Descrizione:
Contatore di particelle Rion Modello KL 20K. Misura della concentrazione di particelle in acqua pura.
Quotation
siehe Storeview
Maggiori Informazioni
Objektnummer B00009263
Seriennummer 009263
Nome oggetto Rion KL20K
Stato Archived Product

Gruppo prodotti: Gelöscht

Non è stato trovato alcun prodotto corrispondente alla selezione.

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

Firma:              Rion

Modell:            KL20K

Kommentar:   Dokumente dt.

OUTLINE

The liquid-borne particle counter KL-20K is designed to measure the concentration of particles in pure water, using the light-scattering method. Particle count is displayed in five size ranges. Using peripheral equipment and the standard built-in RS-232-C or other optional interfaces, auto­matic recording of measurement results and transfer of data to other equipment is possible.

o      Features

• Measurement can be performed either off-line with a sampler.

• Particle count for five particle sizes (0.2, 0.22,0.25,0.3,0.5 µm and above) in a single measurement.

• Battery backup retains measurement settings also during power-off.

• Alarm function can give audible warning when number of particles exceeds a certain level (selectable in 4 steps). A relay for control of external devices is also provided.

Block Diagram

o      Peripheral Equipment Connections

This section shows a few representative connection examples.

For samples with low viscosity or high boiling point, suction type equipment (such as the syringe sampler KZ-30S) is suitable. For low-quantities of sample liquid and high-precision measurements, a syringe sampler should be used.

• Off-line measurement (suction measurernent)      

  • Interfaces

The RS-232-C interface KZ-20S is supplied as standard equipment with the KL-20K. Two other Interfaces are available as options. Either of these can be installed internally in place of the KZ­20S.

GP-IB (general-purpose interface bus) interface       KZ-22S

D/A converter interface           KZ-24S or KZ- 26S

  • Partlcle slze detection

Sizing of particles is based on the amount of light scattered by the particles. The threshold for particle size discrimination is calibrated at the factory, using particles of known diameter and refractive index (polystyrene latex spheres). Therefore, measured size is optically equivalent to the size of polystyrene latex particles.

SPECIFICATIONS

Measurement principle                               Light-scattering

Measurement object                                  pure water,

                                                                  liquids which do not penetrate the liquid contact          materials

Sample flow rate                                        10 ml/min

Sample liquid temperature                         10 to 30°C (no condensation of cell optical       assembly during measurement)

Sample liquid pressure range                    0 to 3 kg/cm 2

Liquid contact materials                             PTFE, CTFE, PFA, Quartz (glass cell)

Measurable particle size                            0.2 to 0.5 µm (for particles with refractive index           1.595 in pure water)

Particle size measurement range             Sizes over 0.2,.22,.25, 0.3, 0.5µm

Maximum particle concentration                1200 particles/ml (with 0.2-µm particle, at        coincidence loss of 5%)

Coincidence loss                                       Less than 5% (for 1200 particles/ml)

Sample liquid volume                                 Manual measurement:    Arbitrary (depending on          measurement time or syringe sam­pler)

                                                                  Automatic measurement: 10 ml, 100 ml Measurement time    Sample liquid volume setting                                            Measurement time

                                                                                 10 ml                                 Approx. 1 min

                                                                                100ml                               Approx. 10 min

Measurement modes                                 Manual, one-time automatic (HOLD),

                                                                  repeated automatic (REPEAT)

                                                                  HOLD:      Measurement value is retained until next      measurement is started.

                                                                  REPEAT: Measurement is restarted automatically      with 10-second pause interval.

Display                                                       Particle count:    Shows the total number of particles               with size selected by the PARTICLE                                                 SIZE selector or larger (range 0 to                  99999).

                                                                  REMOTE:          Lights up when the unit is remote                   controlled using the standard RS-                                                        232-C interface or op­tional GP-IB                   interface (KZ-22S). Flashes when                                                            the unit is set to hold by external                     control.

                                                                  LD NG:               Lights up when the output of the                     internal laser diode falls below the                                                            normal level.

                                                                  CELL NG:           Lights up when normal                                                                                          measurement is not possible due to                 contamination or condensation                                                                       in the detection cell.

                                                                  COUNT:             Lights up while a measurement is                  being performed.

                                                                  Over-range mark:

                                                                                            Lights up when the particle count                   has exceeded 5 digits.

Alarm                                                          Beeper and relay are activated when particle count     exceeds a preset level.

                                                                  Maximum relay capacity

                                                                  With resistive load:   30 V, 5 A (DC), 250 V, 5 A (AC)   With inductive load: 30 V, 3 A (DC), 250 V, 2 A (AC)

                                                                                                   (cos Ø = 0.4, L/R = 7 ms)

Alarm setting range                                    10, 100, 1000 or 10000 particles

Input/output terminals                                 CONTROL connector

                                                                  Output:       Particle size signals, model number,                     counter gate, clear, auto/manual mode,                                            10 ml/100 ml, LD status, cell status

                                                                  Input:           External start, external stop, external                     HOLD, ready/busy signals

                                                                  Interfaces

                                                                  RS-232-C:                         (standard)

                                                                  GP-IB:                   KZ-22S   (option)

                                                                  D/A converter:     KZ-24S   (option)

                                                                  (One of above selected for internal installation)

Ambient conditions for operation               5 to 35°C, less than 85% RH

Power requirements                                  100 V AC, 50/60 Hz, approx. 45 VA

Dimensions, Weight                                  10 (H) x 29 (W) x 40 (D) cm, Approx. 10 kg