Metrohm Titrino 785
Objektnummer | B00016767 |
---|---|
Seriennummer | 016767 |
Nome oggetto | Metrohm Titrino 785 |
Stato | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
manufacturer : Metrohm
model : Titrino 785
annotation : Dokumente engl.
The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
Front view of instrument:
1 Exchange Unit
2 Display
3 Control keys and indicator lamps on the Titrino
4 Setting of display contrast
5 Controls the dosing rate during manual dosing with <DOS> and subsequent filling
6 Opening for the memory card
Rear view of instrument:
7 RS232 i nterfaces
8 Remote lines (input/output)
9 Connection of electrodes and temperature sensor
10 Connection for separate keypad
11 Connection for power cable
12 Connection for stirrer
13 Rating plate
Technical specifications
Modes DET: Dynamic Equivalence Point Titration
MET: Monotonic Equivalence Point Titration SET: Set End point Titration
MEAS: Measurement
CAL: pH calibration
TIP: Links commands to titration procedure
Measuring inputs 2 high-impedance measuring inputs for pH, redox and ISE
electrodes .
1 reference input for a separate reference electrode. May also be used as a differential amplifier.
1 measuring input for polarized electrodes.
1 measuring input for temperature sensor Pt100 or Pt1000.
Measuring range
pH value (pX) 0...±20.00, resolution 0.01
Voltage 0...± 2000 mV, resolution 1 mV, error limit 0.1 % fullscale
Current 0...± 200 µA, resolution 1 µA
Temperature 150.0...+450.0 °C, resolution 0.1 °C
Polarizer Ipol 0...± 127 µA, resolution 1 µA
Upol
0...± 1270 mV, in steps of 10 mV
Measuring amplifier
(at 25 °C and Titrino warmed-up)
Input resistance >10 13 Ω
Offset current <3 10 13 A
Deviation of offset voltage 15 µV/K
Dosification
Volume of buret cylinder 1, (2), 5, 10, 20 or 50 ml
Resolution 10 000 steps per buret cylinder
Auxiliary burets 2 additional burets: 776 or 765 Dosimat
Materials
Housing Polybutyleneterephthalate (PBTP)
Keypad cover Polycarbonate (PC)
Display Graphical LCD, 192 x 64 Dots
Field: 100 x 37 mm
LED back-lit
Memory card
Standard JEIDA ICMC 4.0 / PCMCIA (68 pins)
Data Method memory
Storage capacity 128 KB (256 KB maximum)
SRAM card Write and read, battery supplied
Flash Read only with 55 Metrohm methods
Internal memory Method memory for up to 100 methods
Silo memory for sample data and results
RS232 interface 2 separate interfaces, each can be configured
for printer, balance or computer connection: completely controllable from external control unit
Remote input/output lines
for Sample Changer, robot connection.
With optional Remote Box:
Connection of barcode reader and PC keyboard.
Stirrer control Switch the stirrer on/off either manually or coordinated with
the titration sequence
Ambient temperature
Nom. operation range 5 ... 40 QC
Storage 20 ... 60 QC
Transport 40 ... 60 QC
Safety specifications
Designed and tested in accordance to IEC publication 1010, safety class I. This manual contains information and warnings which have to be followed by the user to ensure safe operation and to retain the apparatus in safe condition.
Mains connection
Voltage 100...240 V
Frequency 50 ... 60 Hz
Power consumption 15 W
Fuse 2 x 1 ATH (to be replaced by Metrohm Service only usin the same type)
Additional electronic overload protection
Dimensions with Exchange Unit
Width 150 mm
Height 450 mm
Depth 275 mm
Weight , incl. keypad app. 3.6 kg