Soldé

Shimadzu Doppel-Ionen

Seriennummer: 016642
Status: Archived Product
Preis auf Anfrage
Excl. 19% VAT , excl. Shipping Cost
siehe Storeview
Descrizione:
Shimadzu Ion-Chromatograph. 2 x Conductivity detector CDD-6A. Controller SCL-10AVP. 2x Double piston pumps LC-10ADvp. Degasser DGU-14A. Column oven CTO-10AVP.Autosampler SIL-10CE. Software ClassVP. Computer. Screen.
Quotation
siehe Storeview
Maggiori Informazioni
Objektnummer B00016642
Seriennummer 016642
Nome oggetto Shimadzu Doppel-Ionen
Stato Archived Product

Gruppo prodotti: Cromatografia Ionica

Non è stato trovato alcun prodotto corrispondente alla selezione.

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

manufacturer:       Shimadzu

model:                   Doppel-Ionen GC

annotation:           Dokumente engl.

CDDA-6A Conductivity Detector


1 Display

2 Display select switch

3 Polarity select switch

4 Response select switch

5 GAIN select switch

6 RANGE select switch

7 POWER pilot lamp

8 ZERO OVER pilot lamp

9 REC. SHORT pilot lamp

10 ZERO

11 FINE ZERO

12 REC. SHORT

13 MARK


Specifications

Cell volume:                           0.25µL

Cell constant:                         25cm -1

Materials in contact with

liquid:                                     SUS 316, PTFE, PCTFE

Cell pressure tolerante:         30kg/cm 2

Measuring sensitivity:             0.1 ~ 5120 µScm -1 FS

                                              Determined by [gain setting] X [range setting]

Gain:                                       0.1, 1, 10µScm -1 3-step selection

Range:                                   1, 2, 4, 8, 16, 32, 64, 128, 256, 512 10-step selection

Zero adjustment:                    Auto zero by pushbutton switch

Zero adjustment range:          4000µScm -1 (with the gain set of 0.1 or 1µScm -1 )

                                              9000µScm -1 (with the gain set of 10µScm -1 )

Zero adjustment accuracy:    ± 10/N% (N: Range)

Baseline positioning:              By FINE ZERO key switch (effective for RECORDER output terminals only)

Fine zero adjustment range: Beyond the recorder full scale

Response:                             FAST, STD, SLOW 3-step selection

                                              Corresponding time constants are 0.05, 0.75 and 1.5 sec.

Noise:                                     0.004µScm -1 max.

With the response at STD when 2.5mM phthalic acid + 2.4mM Tris (hydroxymethyl) aminomethane solution at PH 4.0 (con­ductivity: approx. 250µScm -1 ) is fed at the rate of 1.5mL/min. into the Shim-Pack IC-A1 Column to be used in the HIC-6A

                                              Ion Chromatograph.

Drift:                                       0.025µScm -1 /h max. (Under the same conditions as above and
at constant ambient temperature.)

Drift by ambient tempera-      0.025µScm -1 /°C max. (Under the same conditions as noise

ture change:                           measurement)

Recorder:                               1mV, 10mV

Integrator terminal output:      The signals of 10, 1 & 0.1mV/µScm -1 are output according to the gain settings of 0.1, 1 & 10µScm -1 .

Optimum operating

temperature range:                10°C ~ 35°C

Size:                                       Controller         25 (W) X 16 (H) X 40 (D) cm

                                              Cell                   10 (W) X 19 (H) X 4 (D) cm

Weight:                                   Controller:        7kg       Cell: 0.8kg

Optimum operating

temperature range:                10°C ~ 35°C


SCL-10AVP

Features

This system controller is designed to be connected to the various units in the Shimadzu LC-10A/10Avp High Performance Liquid Chromatography series, and control their functions.

This controller centralizes control of a wide variety of system configurations, and can also permit fully automated operation.

•    By means of specially designed fiber optic link Interfaces, the controller can be connected to the pump, column oven, detector, auto injector and fraction collec­tor in the LC-1 OA/1 OAvp series.

Use of a sub-controller will also permit the controller to be connected to and con­trol options such as special valves and a helium degasser.

•    An fiber optic link interface for connection to a Chromatopac is supplied. This interface permits operation in tandem with a Chromatopac, or system control by BASIC programs written an the Chromatopac.

