Philips PU 9716/15
| Objektnummer | B00015465 |
|---|---|
| Seriennummer | 015465 |
| Nome oggetto | Philips PU 9716/15 |
| Stato | Archived Product |
Gruppo prodotti: Spettrometri Infrarosso
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
|
Country |
Possible payment methods |
Comment |
|
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
|
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
|
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
|
Status |
Condition |
Comment |
|
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
| In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
|
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
|
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
|
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
manufacturer: Philips
model: PU 9716/15
annotation: Dokumente engl.


Specifications:
Source lndirectly heated silicon carbide bar.
Optics Nominal Beam separation: l40mm
Beam centre height: 56mm
Sample 157mm wide x 327mm deep x l49mm high .
Compartment Focus 38 .5mm from left-hand wall.
Monochromator Orthogonal Ebert, focal length l80mm and aperture f/5.
One grating 50mm x 50mm, 100 lines/mm and l400cm -1 first order blaze. Grating used in first order from 2000 to 600cm -1 and in second order from 4000 to 2000cm -1 . Unwanted orders removed by four interference filters and one glass filter.
Detector Small area detector
Range (cm) 4000cm -1 to 600cm -1
Presentation Strip chart 200mm wide. Scale change 2 :1, at2000cm -1 .
Chart grid options:
A4-l50mm x 280mm chart grid
Standard-l50mm x 420mm chart grid
cm -1 Accuracy Better than ±4cm -1 , 4000 to 2000cm -1 .
Better than ±2cm -1 , 2000 to 600cm -1 .
cm -1 Repeat- ±0.4cm -1 for 4000 to 2000cm -1 .
ability ±0.2cm -1 for 2000 to 600cm -1 .
Resolution Spectral slitwidth on High resolution programme:
1.5cm -1 at 1000cm -1
3.0cm -1 at 3000cm -1
and on High energy programme:
3.5cm -1 at 1000cm -1
7.0cm -1 at 3000cm -1
Photometric As PU9706. `Spectraset`preset scan conditions facility
system
%T Linearity Better than ±0.2%T as measured on the 1V output
%T Repeat- Better than ±0.5%T as measured on
ability the 1V output
Output Sockets 0 - 1V - %T output + SP3-050 socket
Pen recorder Dead band less than 0.3% of F.S.D.
Agreement with 1V socket better than
±0.25%T
Transmittance Continuously variable digital calibrated expansion of
expansion %T up to x 100
Back off Back off ± 150% T
Beam balance ±1.5%T
Stray light Generally less than 0.3%
Scan times 1, 3, 7, 20 and 40 minutes and fast reset speed
Wavenumber x 1 and x 5
expansion
Overall 754mm wide x 627mm deep x 328mm high
dimensions
Weight 47kg
Services 110, 120, 200, 220 or 240V:200VA,
50 or 60Hz, ±10% permissible range
Purging Not standard
ACCESSORIES
The following accessories may be used with the PU9700 Series instruments:
Purge Kit
This kit provides sealing of the covers, cell compartment lid and a heat sink for the heat from the source thus allowing dry air purging of the instruments. lt should be factory fitted as a optional extra to the PU9706, PU9716, PU9714 and PU9712 instruments.
SP3-060 Air Dryer
This unit supplies 40 litres/minute (1. 55 cu.ft/minute) of dry air at below 70°C dewpoint. The services required are:
(a) Mains electrical supply of 115V or 230V (50 or 60Hz), 75VA.
(b) Supply of compressed air at pressures between 4.2 and 8.4 kg/cm 2 (60 and 120 lb/in 2 ).
The Air Dryer is supplied complete with filters, gauges and controllers. Full details are given in the Instruction Booklet which accompanies the accessory.
lnfra-red Data Station
The PU9700 can be driven by the IR Data Station via the ACCESSORY port on the accessory interface panel at the rear right-hand side of the main instrument (Fig 1 .2). Full details are given in the manual which accompanies the accessory.
PU9702 Absorbance Accessory
The PU9702 Absorbance Accessory is used with the PU9700 Series of IR spectrophotometers to convert the recorder signal from transmission to absorbance in the range OA to 2A. lt is connected to the ACCESSORY socket on the Accessory Interface Panel. Full information is contained in the PU9702 Users Manual 9499 230 10711 that accompanies the accessory.