Thermo Separation CE-Ultra
| Objektnummer | B00013614 |
|---|---|
| Seriennummer | 013614 |
| Nome oggetto | Thermo Separation CE-Ultra |
| Stato | Archived Product |
Gruppo prodotti: Elettroforesi Capillare
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
|
Country |
Possible payment methods |
Comment |
|
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
|
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
|
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
|
Status |
Condition |
Comment |
|
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
| In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
|
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
|
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
|
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
Firma: Thermo Separation
The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.

Introduction
Capillary electrophoresis (CE) is the movement of ions through a capillary subjected to an electrical field. Sample ions are introduced hydrodynamically (using pressure differences) or electrokinetically (using charge) to draw sample into a capillary. Once the sample is injected, the end of the capillary is removed from the sample reservoir and immersed in the running buffer. Separation is based an the charge-to-mass ratlos of the constituent ions. As the sample migrates through the capillary, a variable-wavelength detector monitors the presence or absence of the constituents of interest. The detection data is then sent to a data acquisition system, which creates a recording of the electrophoretic separation, an electropherogram or E-gram TM , of the run.
SpectraPHORESIS ULTRA System Operation
A typical automated CE system contains seven functional components: 1) fused-silica capillary, 2) electrolyte vessels, 3) electrodes, 4) pressure system. 5) regulated high-voltalte power supply, 6) detector, and 7) data acquisition system. The following Instrument modules comprise the SpectraPHORESIS ULTRA System:
SpectraPHORESIS VialServer (Autosampler): The VialServer holds up to 110 sample or buffer vials in peeparation for CE analyses. The optional Tray Temperature Control function maintains up to 95 vials at temperatures from 4° - 60°C. The vial positions can be prearranged and easily coordinated with the acquisition method defined in the PC1000 System Software. The VialServer's Sample/Buffer Preparation function allows for automated pre-run huffer preparation and post-run huffer replenishment.
SpectraPHORESIS ULTRA (CE Module): Electrophoretic separation occurs in the CE module. The CE module shuttle automatically slides into the side access door in the VialServer to ohtain buffer and sample vials as specified in the sequence method. The shuttle then slides hack into the CE module to transport the required vials to the front and rear vial positions in the injection tower. There they are raised onto the ends of the fused-silica capillary housed in the capillary cassette. The cassette is inserted into the capillary oven, which controls the capillary temperature throughout the run.
The capillary is threaded between two ball lenses. Prior to injection, and as the vial is lifted into the injection tower, the vial septa are pierced by the needle. The vials are then raised into the final preinjection position, causing the electrode, and capillary to protrude through the vial septa and become immersed in the vial contents. The injection tower needle septa combined with vial and cassette septa work together to maintain a stable injection pressure, minimizing backflushing into the needle and loss of precious sample. This unique tri-seal configuration contributes to the 2% RSD precision offered by the CE ULTRA System. During hydrodynamic injection, the positivepressure pneumatic system forces sample into the capillary. During electrokinetic injection, the high-voltage power supply provides potential up to 30kV to draw the sample into the capillary.
SpectraPHORESIS UV3000: Separated sample constituents pass through the dual-ball sapphire lens assembly located in the lower portion of the cassette. Either deuterium (UV 190-365 nm) or tungsten (visible 366-800 nm) light emerging from the capillary is sent to the sample diode in the SpectraPHORESIS UV3000 scanning detector. A reference signal is simultaneously sent to the detector reference photodiode. The detector can run in single, dual, triple wavelength, or scanning modes.
PC1000 Software: The SpectraPHORESIS ULTRA modules are connected to a personal Computer where instrument control, acquisition and calculation method setup, sequence and vial assignment, data acquisition, data processing, and reporting are Bone using Thermo Separation Products" PC1000 Software Ver 3.5, or higher. Refer to the PC1000 System Software Reference Manual shipped with your SpectraPHORESIS ULTRA System for further information an PC1000 software.
