Soldé

Shimadzu UV 1601

Seriennummer: 023438
Status: Archived Product
Preis auf Anfrage
Excl. 19% VAT , excl. Shipping Cost
siehe Storeview
Descrizione:
Shimadzu UV / Vis spettrofotometro UV-1601. 190-1100 nm, 230 V, 50/60 Hz, 160 W. display integrato e pannello di controllo. Doppio trave. Con il supporto della cella per cuvette mm 10x10. Diverse interfacce. Slot per scheda IC. Con il computer. Win95. Software Hyper Shimadzu UV. Marchio CE.
Quotation
siehe Storeview
Maggiori Informazioni
Objektnummer B00023438
Seriennummer 023438
Nome oggetto Shimadzu UV 1601
Stato Archived Product

Gruppo prodotti: Fotometri / Spettrofotometri

Non è stato trovato alcun prodotto corrispondente alla selezione.

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

Hersteller:    Shimadzu

Modell:          UV-1601

 

 

Nachfolgende Abbildungen und Beschreibungen sind modellbezogen und aus Prospekten entnommen.

Sie geben nicht den Lieferumfang des Systems wieder.

Den exakten Lieferumfang entnehmen Sie bitte aus dem Angebotstext.

 

 

 

 

Hardware-Spezifikationen

 

 

Messwellenlängenbereich

 

190 ~ 1100nm

 

Spektrale Bandbreite (Auflösung)

 

2 nm

 

Wellenlängenanzeige

 

0,1 nm-Einheiten

 

Wellenlängeneinstellung

 

0,1 nm-Einheiten (1 nm-Einheiten in Scan)

 

Wellenlängengenauigkeit

 

± 0,5 nm Bord automatische Wellenlängenkalibrierung Mechanismus

 

Wavelength Wiederholbarkeit

 

± 0,1 nm

 

Wellenlängen-Scan-Geschwindigkeiten

 

GOTO WL-Befehl:   ca. 6000nm / min

Sehr schnell:            ca. 3200nm / min

Schnell:                    ca. 2200nm / min

Medium:                   ca. 370 nm / min

Langsam:                 ca. 260 nm / min

Sehr langsam:          ca. 160 nm / min

 

Abtastintervall

 

1,0 nm (2,0 mm): 910nm> = Lambda-Bereich> = 500 nm

0,5 nm (1,0 mm): 500 nm> = Lambda-Bereich> = 200 nm

0,2 nm (0,5 mm): 200 nm> = Lambda-Bereich> = 100 nm

0,1 nm (0,2 mm): 100 nm> = Lambda-Bereich

Das Abtastintervall in () gibt die ein zu "Very Fast" Wellenlängenscangeschwindigkeit.

(Wo in ist "Lambda-Bereich" die Differenz zwischen dem Messstartwellenlänge und dem Messende Wellenlänge.)

 

Lichtquelle Schalt

 

Automatische Umschaltung mit Wellenlängenbereich. Kann überall im Bereich von 295.0nm zu 364.0nm eingestellt werden (340.8nm wird empfohlen)

 

Streulicht

 

Weniger als 0,05%

 

Belichtungsmessung

 

Doppelstrahloptik

 

Photometriebereich

 

Absorption: -0,5 ~ 3,999 Abs (bei ​​unkorrigierten baselie Kurve innerhalb 0.5Abs)

Durchlässigkeit: 0 ~ 300%

 

Registrierbereich

 

Absorption: -3,99 ~ 3.99Abs

Durchlässigkeit: -399 ~ 399%

 

Photometrische Genauigkeit

 

± 0.004Abs (at 1.0Abs)

± 0.002Abs (at 0.5Abs)

 

Photometrische Reproduzierbarkeit

 

± 0.002Abs (at 1.0Abs)

± 0.001Abs (at 0.5Abs)

 

Auto-Null-Funktion

 

[AUTO ZERO] -Taste ermöglicht One-Touch-Einstellung

 

Basislinienstabilität

 

± 0.001Abs / h

 

Baseline Ebenheit

 

± 0.002Abs

 

Geräuschpegel

 

0.002Abs

 

Basislinienkorrektur

 

Automatische Korrektur mit Computerspeicher

2-Level-Baseline-Korrektur (Hochpräzisionsinstrument Basislinienkorrektur-Funktion)

 

Lichtquelle

 

50W Halogenlampe (mit langer Lebensdauer 2000 Stunden), Deuterium-Lampe (Buchsentyp), Bord automatische Lichtquelle Positioniermechanismus

 

Monochromator

 

Verwendet aberrationskorrigierende konkaven loderte holographische Gitter

 

Detektor

 

Silizium-Photodiode

 

LCD

 

B acklit (320x240 dot), Kontrast einstellbar

 

Probenraum

 

Innenmaße 110x230x105 (mm)

(BxTxH) (Teiltiefe 155mm)

Die Entfernung zwischen Balken 100mm

Abnehmbare Typ-2-Verschraubung

Strahlabmessungen 1x10mm (BxH)

(in der Mitte des Probenraum)

 

Energieversorgung

 

100, 120, 220, 240

50 / 60Hz 160VA

 

Abmessungen

 

550x470x380 (200 bei Abschluss LCD-Einheit) (mm) (BxTxH)

 

Last

 

18kg

 

Umgebungstemperatur

 

Raumtemperatur 15 ~ 35 ° C

 

Luftfeuchtigkeit

 

Luftfeuchtigkeit 45 ~ 80%

 

 

 

 

 

 

 