Archivio

Tecan SpectraFluor Plus

Seriennummer: 018170
Stato: Archived Product?The unit is no longer available.
Spedizione:
Descrizione:
siehe Storeview
Maggiori Informazioni
Objektnummer B00018170
Seriennummer 018170
Nome oggetto Tecan SpectraFluor Plus
Stato Archiv

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

Manufacturer:             Tecan

Model:                         Spectrafluor pl

 

 

 

The SPECTRAFLUOR PLUS is a fully automatic, computer controlled fluorometer for measuring samples in a microplate. The instrument fulfills the requirements for numerous fluorescence, absorbance and glow type luminescence applications in research and routine investigations.

The filters are arranged in filter slides; four excitation filters and four emission filters respectively. The instrument has no on-board software and is exclusively controlled via an external computer using WindowsTMformat.

 

 

 

1 = power LED

2 = emission filter

3 = excitation filter

4 = plate carrier

 

 

 

 

Instrument Features

 

General

The following types of measurement are provided with the SPECTRAFLUOR PLUS microplate reader.

 

 

Measurement Type             Read Mode                           Microplates to be Used

Fluorescence                         Standard and time-                6 - 1536 wells

                                               resolved, top and bottom

Absorbance                            Single and dual                       ≤ 384 wells

                                               wavelength

Glow type luminescence                                                       96 wells

 

 

The table below lists the technical specifications for the instrument.

 

Parameters                           Characteristics

Measurement                         Software controlled

Interface                                 Serial interface: RS 232

Filter handling                         External filter exchange

                                               4 excitation filters and 4 emission filters in separate filter slides.

                                               Up to 8 excitation and emission filter slides can be defined            when using software.

Plate definition                        Via scanning software

Temperature control              From ambient +5°C up to 42°C

Plate shaking                          Orbital or linear shaking - both offer 3 speeds

Light source                           High energy Xenon flash lamp

Optics                                     Quartz fibers and lenses

Detector                                 Side window, low dark current photomultiplier tube

Power supply                         Auto-sensing:100-120V / 220-240V 50/60 Hz

Power consumption               200 VA

Main fuse                                T 2A / 250 (slow blow)

 

Physical

Outer dimensions                  Width: 380 mm

                                               Depth: 415 mm

                                               Height: 170 mm

Weight                                    18.8 kg

 

Environmental

Ambient temperature

Operation                               15°C –30°C    59°F- 86°F

Non-operation                         -20°C -+60°C-4°F - +140°F

Relative humidity

Operation                               90% non-condensing

Overvoltage category             II

Pollution degree                     2

Usage                                     Commercial

Noise level                              < 70 dBA

Method of disposal                 Electronic waste (infectious waste)

 

 

Fluorescence

Parameters                   Characteristics

Sensitivity                      ≤ 1.5 pg fluorescein / well (200 µl / well) or ≤ 4 fmol / well

                                      in a Greiner 96 well black plate using 100 flashes / well

Wavelength Range       230 - 700 nm

Standard Filters            Wavelength         Band Width1)

Excitation                      1. 360nm             35nm

                                      2. 430nm             35nm

                                      3. 485nm             20nm

                                      4. 590nm             20nm

Emission                       1. 465nm             35nm

                                      2. 535nm             25nm

                                      3. 595nm             35nm

                                      4. 635nm             35nm

Read Mode                   Top and bottom (bottom for transparent plates)

Integration Time2)          20 - 2000 µs

Lag Time2)                    0 - 2000 µs

 

1) Band width for standard filters may have some tolerance ±10%

2) Needs to be adjusted for Time Resolved Measurements. Lag Time is typically ≥ 40µs

Gain setting                     Values                     Measurement range

Manual                               1 -255                       0 - 60,000 RFU

Optimal                              Automatic                 0 - 60,000 RFU

Extended Dynamic            Automatic                 0 - 6,000,000 RFU

Range                                                                 (using XFLUOR)

 

 

No. of                    Definition                                Characteristic             Meas. Time for Flashes/ Well                                                                                        96 Wells

1 - 3                        Plate carrier does not stop       max. speed                   ≤ 30 sec

                               at the measurement point

                               (on the fly)

4-255                      Plate carrier stops at the          max. precision              Dependent on the      meas. point                                    flash parameter                                                                         setting

 

 

 

Absorbance

The specifications only apply to 96 well plates.

Absorbance can not be measured in 1536 well plates.

The absorbance filters are not within the instrument's standard configuration.

 

 

Parameters                  Characteristics

Wavelength range        230 - 750 nm

Meas. range                  0-3 OD

Resolution                     0.0001 OD

Accuracy                       0 - 2.0 OD    ± 2.0 % and ± 0.010 OD        492 nm

Precision                       0 - 2.0 OD    ± 2.0 % and ± 0.005 OD        492 nm

Linearity                        0 - 2.0 OD    ± 2.0 %                                   492 nm

                                      0-1.5 OD      ± 2.0 %                                   260/340 nm

Standard filters             405 nm / 450 nm / 492 nm / 620 nm

                                      (not included in standard delivery)

 

Glow Type Luminescence

 

The specifications only apply to 96 well plates.

Luminescence can only be measured in 96 well plates.

 

Parameters                  Characteristics

Sensitivity                      < 0.4 10-6DEA activity units/well of molecular biology grade alkaline       phosphatase (20 µl/well) using CDP-Star (Tropix) chemiluminescent            substrate mix with Emerald-II enhancer (100 µl/well) in a Greiner 96                     well white plate. Use of the cover plate greatly reduces the cross-talk.

                                      0.4 10-6DEA activity units/well correspond to 200 fg/well of alkaline        phosphatase when using a lot of 2000 DEA activity units/mg protein.

Wavelength range        400 - 700 nm

Gain                              Optimal:       preset value

                                      Manual:         1 -255

 

Integration setting               Integration time                    Measurement range

Manual                                    1 - 5000 ms                            0 - 55,000 RLU

Maximum Dynamic                automatic (1 - 1000 ms)        0 - 55,000,000 RLU

Range                                                                                    (using XFLUOR)