Soldé

Varian Liberty AX II

Seriennummer: 015337
Status: Archived Product
Preis auf Anfrage
Excl. 19% VAT , excl. Shipping Cost
siehe Storeview
Descrizione:
Varian Sequentielle ICP-AES Liberty AX serie II. AGM2. Autosampler SPS5. Hotizontal plasma torch. 175-940nm. Peristaltic pump. Software. Computer. Printer.
Quotation
siehe Storeview
Maggiori Informazioni
Objektnummer B00015337
Seriennummer 015337
Nome oggetto Varian Liberty AX II
Stato Archived Product

Gruppo prodotti: ICP-Systems

Non è stato trovato alcun prodotto corrispondente alla selezione.

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

manufacturer:       Varian

model:                   Liberty AX II

annotation:           Dokumente engl.

Liberty Series II

Specifications

Environmental

Condition

Altitude

Temp

T (°C)

Humidity (%RH)

Non-condensing

Non-operating

(transport)

0-3050m

(0-10000`)

5-45

20-80

Non-operating & meeting dielectric strength tests

Sea level

40

90-95

Operating but not necessarily meeting performance spec’s

0-3050 m

(0-10000´)

5-31

31-40

80

≤{80-3.33(t-31)}

Operating within performance specifications

0-3050 m

(0-10000`)

10-35

8-80

For Optimum analytical performance it is recommended that the ambient temperature of the laboratory be between 20-25 °C and be held constant to within ±2 °C throughout the entire working day.

Suitability

The Instrument is designed for indoor use only. It is suitable for the following categories:

q          Installation category il

q          Pollution degree 2

q          Safety Class 1 (EN 61010-1) 1.3.2 Power

Power

All power supplies should be single phase AC, 3 wire system (active, neutral, ground or two active and ground) and should be terminated at an appropriate connection receptacle that is within reach of the system power Gable assembly.

System unit

Required supply voltage

Nominal Rating

Spectrometer

(Low power)

208, 220 or 240 VAC, ±10%

230 VAC +6/-14%

230 VAC +14-6%

Frequency 49 to 61 Hz

650 VA

(High power)

187 to 264 VAC,

Frequency 49 to 61 Hz

4.4 kVA

Computer

115,120, 220,240 VAC, 50/60 Hz

(Typical)

300 VA

Laser printer

115,120,220,240 VAC, 50/60 Hz

(Typical)

85 VA

Connections

Low power mains inlet coupler: 6A 250 VAC IEC type

Low power mains power cord:

  Australia 10A 250VAC Complies with AS3112

  USA    10A 125VAC Complies with NEMA L6-15P

  Europe 6A 250VAC Complies with CEE7 sheet vii or

N FC61.303VA

High power mains inlet:

  Flying lead 2 m (6 ft 6 in) long

  3 core, single phase, 3/74/0.3 mm, 6 mm 2 (approx 10AWG)

High power mains power cord connector:

Connection to be made according to local regulations.

Other electrical connections:

IEEE 488

Chart recorder Output, 3.5 mm tip/sleeve jack

Fuses

1FS1 & 1FS2 T6.3AH250V, IEC 127 sheet 5, 5 x 20 mm

1C131              Circuit breaker 30A Fast trip

Note     For safety reasons, any other internal fuse or circuit breaker is not Operator accessible, and should only be replaced by Varian authorized personnel.

Fuse Information an the rear of the Instrument is the most up to date.

Gas supplies

Liquid or gaseous argon may be used with the Liberty Serien 11 System. The use of liquid argon is recommended, since it is purer, more convenient and cheaper per unit volume.

Purity

99.996%

Oxygen

< 5 ppm

Nitrogen

< 20 ppm

Water

< 4 ppm

Permissible pressure range

410 to 690 kPa (60 to 100 psi)

Recommended pressure

550 kPa (80 psi) regulated

Reauired flow raten

0.7 to 32.25 litres per minute

Connection:

1.8 m (6 ft) length of 6.3 mm (1/4 in) ID hone with 9/16 in-18UNF thread fitting.

Other gas connections:

Torch compartment: barb type connector for argon torch main flow.

Sample compartment: push-on connector for argon torch auxiliary flow

push-on connector for argon nebulizer

Water supply

                                                                      Radial                        Axial

Cooling requirements                                    0.2kW                         1 kW

Minimum flow rate                                          0.5 L/min                    1.6 L/min

Inlet temperature, recommended                 20 °C                          20 °C

Inlet pressure, minimum                                55 kPa (8 psi)             55 kPa (8 psi)

Inlet pressure, maximum                               400 kPa (60 psi)        400 kPa (60 psi)

Inlet and outlet connection for both radial and axial systems:

1.8 m (6 ft) length of 6.4 mm (1/4 in) ID hose.

