Metrohm 726
Objektnummer | B00018445 |
---|---|
Seriennummer | 018445 |
Nome oggetto | Metrohm 726 |
Stato | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
Firma: Metrohm
Nachfolgende Abbildungen und Beschreibungen sind modellbezogen und aus Prospekten entnommen.
Sie geben nicht den Lieferumfang des Systems wieder.
Den exakten Lieferumfang entnehmen Sie bitte aus dem Angebotstext.
Technische Daten
Der Titroprocessor 726 ist ein kompaktes Titrier- und Messgerät für die Anaiytik in Labor und Betrieb. Anschlussmöglichkeiten für 1 oder 2 Probenwechsler, 4 potentiometrische Sensoren ( pH / mV), 2 polarisierbare Elektroden, 2 Temperatursensoren usw.
32-Bit Mikroprocessor , 16 MHz Taktfrequenz, 2 MByte batteriegestütztes SRAM, 2 MByte EPROM.
Echtes Multitasking mit dynamischer Speicherverwaltung.
Abmessungen
Höhe: 25 mm (mit eingebautem Drucker)
175 mm
Breite: 405 mm
Tiefe: 500 mm
Gewicht
12,6 kg (mit eingebautem Drucker)
11,1 kg (ohne Drucker)
Material
Gehäuse: PUR (Polyurethan)-Hartschaum mit Flammschutz für Brandklasse UL94VO, FCKW-frei
Netzversorgung
Netzspannung: 100...120 V ± 10 %, 220...240 V ± 10 %
Netzfrequenz: 50...60 Hz
Energieverbrauch: max. 160 VA
Sicherung: 2 ATH (träge, mit grossem Schaltvermögen)
5 mm Ø, 20 rnm lang
Bildschirm
Typ: LCD-Bildschirm, hinterleuchtet
Bildschirmdiagonale: 257 mm
Auflösung: 640 x 400 Punkte, 4 Graustufen;
28 Zeilen zu je 80 Zeichen
Wiederholfrequenz: 72 Hz
Kontrast: einstellbar
Tastatur
Schaltpillentastatur mit Druckpunkt, spritzwasserdicht
Kommunikationsschnittstellen
2 x RS232 für serielle Datenkommunikation mit Waage, PC oder Drucker
Norm: EIA RS232C, CCITT V.24, ISO 2110, DIN 66020
Buchse: D-Sub 25-polig
max. Baudrate: 9600
Remote-Schnittstelle für die Steuerung externer Geräte
Typ: Logik Steuerleitungen
Leitungen: 8 Output, Opencollector , V CE = 40 V, I C = 20 mA
8 Input, Pullup auf 5 V
1 Speisung, 4...5.5 V, 200 mA
1 GND
Drucker (Variante)
Typ: Thermodrucker
Papier: Endlos-Thermopapier A4, perforiert
Auflösung: 84 dpi im Textmodus
160 dpi im Grafikmodus
Anschlüsse Barcode-Leser-Anschluss für Dateneingabe
Buchse: D-Sub Buchse, 9-polig nach RS232C
Übertragungs-
einstellungen : Baudrate 9600
Datenbit 8
Stoppbit 1
Handshake RTS/CTS
Zeilenende: CR/LF
Eingangspuffer: 40 Zeichen
Zeichensatz: 7-Bit ASCII
Pinzuordnung: Pin 1 ernpty Pin 6 empty
Pin 2 TxD (Output) Pin 7 RTS (Output)
Pin 3 RxD (Input) Pin 8 CTS (Input)
Pin 4 empty Pin 9 +5 V
Pin 5 GND (0 V)
Verdrahtung: 726-Buchse Barcode-Leser
TxD → RxD
RxD ← TxD
RTS → CTS
CTS ← RTS
GND ─ GND
+5 V → +5 V
4 x Metrohm Dosimat 685 oder Dosino 700
Buchse: D-Sub Buchse, 9-polig
1 oder 2 Rühreranschlüsse für Metrohm Rührer 728 oder ähnlich Ausgangsspannung: 10,8...13,2 V
Strom: 2 50 mA
kurzschlusssicher und schaltbar
Metrohm External Bus Anschluss für:
2 x Probenwechsler 717
2 x Dosimat-Interface 729 für je 4 x Dosino 700 oder
Dosimaten 685
Druckeranschluss (Variante), Parallelport für Standarddrucker Buchse: D-Sub Buchse, 25-polig
Messteil
1 oder 2 (Variante) Messgruppen mit je
2 hochohmige Messeingänge für pH- , Redox- , ISE-Sensoren
1 Referenzeingang für separate Referenzelektrode
(obige Anordnung kann als Differenzverstärker eingesetzt werden)
1 Messeingang für polarisierte Elektroden
1 Messeingang für Temperatursensoren Pt 1000 oder Pt 100, mit automatischer Erkennung
1 Polarizer
2 parallele A/D-Wandler, 20 Bit, 50 ms Messrate
1 Rühreranschluss (siehe oben)
Galvanische Trennung
Alle Signale eines Interface sind auf ein gemeinsames Referenzpotential bezogen. Der Referenzpunkt ist gegenüber der restlichen Schaltung galvanisch getrennt. Jedes Interface ist einzeln getrennt.
