Hewlett Packard 5890 II
Objektnummer | B00016458 |
---|---|
Seriennummer | 016458 |
Nome oggetto | Hewlett Packard 5890 II |
Stato | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
Firma: Hewlett Packard
The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.
HP 5890 Series II and HP 5890 Series II Plus Gaschromatographs
(Note: The HP 5890 II Plus is a version of the HP Series II configured specifically for the use of EPC).
Technical Data
Instrument Dimensions and Weight
Height: 18-3/8 inches (465 mm)
Width: 25-7/8 inches (655 mm
Depth: 20-1/8 inches (511 mm)
Weight: 90 pounds (41 kg)
Power Requirements:
Voltages: 120/200/220/240
Range: +5, -10% each
Frequency: 47.5-66 Hz
Consumption: 2200 VA max
Output: 7500 Btu/hr max
Environmental
Operating range is: 0-55°C ambient (20-27°C optimum)
5-95% humidity (50-60% optimum)
Heated Devices
Five heated zones standard: Two detectors
Two inlets
One auxiliary
Methods stored: Two
Flow Sensor (not available with EPC or Series II Plus)
Range.- 0-35 sccm [sccm = standard cc/minute]
Accuracy: ± 3 sccm für He and H 2
± 5 sccm for N 2 and ArCH 4
Extended range: 100-150 sccm
Accuracy: ±15 sccm
Detector Signal
For external processing by a recorder, integrator, or computer:
Signal Path |
Signal Bandwidth |
Minimum Peak Width* |
INET digital |
4 Hz |
0.32 sec |
0-1 mV analog |
~ 4 Hz** |
** |
0-1.1 V analog |
2.6 Hz |
0.50 sec |
* That can be calculated accuratey by an external measuring device (e.g., integrator) operating at ≥ 4-Hz bandwidth frequency.
**Actual bandwidth depends on the input impedance of the measuing device.
Standard Inlets on the HP 5890 Series II
Split/Splitless Capillary
Paked witb Septum Purge
Temperature-Programmable Cool On-Column Capillary
Split/Splitless Capillary
Range to 400°C in 1°C increments
Back-pressure design permits independent adjustment of split flow rate without affecting column flow, 030 psi head pressure gauge and regulator standard; 015, 060, and 0100 psi regulators and gauges also available
Septum purge built in at 3 ml/min
Accepts columns up to 1.2-mm od
Accepts 114-inch glass column for on-column injection
Multimode design includes split and splitless injection
Air fan built into mainframe to assist in cool-down of inlet
Splitless purge time variable in 0.01-min elements
Packed with Septum Purge
Range to 400°C in 1°C increments
Flow control/forward-pressure design; 0100 psi head pressure gauge standard; 010 and 060 psi gauges also available
Septum purge built in at 1.5 ml/min
On-column injection available with configuration A, 1/4-inch od glass columns
Individual liners for use with 114-inch metal columns as well as for Series 530-µ columns
Liners available for use with replaceable glass inserts
Air fan built into mainframe to assist cool-down of inlet
Tempera ture-Program mahle Cool On-Column Capillary
Temperature-programmable to 450°C with up to three ramps at up to 100°C/min; also available with oven track mode
Forward-pressure design, 030 psi head pressure regulator and gauge standard; 010, 060, and 0100 psi regulators and gauges also available
Septum purge built in at 14 mI/min.
Accepts columns with 0.32- to 1.2-mm od (min id = 0.2 mm)
Automated injection onto 250-p, 320-1p, and 530-p capillary columns with HP 7673A
Air fan built into mainframe to assist cool-down of inlet; optional cryogenic cooling
Detectors
Thermal Conductivity Detector
Electron Capture Detector
Flame Ionization Detector
Nitrogen Phosphorus Detector
Flame Photometric Detector
Thermal Conductivity Deteotor Range to 400°C
Single-filament (single caluran) design has fluidic switching of reference and analytical carrier flows; a passivated tungstenrhenium filament is used in a 3,5-µl pell, operating at a constant temperature dif erence relative to the deteotor block temperature
Minimum detectable: <400 pg/ml carrierequivalent to <1 ppm neon in 1 ml air
Linear dynamie range: < ± 5% over 10 5 range
Digital gain setting time-programmable through the keyboard
Conditions: detector 100°C, 45 ml/min switching and 30 ml/min analytical flow of helium, propane Sample
Electron Capture Detector
Range to 400°C
A coaxial design with a 15 mCi source (555 MBq) of Ni-63 plated an the inferior of the lower block
Constant current mode of operation features switoh selection of pulse parameters for using either nitrogen, hydrogen, helium, or argon/methane oarrier gas
Minimum detectable level: <0.04 pg/sec lindane
Dynamic range: >10 4 for lindane
Conditions: detector 250°C, 60 ml/min nitrogen carrier
Flame Ionization Detector
Range to 450°C
Grounded jet and current limited design for operator safety
Push-button Flame ignition Fused silica columns insert within 2 mm of the jet tip
Sensitivity:
>18 mCoul/gm carbon: nitrogen oarrier, 0.018-in. id jet
>15 mCoul/gm carbon: helium carrier, 0.018-in. id jet
>22 mCoul/gm carbon: nitrogen carrier, 0.011-in. id capillary jet
>18 mCoul/gm carbon: helium carrier, 0.011-in. id capillary jet
Minimum detectable: <5 pg carbon/sec, nitrogen carrier at S/N=2
Linear dynamic range: < t 10% over a 10 7 range with 0.018-in, id jet
Conditions: column flow 50 ml/min, 45 ml/min Fit, 650 ml/min air, propane sample
Nitrogen Phosphorur Deteetor
Range to 400°C
Minimum detectable:
<0.4 pg Nilsec as azobenzene
<0.2 pg P/sec as malathion
Dynamic range: 10 4 (N and P)
Selectivity: 35,000 to 1: g N/g C (as octadecane)
75,000 to 1: g P/g C
Conditions: 30 ml/min column flow, 3 m1/min H 2 , 80 mi/min air
Flame Photometric Detector
Range to 300°C
100530 µ id capillary columns fit through the transfer line to the base of the flame
Two combustion modes: air/hydrogen or oxygen/hydrogen
Detector signal display to view when flame is lit
Push-button flame ignition
Sulfur mode: Sensitivity to 20 pg S/second Range
10 3 Selectivity
10 5 g S/g C
Phosphorus mode: Sensitivity 0,9 pg P/second
Range 10 4
Selectivity 10 6 g P/g C
Column Oven
Usable volume: 11 in. x 12 in. x 6.5 in. (h x w x d)
279 mm x 305 mm x 165 mm (h x w x d)
Column span: 228.5 mm, 9 in. (coil size)
Automatic cooling under processor control
Operatmg range: 4°C above ambient to 450°C:
- 80 to 450°C with cryogenic cooling
Setpoint entry: 1°C for temperatures
0.1°C for programm rates
Programming: Rates 0.1 to 70°C per min
650 min maximum run time
Three ramps with initial/final holds
*Achievable rates depend on zone temperature, voltage and columns.
Technical Performance
Accuracy (true temperature relative to setpoint)
Specification: ±1% (°K) from 4°C above ambient to 450°C
Stability (effect of ambient change on actual temperature)
Specification: <0.01°C for 1°C ambient change
Calibration (setting true temperature at a setpoint)
Oven can be recalibrated to ± 0.01°C with appropriate instrumentation
Gradients (temperature variations within a column)
Specification: less than 2°C within a 9-in. coil anywhere within the operating range