Weights and dimensions

Weight

Packed                 325 kg (717 lb)

Unpacked             226 kg (500 lb)

Dimensions (W x D x H)

Packed                 1660 x 960 x 1420 mm (66 x 38 x 56 in)

Unpacked             1380 x 710 x 1100 mm (55 x 28 x 42 in)

AGM-2 Module

At wavelengths below 189 nm, oxygen in the optical system of the Liberty ICP-AES beginn to absorb the energy emitted from the plasma. To perform measurements below 189 nm requires the purging of the optical system with an inert gas. This removes any oxygen present, allowing the far UV wavelengths to be used.

The Auxiliary Gas Module (AGM-2) is an accessory designed to extend the wavelength range of the Liberty 100 ICP-AES spectrometer. lt does this by purging both the monochromator and the pre-optics snout assembly with nitrogen. This extends the wavelength range down to 175 nm, allowing the Liberty 100 to be used for the determination of important additional elements, such as sulfur.

The Liberty 100, fitted with the AGM-2 accessory is the best alternative if the far UV region is used infrequently. The Liberty 200 vacuum monochromator provides the best detection limits and stability in the far UV if that region is often used.

SPS-5 Autosampler

The SPS-5 Autosampler is designed for use with several of Varian's Instruments:

q          Liberty ICP-AES spectrometers

q          Cary UV-VIS-NIR spectrophotometers

q          SpectrAA spectrometers

This manual describes how to unpack and install the sampler. For details on the software Interface to the sampler you should consult the relevant Instrument manual.

The SPS-5 can fit up to 500 samples, with the extension platform fitted and five type 10 racks. Several different rack types are available. Three racks may be fitted on the standard unit. The extension platform must be fitted to allow five racks to be used.

The SPS-5 features the automatic Varian Magnetic Rack Recognition system. This means that the SPS-5 knows which rack type is in each rack position. This is done with a magnetic identification system.

A cover is available for use with corrosive or dangerous samples.

Note Moving the SPS-5 arm or probe arm of the SPS-5 during Operation will sause a loss of alignment. To realign, reset the SPS-5 as described in section 3.4.

Specifications

Environmental

Your accessory is designed for indoor use only. lt is suitable for the categories stated on the front of this manual.

The site should be selected to avoid dusty or corrosive atmospheres.

Note     Extra maintenance may be required on the SPS-5 if it is operated in excessively dusty or corrosive conditions.

Condition                              Attitude                 Temp                  Humidity (%RH)

                                                                            t(°C)                     non- condensing

Non-operating                        0-2133 m                5-45                      20-80

(transport)                              (0-7000')

Non-operating &                     sea level                 40                         90-95
meeting dielectric

strength tests

Operating but not                   0-2000 m                5-31                      ≤80

necessarily meeting               (0-6562')                 31-40                    ≤(80-3.33(t-31)} performance spec's

Operating within                     0-853 m                  10-35                    8-80

performance                          (0-2800')

specifications                         853-2133 m            10-25

(2800-7000')

Power

Voltage             100, 120, 220 or 240 volts AC ±10%

230 +14% -6% volts AC

230 +6% -14% volts AC

Frequency       50-60 Hz ± 1 Hz

Consumption   ~ 80 VA

Electrical connections

Mains inlet coupler: 6A 250VAC IEC type

IEEE 488

Accessory, diluter: 15 way D-range type.

Mains power cord connector:

  Australia 10A 250VAC Complies with AS3112

  USA    10A 125VAC Complies with NEMA 5-15P

  Europe 6A 250VAC   Complies with CEE7 sheet vii or

NFC61.303VA

Fuses

T2AH250V IEC 127 sheet 5, 5 x 20 mm (220-240 VAC)

T2.5AH250V IEC 127 sheet 5, 5 x 20 mm (100-120 VAC)

Note     For safety reasons, any other internal fuse or circuit breaker is not Operator accessible, and should be replaced only by Varian authorized personnel.

Fuse Information an the rear of the Instrument is the most up to date.

Weights and dimensions

Weights

                                      Packed                       Unpacked

                                      kg (1b)                        kg (lb)

SPS-5                            35 (77)                        19 (42)

Diluter                             5 (11)                          3 (7)

Dimensions (W x D x H)

                                      Packed                       Unpacked

                                      mm (in)                      mm (in)

SPS-5 only                     580 x 670 x 660          530 x 450 x 600

                                      (23 x 26 x 26)             (21 x 18 x 24)

diluter only                      300 x 180 x 570          160 x 80 x 410

                                      (21 x 7 x 22)               (6 x 3 x 16)

SPS-5 with diluter                                            530 x 510 x 600

(arm raised)                                                     (21 x 20 x 24)

SIPS-5 with rear                                               530 x 600 x 600

rack extension                                                  (21 x 24 x 24)

(arm raised)

Part of labexchange logo
Search engine powered by ElasticSuite