Die rein messtechnisch bedingte Trennung erfüllt nicht die Anforderungen einer Sicherheitstrennung,
Messbereiche
potentiometrisch: -2500 mV...2500 mV
pH-20,000...20,000
amperometrisch -250µA...250µA ( Upol )
voltametrisch -2500 rnV ...2500 mV ( Ipol )
Temperatur: -150 °C...450 °C
Messfehlergrenzen
Eingänge abs. Fehler Auflösung
Ind A1, A2, B1, B2 < ± 0 ,8 mV 5 µV
Diff A, Diff B < ± 4 ,0 mV 5 µV
Pol (µA, Upol ) < ± 3 µA 0,001µA
Pol (mV, Ipol ) < ± 20 mV 5µV
Pt100/1000
0°C 100°C < ± 0,3°C 0,005°C
Referenzbedingungen
Die Messfehlergrenzen gelten ab Werk bei einer Umgebungstemperatur von 25°C (±- 5°C) und nach 5 Min. Einschaltdauer.
Die Messfehlergrenzen der Temperaturmessung gelten inklusive der Überbergangswiederstände der Buchse und inklusive einem Sensorkabel des Metrohm Pt100 Temperaturfühler NS, aber exklusive die Fehlergrenzen des Sensors selbst.
Quellengenauigkeit Polarizer
Polarisationsspannung ( Upol )
Messbereich: -1270 mV...1270 mV
Linearität: < ± 20,0 mV
Auflösung: 10,0 mV
Polarisationsstrom ( Ipol )
Messbereich: -127 µA...127 µA
Linearität: < ± 3,0 µA
Auflösung: 1,0 µA
Umgebungs-Temperatur
Nomineller Funktionsbereich 5 40 °C
Lagerung, Transport -25...60 °C
Umgebungs- Luttfeuchtigkeit
bei < 40 °C max. 95% relative Luftfeuchtigkeit
bei 45°C max. 70% relative Luftfeuchtigkeit
bei 60°C max. 35% relative Luftfeuchtigkeit
Titrier- und Messmodi
DET dynamische Endpunktitration ; U, pH , I pol , U pol
MET monotone Endpunkttitration ; U, pH , I pol , U pol
SET Titration auf vorgegebenen Endpunkt; U, pH , I pol , U pol
SEC Konditionieren auf vorgegebenen Endpunkt; U, pH , I pol , U pol
KFT Karl-Fischer-Endpunkttitration ; I pol , U pol
KFC Konditionierung nach Karl Fischer; I pol , U pol
MEAS Messung ; U, pH, I pol , U pol T
Speicher
Programmspeicher: EPROM, max. 4 MB
Datenspeicher: 2 MByte SRAM, batteriegestützt
Bestimmungsspeicher: für bis zu 20 Bestimmungsdatensätze,
mit jeweils bis zu 5 Titrationen oder Messungen
Methodenspeicher: für frei definierbaren Ablauf, bis zu 99 Befehlszeilen
Silospeicher: Probendatenspeicher für bis zu 160 Probendatensätze
Dateispeicher: freier RAM-Bereich zur Ablage von Bestimmungen, Methoden,
Probensilos
Datenkarte
Typ: Momory Card nach Standard JEIDA 4.X / PCMCIA
2.X (68 Pins)
Speicherkapazität je nach Modell, 128 KByte; andere
Versionen bis 2 MByte
Sicherheits- Spezifikation
Konstruktion und Prüfung gemäss IEC 1010 / EN 61010 / UL 3101-1,
Schutzklasse I
Schutzgrad IP 52
Die Gebrauchsanweisung enthält Informationen und Warnungen, die vom Benutzer befolgt werden müssen, um den sicheren Betrieb des Gerätes zu gewährleisten.
Elektromagnetische Verträglichkeit (EMV)
Störaussendung: Das vorliegende Gerät erfüllt die Fachgrundnormen EN 50081-1 01.92, EN 55011 (Klasse B), EN 55022 (Klasse B), und NAMUR.
Störfestigkeit: Die Normen Erfüllt EN 50082-2; 1995, IEC801-2 bis IEC801-6, EN 60555-2 und NAMUR werden eingehalten.