PC 1000 processes the data acquired from the SpectraPHORESIS UV3000 to perform peak detection and quantification. The software also provides real-time Status information throughout the run and can create a post-run report and an electropherogram of the Separation. Post-run reprocessing of previously acquired data is also possible for further method refinement.
Specifications
Safety Features: Interlock disabling of high-voltage CE module
Power Requirements: 100/120/240 VAC nominal. 50/60 Hz
Ambient Environment: 10 to 40 °C; 5%-95% RH non-condensing
Bench Space without PC: 44 x 92 52 cm (H x W x D) cm total bench spare without computer
SpectraPHORESIS VialServer
Maximum Vial Capacity: 110 vials, 95 with optional tray thermostatting
Tray Temperature Control: Optional 0-60 °C, Peltier controlled
Needle Height: Programmable in 0.1 mm increments
Vial Dimensions/Volume: 35 x 11.5 mm. Blass or plastic (ca. 1.8 mL glass, ca 650 µL and 100 µL plastic, screw capped. 100 µL. glass microinserts).
Sample/Buffer Preparation
Reagent Vials: Up to three vials per method
Mixing/Heating: Single vial. air-driven vortex heater/mixer. settahle in 1°C increments form 30° to 100 °C
Dilution Precision: Typically <1.0% RSD for 1/100 dilution (10 µL to 1 rnL)
Dispensing Precision: Typically <0.2% RSD for 1 mL dispensing from external solvent reservoir
Dimensions and Weight: 37 x 41 x 51 cm (H x W D); 16 kg Standard. 24 kg with all options installed
SpectraPHORESIS ULTRA (CE Module)
Vial Shuttle Capacity: Up to 2 x 5 vials
Minimum Sample Volume: 20 µL pressure injection. 50 µL electromigration sampling (using microinserts)
Wash 100 psi pressure, applicable to either of the two capillary ends
Capillary Temperature Control:
Peltier controlled, 10°C to 60°C in 0.1°C increments. Maximum ringe of cooling 10°C below amhient. Modes: constant temperature. temperature gradients, constant resistance
Run Modes:
Constant voltage, constant current, voltage gradients, positive or negative polarity, time-programmed switches. Apply pressure at one or both ends of a capillary with the voltage or current applied.
Capillary Cassette: Bar code labeled. Minimum capillary length ca. 20 cm. Maximum length 100 cm. Shortest detection length (Ld) 5 cm.
Dimensions and Weight: 44 x 36 x 51 cm (H x W x D); 27 kg.
SpectraPHORESIS UV3000 (Detector)
Modes: Time-programmable single wavelength, multiple wavelengths. scanning
Wavelength Range: 190 - 800 nm
Wavelength Accuracy: ± 1 nm
Wavelength Precision: 0.1 nm
Lamps: Deuterium and tungsten, pre-aligned
Optics,: Fiber optics. ball lenses
Rise Time: 0.0 to 9.9 sec
Noise: 2.5 x 10 -5 AU @ 210 nm
Dimensions and Weight: 37 x 15 x 47 cm (H x W x D); 11 kg
PC1000 System Requirements
COMPUTER HARDWARE
Refer to the TSP README file for the specific system requirements. These include
CPU 486 66MHZ or higher
Monitor SVGA or higher
Math Coprocessor Required
Memory 16 Mbytes (minimum), 32 MB strongly recommended
Disk Space 10 Mbytes for software
Data requirement determined by use (Scanning files use 1MB/hr during data acquisition.)
COM Ports UART 16550 AF supporting extended hardware buffering
SOFTWARE REQUIREMENTS
Operating System OS/2 Warp! Ver 3.0 or higher
SOFTWARE CAPACITIES
Run Time per Injection 655.35 minutes maximum;
120 minutes maximum for scanning
Injections per Sequence 999
Calibration Levels Up to 15
Components 750 identified peaks or 1500 unidentified
Integration 3 signals selectabie for integration
Internal Standards Up to 3
Sample Names 16-character name with 65-character description
Method Names 99 characters including path and fite extension
Component Names 15 characters