Shimadzu AA 6701
| Objektnummer | B00008833 |
|---|---|
| Seriennummer | 008833 |
| Nome oggetto | Shimadzu AA 6701 |
| Stato | Archived Product |
Gruppo prodotti: Assorbimento Atomico
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
|
Country |
Possible payment methods |
Comment |
|
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
|
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
|
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
|
Status |
Condition |
Comment |
|
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
| In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
|
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
|
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
|
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
The AA-6701/6601 Atomic Absorption Spectrophotometer incorporates two background correction function, the D 2 method (Deuterium Lamp method) and the SR method (Self-Reversal method), enabling selection of the appropriate method for the measurement sample.
Listed below are the measurement modes which are supported.
(a) Flame Measurement (Continuous Sampling)
(b) Flame Measurement (Microsampling)
(c) Furnace Measurement
The AA-6701 allows you to change the measurement mode while automatically switching the atomizer, thus enabling quick and easy switching back and forth between flame measurement and fumace measurement.
In addition, wide ranging measurement operations are available, from manual operation to automatic continuous measurement of multiple elements with the use of an autosampler (ASC-6000). This enables selection of the appropriate combination to match the number of elements and samples to be analyzed as well as the skill of the operator.
Software operations are easily performed, thanks to the operating environment employing the MS-Windows GUI (Graphical User Interface). In addition, the MDI (Multi Document Interface) is adopted, so that parameter setting windows, calibration curve window, measurement results window and other necessary windows may be displayed on the screen simultaneously, to enable operations while maintaining a good perspective of the overall conditions.
Specifications
|
Spectrophotometer |
Wavelength range |
190 to 900 nm |
|||||
|
Mounting |
Aberration corrected Czerny-Turner mounting |
||||||
|
Bandwidth |
0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm (6-step automatic) |
||||||
|
Lamp |
Number of HC lamps |
8-lamp turret, 2 lamps simultaneously lit (1 for measurement and 1 warming up for next measurement) |
|||||
|
Background correction |
BGC-SR (high-speed self-reversal method) and BGC-D 2 (D2 lamp method;high-speed, dual-frequency simultaneous measurement) |
||||||
|
Lamp mode |
Emission, Non-BGC, BGC-SR, BGC-D 2 and D 2 |
||||||
|
HC lamp database |
Lamps in use may be registered for tracking their hours in use. |
||||||
|
Data processing |
Software requirements |
MS-Windows Ver 3.10 or later |
|||||
|
Operational parameter display |
MDI (Multiple Document Interface) method |
||||||
|
Window sizes |
Automatic adjustment by Quick View functions |
||||||
|
Measurement mode |
Flame continuous, flame microsampling, and furnace modes |
||||||
|
Concentration computation |
Calibration curve method (choice of 1st, 2nd, and 3rd order functions), Standard addition method (1st order function) |
||||||
|
Repeat analysis |
Up to 100 repetitions. Mean value, standard deviation (S.D.) and coefficient of variation (C.V) displayed. Automatic exclusion of deviant values by setting S.D. and C.V. |
||||||
|
Baseline correction |
Automatic drift correction using electrical double beam signal processing in both flame microsampling and furnace modes. |
||||||
|
Sensitivity correction |
Recalculation after drift correction |
||||||
|
Tabular data processing |
Calculation of final concentrations from sample volumes, dilution ratios, fill-up volumes in sample dissolution factors. |
||||||
|
Multitasking |
Spreadsheet software or word processing software may be executed during measurement. |
||||||
|
Recall of operational parameters |
Template functions available. |
||||||
|
Procedure/result display |
Concentrated MRT (Measured Result Table) system |
||||||
|
Report generation |
Quick report, Run time report, and Summary report |
||||||
|
Power requirements |
100, 120, 220, 240 VAC, switch selectable. 50/60 Hz. 600 VA. The computer and its peripherals require additional power. |
||||||
|
Dimensions and weight |
1080 W x 550 D x 460 H mm, 100 kg (including the gas controller and not including the computer and its peripherals) |
||||||
|
Burner Unit |
Type |
Air cooled premix type |
|||||
|
Burner head |
Stainless steel (titanium optional) |
||||||
|
Nebulizer |
Pt-Ir capillary with Teflon orifice |
||||||
|
Chamber |
Polypropylene |
||||||
|
Positioning |
Automatic search of optimum burner height (AA-6701/6701 F) Automatic flame-furnace switching by motor (AA-6701/6701 F) |
||||||
|
Gas Control |
Flow rate/pressure control |
Automatic control and automatic optimization |
|||||
|
Safety measures |
Automatic gas leak check, Air-C 2 H 2 flame priority ignition, Air-N 2 0 automatic changeover with flow rate monitor using optical sensor, Prevents flowing out of unburnt fuel, Prevents usage of wrong burner head, Flashback prevention with pressure monitor, Automatic flame extinction when power failure, even an instantaneous one, occurs., Drain tank monitor to prevent flashback |
||||||
|
Heating temperature range |
Ambient to 3000°C |
||||||
|
Heating control |
Drying: Current control with automatic temperature calibration Ashing/atomization: Temperature control with optical sensor |
||||||
|
Heating conditions |
Up to 20 stage, programming; RAMP and STEP, heating modes; choice of two, inner gases; high sensitivity mode available; up to 20 times, concentration boosting in the furnace; 0 to 1.5L/min. inner gas flow rate setting in 0.1 L/min. increments |
||||||
|
Cooling water/argon gas consumption |
Cooling water: 1.0 L/min. or higher, Argon: 3.0 L/min. maximum |
||||||
|
Power requirements |
200 VAC, 50/60 Hz, 6 kVA max. |
||||||
|
Dimensions and weight of power supply unit |
642W x 247D x 460H mm, 44 kg (to be placed in the rear space of the AA main; no additional space required. ) |
||||||
Guaranteed Sensitivity |
|||||||
|
|
Flame analysis |
Furnace analysis |
|||||
|
Abs |
0.45 Abs. for 4 ppm Cu |
0.18 Abs. for 20 µL of 1 ppb Mn |
|||||
Personal Computer Requirements |
|||||||
|
OS: DOS, MS-Windows ver 3.10, CPU: 80486DX-33 or higher, CRT: VGA or higher, RAM: 8 Mbyte or larger, Hard disk: 120 Mbyte or larger, 3.5" Floppy disk drive: One or more units |
|||||||