Labsystems Multiscan plus
Objektnummer | B00013111 |
---|---|
Seriennummer | 013111 |
Nome oggetto | Labsystems Multiscan plus |
Stato | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.
INTRODUCTION
Multiskan PLUS is an eight-channel vertical light path filter photometer designed to measure the absorbances of liquids in flat-bottomed, 8x12 well microplates, and flat-bottomed Microstrips® of 1x8, 3x8 or 1x12 wells.
The wavelength (400-750 nm) is selected using high quality interference filters.
Four measurement modes and eight calculation modes are included. Any of the calculation modes can be applied to the results of any of the measurement modes.
Multiskan PLUS is designed to be used in various colorimetric and turbidimetric measurements typical of enzyme immunoassay (EIA) and microbiological work.
Key advantages of Multiskan PLUS are its fast reading Speed and versatile and easy-to-use in-built software.
PRINCIPLE OF OPERATION
The optics of Multiskan PLUS are based an Labsystems" concept of vertical photometry in which the light beam passes through the whole sample.
In vertical photometry, the absorption of light is proportional to the amount of light absorbing substance in the well.
Absorption is expressed by the following equation:
A = a/S m,
where A = absorbance, a = molar absorptivity of the substance, m = mass of absorbing substance and S = cross-sectional area perpendicular to the light path.
The advantages of the vertical light path measurement are:
(1) inaccurate pipetting of nonabsorbing liquids does not affect the measured absorbance values
(2) evaporation of nonabsorbing liquids during the reaction does not affect the measured absorbance values and
(3) a certain degree of inhomogeneity in the solution, for example as a result of layering in turbidity measurements, does not affect the results.
SPECIFICATIONS
Weight: 11 kg
Overall dimensions: 140 mm x 420 mm x 320 mm
Mains power supply: 115 V, 50/60 Hz, nominal (operating range 90...130 V)
230 V, 50/60 Hz, nominal (operating range 180...260 V)
Power: 1.3 A (115 V)
0.7 A (230 V)
Fuses: 2 x 3.5 A, slow-blow type
Operating temperature range: +10°C to +40°C (tested according to IEC 68-2-1, test Ab,cold; IEC 68-2-2, test Bb, dry heat and IEC 68-2-3, test Ca, RH 80%, damp heat, steady state)
Storage temperature range: -25°C to +50°C
Transportation temperature -40°C to +70°C (tested according to IEC 68-2-1,
range: test Ab, cold test; IEC 68-2-2, test Bb, dry heat)
Performance range:
spectral range 400 - 750 nm
absorbance range 0 - 2.000 absorbance units
Speed of operation:
warm-up time: 1 minute
reading speed: 5 s/plate (continuous mode)
Microplates: The manufacturer recommends the use of flatbottomed microplates.
(Specifications are set for flat-bottomed microplates.)
Resolution: 0.001 A (absorbance units)
Accuracy: ±2.0% or ±0.007 A, whichever is greater (0-2.000 A in the step mode) (0-1.500 A in the continuous mode)
Precision: CV = 0.5% or 0.007 A, whichever is greater (0-2.000 A in the step mode) (0-1.500 A in the continuous mode)
Linearity: ±2.5% or +0.007 A, whichever is greater (0-2.000 A'in the step mode) (0-1.500 A in the continuous mode)
Long term Stability: Because of the autoblanking procedure the Instrument is stable.
Display: 16 character alphanumeric LC display
Output connections: external computer or
external printer, results in plate format
Interface: RS-232C serial interface
baud rate: 150, 300, 600, 1200, 2400 4800, 9600, 19200
character format: 1 start bit
8 data bits
1 stop bit
no parity
Optical system: quartz tungsten halogen lamp (OSRAM 64607, 8V/50W), aperture, chopper wheel, condenser ens, semi-transparent mirror, interference filter, fibre bundle, focusing lenses, upper lenses and detectors
Interference filters: high quality interference filters with minimum aperture of 10 mm. 405 ± 2 nm, 450 ± 2 nm and 492 ± 2 nm supplied as standard, half bandwidth 3-9 nm. Other filters available, to order, in the range 400 to 750 nm.
Operating modes: Measurement modes: absorbance, dual wavelength absorbance, two point, kinetic
Calculation modes: factor, linear standard, standard line, limit, double limit, range, column subtraction, point to point
Computer control mode
Keypad: 27 keys with click action switches