•    An optional RS-232 Gable can be used to connect the controller to a PC, for PC control. (When the controller is under PC control, the fiber optic link interface cannot be used.)

•    The controller's LCD screen permits setting analysis conditions for the connected devices, provides displays relating to control operation, and displays the status of each device for monitoring puposes during analysis.

    Additionally, the screen's graphic functions can provide graphic displays of scan data from the detector, chromatograms from the fraction collector, and other data that facilitate parameters settings.

Specifications

1. Display panel

 

LCD wich back light

320 x 240 dots

 

2. Operation panel

 

Keys

Fixed function keys

22 keys

 

Variable function keys

5 keys

 

3. Flexible disk drive

 

Drive

Double sided high density

 

Diskette

3.5" 2HD

 

Capacity (formatted)

approx. 1.6 MByte.

 

4. Input terminals

 

External START signal input (MAN.INJ IN)

x 1 (relay contact input)

Actuated by time program or by

START, STOP or ERROR

Signals, as applicable

 

ERROR signal input (ALARM IN)

x 3 (relay contact input)

 

Common output terminals (EVENT1, 2, 3, 4)

x 4 (relay contact output)

 

External power outlet control (AC REMOTE)

x 1 (for multiple terminal box)

 

Optical link (OPT LINK) connector

x 1 (for Chromatopac)

 

RS-232C connector

x 1 (for connection to PC)

 

REMOTE connector

x 8 (for controlling HPLC units)

 

5. Compatible HPLC units

 

Pumps

Up to 3 units

 

Auto injector

1 unit

 

Column oven

1 unit

 

Detectors

1 or 2 units

 

Fraction collector

1 unit

 

Sub-controller / valve Interface

1 unit

 

Equipment below can be connected via sub-controller (or valve interface)

 

2-position rotary valves

2 units

 

6-position rotary valves

2 units

 

Solenoid valve unit

1 unit (cannot be connected via valve Interface)

 

Helium degasser

1 unit (cannot be connected via valve interface)

 

Syringe unit

1 unit (cannot be connected via valve interface)

 

6. Parameter fites

 

Analysis files (parameters, time programs)

Number of files

20

 

Time program steps

400 (total of all files)

 

Program time

0.01 - 9999.9 minutes

 

Fraction collector files (parameters, time programs)

Number of files

10

 

Time program steps

100 (total of all fites)

 

Program time

0.33 - 9999.9 minutes

 

Auto injector analytical sequence file

Number of steps

102

 

Auto injector

pre-treatment files

Number of files

20

 

Program memory capace

250 Bites for each file

 

7. Data files

Scan data files

Number of files

3 (one for background, two for samples)

* Reception and display of scan data from SPD-10A(V)vp possible

Chromatogram files

Number of files

1 (for the fraction collector)

Reception and display of chromatograms from FRC-10A possible

8. Pump control

control modes

Isocratic, high-pressure gradient, low-pressure gradient, constant pressure

Parameters

Flow rate, pressure, concentration, upper and lower pressure limits

Gradient profiles

Step, linear and exponential functions (all at multiple levels)

9. Auto injector control

Parameters

Sample injection, number of repeat injections, analysis duration or analysis file No., pre-treatment file No-, fraction collector file No.

10. Column oven control

Parameters

Oven temperature, upper temperature limit

11. Detector control

Parameters

Detection wavelength, RANGE, time constant, lamp switching, scan conditions, etc.

12. Fraction collector control

Parameters

Fractionation conditions, etc., peak detection conditions, various analysis conditions

13. Data backup

All condition setting files (listed in 6. above) saved by battery backup if power failure occurs

14. Self-diagnosis and safety measures

Memory check

Upper and lower pressure limits, upper temperature limit, monitoring of lamp current, motor speed, etc.

15. Operating conditions, dimensions, weight

Power supply voltage

Nottage range

depends an Part No -

Part No. 228-34350-91: 100-120V AC, 50/60Hz, 320VA, 4A fuse

Part No. 228-34350-92: 100-120V AC, 50/60Hz, 320VA, 4A fuse

Part No. 228-34350-38: 220-240V AC, 50/60Hz, 320VA, 2A fuse

Ambient temperature

4-35°C

Outer dimensions

Width 260 x depth 420 x height 140 mm (excluding protruding parts)

Weight

6kg approx.

Part of labexchange logo
Search engine powered by ElasticSuite