarden voor Verkoop, Levering en Betaling zijn hierop van toepassing. Deze voorwaarden zijn hier te downloaden.

Tussenverkoop is ons voorbehouden.

 

Beschrijving status:

Alle artikelen zijn gebruikte artikelen, tenzij bij een artikel uitdrukkelijk wordt vermeld dat het om een nieuw apparaat gaat.

Status

Conditie

Opmerking

Onmiddellijk verkrijgbaar

gebruikt Het artikel is al gecontroleerd en is in perfecte staat. Het kan onmiddellijk aan u worden verstuurd.
Op voorraad

gebruikt

We hebben het artikel in ons magazijn. Onze technici zullen het artikel voorafgaand aan levering controleren. U ontvangt volledig functionerende artikel.

Gepubliceerd

gebruikt

Het artikel is nog bij de aanbieder. Na uw bestelling wordt het door ons aangekocht, gecontroleerd en aan u verstuurd. Een functioneringscertificaat en een servicerapport zijn bij de levering inbegrepen.

Nieuw

nieuw

Dit is een artikel dat nieuw van de fabriek komt. Hiervoor gelden de garantiebepalingen en wettelijke garantietermijn van de fabrikant.

Firma:              Gilson

Modell:            155

Kommentar:   Dokumente engl.

The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.

Gilson 155 UV/UVIS Detector

Description

The Gilson 155 and 156 UVNIS Detectors are microprocessor-controlled absorbance detectors that provide dual wavelength detection. Both detector models are compatible components of a Gilson modular HPLC system. And, the 155 Detector with its front panel control is also compatible with other commercially available HPLC systems.

The 155 Detector can be controlled via its front panel or via TRILUTIIoN'"" LC Software. The 156 Detector must be controlled via TRILuTION LC since it does not have an integral front panel display. For information an how to control the 156 Detector, refer to the TRILuTioNTM" LC Software User's Guide.

Optics

Gilson's optical design lets you monitor absorbance at any wavelength from 190 to 700 nm.

The detector has a sensitivity range of 0.001 to 2.0 AUFS (Absorbance Units, Full Scale).

Display

The 155 Detector software is self-guiding.

All parameters, values and commands appear an the four-line display.

All of your options are displayed above the six soft keys. Press a soft key to choose and carry out a command. Numeric entries are made using the keypad.

Features

You can choose from the following modes to specify detection conditions and basic operating parameters:

•            single wavelength

•            dual wavelength

•            scan wavelength

•            setup

•            status

•            fite

Technical Specification

Autozero range

Suppresses up to 1.0 AU with 5 mm flow cell installed

Contact control

Event and autozero inputs: remote contact closure can be activated by other Instruments or the Computer

Three channels output data to a recorder.

Display

Four-line display shows modes, parameters, commands and data (155 only)

Drift

After one hour at constant temperature: 3.0 x 10 -4 AU/hour

Flow sensitivity

4.0 x 10 -4 AU for flow step of 2.0 mL/min to 0.5 mL/min with methanol

Front panel

LAMP ON/OFF hard key, LED indicator Iights for POWER, UV, VIS, REMOTE, and ERROR (155 and 156)

Six soff keys, HELP, ESC and CLEAR hard keys, arrow keys, numeric hard keys, and ENTER hard key (155 only)

Lamp warm-up

One hour maximum time

Lamps

UV Iamp: Deuterium, warranted for 750 hours or 180 days shelf-life Visible Iamp: Tungsten/halogen Iamp, warranted for 500 hours

Linearity

1%, when value is within sensitivity range of 0.001 to 2.0 AU

Flow cells

The following quartz flow cells are available:

Pressure

Flow cell

Pathlength

Volume

psi

bar

MPa

Analytical

5 mm

12 µL

500

34.5

3.45

Preparative

0.05 mm

0.16 µL

500

34.5

3.45

Preparative

0.2 mm

0.7 µL

500

34.5

3.45

Capillary

8 mm

35 nL

3000

200

20

Microbore

2 mm

1.6 µL

500

34.5

3.45

Preparative

SFC*

0.5 mm

2.0 µL

10000

689.7

68.97

SFC*

5 mm

10µL

10000

689.7

68.97

*Stainless steel

Manufacturing standards

Meets applicable Safety and EMC certification standards; UL and CE certified.

Monochromator

Dual beam, steppermotor-driven

•            Range: 190 to 700 nm

•            Spectral bandwidth: 9 nm

•            Setting accuracy: ±2 nm

•            Setting precision: ±0.2 nm

Noise with air block installed

220 nm, short term, peak to peak; 2.0 x 10 -5 AU/cm

254 nm, short term, peak to peak; 2.5 x 10 -5 AU/cm

350 nm, short term, peak to peak; 10.0 x 10 -5 AU/cm

415 nm, short term, peak to peak; 8.0 x 10 -5 AU/cm

520 nm, short term, peak to peak; 6.0 x 10 -5 AU/cm

650 nm, short term, peak to peak; 4.0 x 10 -5 AU/cm

Operating modes

Single wavelength, dual wavelength, scan wavelength, status, setup, and fite

Peak width

0.0 or 4.0 to 99.0 seconds

Physical spare

26.5 x 43.5 x 15.6 cm (10.4 x 17.1 x 6.2 in) requirement

Power requirements

Frequency: 50 to 60 Hz

Voltage: 90-240V

Current rating: 1.5A for 90-120V or 1.0A for 220-240V

Online switching regulation of line input from 80 to 260V AC

Sensitivity Range

0.001 to 2.0 AU. Sensitivity is adjustable in increments of 0.001 AU.

Software control

You can issue commands via the GSIOC or RS-232

You can use the GSIOC or RS-232 to do the following:

•            Select all detection parameters

•            Program parameter changes during run to optimize detection

•            Send digital detector output

Static RI sensitivity

Methanol versus cyclohexane at 270 nm: 5.0 x 10 -3 AU

Temperature sensitivity

3.0 x 10 -4 /°C for temperature change from 21 ° to 24°C

Weight

8.6 kg (19 lbs.)

Product groep: HPDE

Back to Top
part-of-labexchange-group
Search engine powered by ElasticSuite
lazyOwllazyOwl
Labprocure