Archivio

Spectra Physics SP 8780

Seriennummer: 015503
Stato: Archived Product?The unit is no longer available.
Spedizione:
Descrizione:
Spectra Physics Autosampler SP 8780 for HPLC. 4 Trays for each 20 Vials. Rheodyne-valve. RS 232-Interface.
siehe Storeview
Maggiori Informazioni
Objektnummer B00015503
Seriennummer 015503
Nome oggetto Spectra Physics SP 8780
Stato Archiv

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

manufacturer :       Spectra Physics

model :                   SP 8780

annotation :           Dokumente engl.

 

 

Specifications

•      -Full-loop injection modes

-           Fixed-loop 20 µL standard; 10, 50, 100, 200, and 500 µL loops available.

-           Precision @ 20 µL, 0.5% or less. Typical values based an 10 runs.

      Provides maximum reproduc­ibility and precision.

•      Variable injection modes

-           Variable volume: 1 µL to 250 µL in 1 µL increments.

-           Precision: 10 µL, 1% or less; 3 µL, 3% or fess. Typical values based on 10 runs.

-           Minimum volume required: 5 µL for a 1 µL injection using Spectra-Physics microvials.

Provide versatility to change injection volumes without chang­ing loops.

•      Cooler option available for 4-15°C operation.

•      1-10 injections per vial.

      Sample vial: 2 mL, glass standard; optional microvial available.

      Conserves precious samples.

•      Communications option: includes RS-232-C, LABNET, BCD, and four programmable timed events.

      Provides interface to laboratory Computers.

•      Bar-code label reading option.

      Provides positive sample identifi­cation and automatic method selection.

•      80-sample capacity in four removable trays.

      Provides continuous operation.

•      Quadrant and position identifi­cation.

      Helps comply with Good Labratory Practices.

•      Sample trays can be numbered 1-20 and read automatically.

      Increase convenience and help you comply with Good Labora­tory Practices.

•      Priority samples can be run without disturbing automatic sequence.
Greater efficiency and convenience.

•      The same vial(s) can be used repetitively for calibration and re-calibration via LABNET or RS232 set up.

      Saves vial positions for samples.

•      Automatic flush after each sample.

      Eliminates sample carryover.

•      Variable sample draw rate.

      Easily accommodates viscous samples.

•      Sample size, number of injections and analysis file can be set for each vial.

      Allow you to run different methods automatically.

•      Gradient delay Feature up to 99.99 minutes.

      Allows you to duplicate methods developed on other LC systems.

•      Equilibrium delay up to 99.99 minutes.

      Ensures column stability.

•      Standard inputs/outputs:

      Inject hold input

      Gradient Start output

      Pump stop output

      Inject signal output

      +5 volt, 30 mA output

      Provicle flexibility for a variety of applications.

 

Power requirements:

90-130 V, 50/60 Hz or 200/265 V, 50/60 Hz.

 

Environment:

+10° to +40°C; 20% to 95% non-condensing relative humidity.

 

Dimensions:

Width: 15.0 inches      (38 cm)

Height: 9.75 inches     (25 cm)

Depth: 22.0 inches      (56 cm)

Weight: 43.0 lbs.         (19.5 Kg)

 

 

Gruppo prodotti: HPLC-Autocampionatori