Dade DAC II
Objektnummer | B00013292 |
---|---|
Seriennummer | 013292 |
Nome oggetto | Dade DAC II |
Stato | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.
1-1: Dade Automatic Centrifuge II
Functional Description
The Dade® Automatic Centrifuge II is a laboratory bench-top, multi-purpose, electromechanical instrument designed to perform cell washing and immunohematology procedures including cell resuspension. lt provides the laboratory with a fast wash cycle capability, thereby lowering laboratory time costs. lt can be easily programmed to automatically perform one, two, three, four or five wash cycles per test sequence. Each cycle consists of programmed saline introduction and subsequent decanting during centrifugal rotation. The System provides a saline monitoring feature to warn the Operator when saline is not being delivered. The centrifuge incorporates variable spin times for several different types of centrifuge operation. Decant RPM is Operator adjustable to give more than 0.1 mL residuals if desired.
An agitation cycle resuspends cells for approximately 5 seconds and then spins the cells for 20 seconds before completion of the cycle. The agitate time, as well as spin after agitate time, is operator-programmable. The front control panel (See Figure 1-2) contains a vacuum fluorescent display of RPM functions and time which allows the operator to continually monitor the status of instrument operation.
Physical Description
The centrifuge is a self-contained Instrument 14.5 inches (36.8 cm) high, by 12.75 inches (32.4 cm) wide, by 16 inches (40.7 cm) deep, and weighs approximately 43 pounds (20 kg). lt is mounted on a heavy metal base with two adjustable front feet and two stationary rear feet. A power cord with a threepronged plug provides primary power to the unit.
The microprocessor-controlled electronics portion of the instrument consists of the Control and Display Printed Circuit (PC) Board. This PC Board coordinates the Operation of the instrument control panel and keyboard, electronic detectors, and control elements throughout the instrument. The board also visually displays the motor RPM, wash count or function and time through the three display view windows located on the front control panel. An "end of cycle" is signaled by an audible signal.
The four basic assemblies or components that comprise the mechanical portion of the instrument (see Figure 1-3) are as follows:
A. Centrifuge inner bowl - The centrifuge inner bowl houses the centrifuge rotor with 12-tube carrier. Access to these components is gained by opening the centrifuge top cover. The inner bowl is easily removed for cleaning. Warning: After replacing the removable bowl, and before installing the rotor, rotate the rotor drive by hand until it turns freely. The saline dispensing adapter, located on the top and inside of the centrifuge cover, is connected to the saline pump by the saline output tubing (see Figure 1-3). The saline drain outlet is located at the rear of the centrifuge bowl and extends out of the rear of the instrument. An overflow drain is located just above the normal drain fitting.
B. Centrifuge rotor - The centrifuge rotor has a capacity of 12 test tubes of either 10 x 75 mm or 12 x 75 mm size. All loadbearing components are manufactured from high grade, corrosion-resistant stainless steel. The rotor may be removed from the instrument by grasping the upper collar of the rotor and lifting. The centrifuge rotor can be placed either on a bench top or in a water bath, as the bottoms of the tube carriers act as supports.
C. Test tubes - Either 10 x 75 mm or 12 x 75 mm glass test tubes can be used in the rotor. Use of a 10 mm adapter (P/N S23265-001) in each tube carrier is required for 10 x 75 mm test tubes.
D. Control panel - The control panel (also referred to as the operator's panel or front panel) is equipped with a vacuum fluorescent alphanumeric readout display of RPM functions and time which allows the Operator to continually monitor the status of instrument Operation. All panel switches are the tactile type, activated by applying finger pressure to the embossed panel surface. A "rocker" power switch is mounted through the front panel.
1-3: Dade Automatic Centrifuge II Wash Enclosure, Bowl and Rotor
Performance Characteristics
The Dade® Automatic Centrifuge II encompasses the following performance characteristics:
The instrument can be programmed to perform one, two,
three, four, or five wash cycles per test sequence.
Spin function can be selected for a preset spin time of 20
seconds.
Manual Operation of the instrument allows for variable spin
times up to a maximum of 9 minutes 59 seconds.
An agitation function is provided for resuspension, then
centrifugation of the resuspended mix.
Front panel keys are tactile type.
RPM, Function and Time readout displays permit constant monitoring of instrument Operation.
An audible signal accompanies the end of all test programs.
The centrifuge Iid interlock prevents instrument Operation if the centrifuge cover is open.
Decant RPM is operator-adjustable for a dry (790 RPM) or wet (750 RPM) cell button.
The centrifuge lid lock prevents operator from opening cover during instrument operation.
The instrument is automatically grounded when connected to properly grounded power source.
Saline volume is microprocessor-controlled.
All programmed functions last selected and entered via the ENT key will be retained even after power is turned off.
Specifications
The specifications listed below are subject to change without notice to allow the introduction of design improvements.
Automatic controls
Rotating
Radius
Speed approx. Time
Auto * Wash 0-3200 RPM N/A 4 sec.
Wash * Fill 3000-3200 RPM 9.2 cm (925-1050 rcf) 16 sec.
Sequence * Pack 3000-3200 RPM 9.2 cm (925-1050 rcf) 14 sec.
* Pause 3200-0 RPM N/A 10 sec.
* Decant 750-790 RPM 4.8 cm (19-34 rcf) 9 sec.
Auto * Spin 3000-3200 RPM 9.2 cm (925-1050 rcf)
Spin
Auto * Agitate N/A (5 second agitation
Agitate & 20 second spin) 25 sec.
Manual Controls
Rotating
Radius
Speed approx. Time
* Spin 3000-3200 RPM 925-1050 rcf Programmable
(1 second - 9 minutes
59 seconds)
* Agitate N/A - Programmable
(1-9 seconds)
* Spin after 3000-3200 RPM - Programmable
agitate (1 second-2 minutes
55 seconds)
Electrical
Catalog No. B5054-2X
* Volts 220-240 VAC
* Amperage (max.) 1.5A
* Cycles 50 Hz
Physical
* Height 14.5 in. (36.8 cm)
* Width 12.75 in. (32.4 cm)
* Depth 16.0 in. (40.7 cm)
* Weight 43.0 Ibs. (20.0 kg) approx.
* Shipping weight 58 Ibs. (26.3 kg) approx.