Stato Archiv

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

manufacturer:       Perkin Elmer

model:                   FIMS 400

annotation:           Dokumente engl.

 

 

FIMS-400

 

The FIMS TM -400 is a compact, easy-to-operate atomic absorption spectrometer dedicated to the determination of mercury. Based on flow injection techniques, FIMS is fully automated, fast, and cost-effective. The FIMS uses a high-performance single-beam optical system with a low pressure mercury lamp and solar-blind detector for maximum performance. The FIMS-400 has two stepper motor-driven peristaltic pumps for greater flexibility when used with optional accessories.

 

 

Features/ Benefits

*                    High sample throughput

*                    Freedom from interferences

*                    Superior detection limits

*                    Large sampling capacity

*                    Lower operating costs

*                    Amalgamation compatible

*                    Compact design

*                    Powerful WinLab32™ for AA software

 

 


FIM S Flow Injection M ercury System

System Description                                                                                                                           

Automated atomic absorption spectrometer with integrated flow injection system for cold vapor mercury AA. Single beam optics with automatic baseline offset correction (BOC) immediately before each measurement, low pressure Hg lamp and solar-blind detector with maximum sensitivity at 254 nm. Built-in supply and control electronics based on surface mounted device (SMD) technology for fully automated operation.

The FIMS-100 system is provided with 1 pump. The FIMS-400 system is provided with 2 pumps.

System Control                                                                                                                                   

Full control of the spectrometer, Fl components, autosampler and other accessories via an industry standard personal computer using the PerkinElmer WinLab32 for AA system software based on the Microsoft ® Windows ® operating environment.

Transport of Reagents and Samples                                                                                              

Stepper-motor driven peristaltic pumps with a maximum of 8 channels each for tubing

with 0.13 to 3.18 mm inner diameter. The speed of revolution is selectable from 20 to 120 rpm.

Sampling and Flow Switching

Fl switching valve with 5 ports and exchangeable sample loops.

Gas Supply                                                                                                                                          

Carrier gas stream (argon) flow controlled from 40 to 250 mL/min by flow meter. Required inlet pressure: 320 - 400 kPa. Automatic switch-off after operating pauses of more than 10 min.

Absorption Cell                                                                                                                                  

Long-path glass absorption cell (240 mm long, 7 mm O.D., 4 mm I.D.) with removable quartz windows.

Cell Heating

Electrically heated mantle around the cell to maintain the cell at approx. 50 °C to prevent vapor condensation.

Data Connection                                                                                                                                 

Built-in IEEE-488 and RS-232C interfaces. Addressing and switch over between IEEE and RS-232C by DIP switches.

Remotes

9 output contacts, which can be activated independently; e.g. for control of the amalgamation accessory.

Power Requirements                                     

115 VAC or 230 VAC, 50/60 Hz, 450 VA.

Technical Standards                                                                                            

Classified as a laboratory instrument. Complies with the applicable European Union directives and standards for safety and electromagnetic compatibility for CE marking. The instrument has been developed and is produced in compliance with ISO 9001.

Weight

19 kg.

Dimensions (W x H x D)                                                                      

42 cm x 26 cm x 41 cm (including sample tray).

Amalgamation Accessory

Amalgamation Accessory

Principle

Amalgamation of mercury on a gold/platinum gauze and fast release of mercury by heating the gauze using an infrared heat source.

Heating Temperature                             

Approximately 600 °C.

Cooling

By means of a flow of compressed air.

Carrier Gas Flow

75 or 300 mL/min.

Operation

Times for collection, heating, cooling and flushing can be freely selected via the FIMS system.

Power Requirements                                                                

115 VAC or 230 VAC, 50/60 Hz, 300 VA.

Weight

4.9 kg.

Dimensions (WxHxD)

25 cm x 25 cm x 24 cm.

 

Gruppo prodotti: Assorbimento Atomico