BIAcore sistema di campionamento. Per la purificazione di biomolecole. 100-120/220-240 V 50/60 Hz marchio CE. Dispositivo con schermo e drive incorporato. Interfacce: RS232 e stampante. Software: Software di valutazione Biacore V.2.0. Biacore controllo software v.1.0a campione. Sistema di dati specifici. Tutto il software è fino a 3,5 \\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\\ ". Attrezzatura floppy ha successivamente fatto 2000fähig.
siehe Storeview
Maggiori Informazioni
Objektnummer B00013933
Seriennummer 013933
Nome oggetto Biacore Probe System
Stato Archiv

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

Firma:                   Biacore

Modell:                Probe System

Kommentar:        Dokumente engl.

The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.

BIACORE P ® P probe , for fast identification of biological activity from fermentationlexpression to purified biomolecule.

Faster Purification Development for Biomolecules

Ø        Easy to use

Ø        Flexible in application

Ø        Robust in performance

Ø        Direct identification of biological activity during Isolation and purification

            -no sample preparation, labet free

Ø        Fast decisions

            -immediate result evaluation

            -instant savelwaste decisions

Ø        Estimation of yield at every stage

Ø        Fast Information an activity changes during modification, extraction or purification

            -follow activity throughout procedures

Ø      Faster than conventional analytical techniques

            -Elisa, PAGE, HPLC, etc

BIACORE® probe provides direct identification of biological activity at any stage in a purifica­tion process, from fermentation/expression to final purified biomolecule. Identification of the target molecule achieved within minutes. The technique is labet free; no sample preparation is necessary.

A Sensor Probe is coupled covalently wich a biological binding partner or affinity ligand - the detecting molecule. This functionalized probe is exposed to sample, the sample exchanged for reference buffer, and the bound analyte is measured. The probe is re-usable.

BIACORE@probe is designed for rapid routine determination of the presence and active concentration of biomolecules in solutions, at any stage from expression and extraction to final purification. Measurement is based on the ability of the biomolecule in question (the analyte) to bind to a Chosen interactant (the detecting molecule). No label is required. Operation is simple and each measurement takes typically less than 5 minutes, making BIACORE®probe well suited to method development and other application areas where the design of the next step in a process is directly dependent on the results of the previous step. Approximately 100 pl of sample is required, and the measurement is non-invasive so that sample can normally be used for further processing.

The detection principle exploits the optical phenomenon of surface plasmon resonance (SPR), which measures the surface concentration of solute at a gold-liquid Interface (see Appendix B for a technical description). In BIACORE®probe, SPR occurs at the gold-coated tip of an optical fibre (the sensor probe) coated with detecting molecule. When analyte binds to the detecting molecule, an SPR response is measured which is a direct measure of the amount of functionally active analyte in the sample. ("Functional activity" is defined in terms of the ability of the analyte to bind to the detecting molecule – BIACOREGprobe will not detect analyte molecules which haue lost this ability.) Typical detecting molecules are high affinity antibodies for specific antigens, although other molecules may be appropriate in specific Gases.

Once the detecting molecule has been attached to the sensor probe, determining the presence or concentration of analyte with BIACORE@probe involves four simple steps (Figure 1-1):

1. Set a zero response with the sensor probe in a suitable reference buffer.

2. Bring the sample into contact with the sensor probe incubate for 30 seconds – 2 minutes.

3. Return the sensor probe to reference buffer and read the response (which is a measure of

the amount of analyte bound to the sensor probe).

4. Incubate the sensor probe with regenerating solution for 30 seconds – 1 minute to remove

     bound analyte in preparation for the next measurement.

The detection principle used in BIACORE probe offers major advantages over conventional methods for biospecific determination of molecules such as ELISA:

•       BIACORE@probe uses a label-free technology: neither the analyte nor the detecting molecule needs to be labelled, saving time, effort and money. In addition, all steps in an assay are revealed with the same detection principle, providing valuable intra-assay controls (e.g. attachment of the detecting molecule to the sensor probe is monitored in the same way as binding of analyte).

•       BIACORE®probe uses a non-invasive technology: since the light used to generate the response does not penetrate the sample, measurements can be made on turbid or opaque samples (e.g. cell culture supernatants, fermentation broth, Grude cell extracts).

•       BIACORE®probe uses an interactive technology: each measurement takes typically less than 5 minutes, so that decisions can be made on the basis of each measurement (e.g. chromatogram fractions can be rapidly scanned to localize the active peak for immediate follow-up processing).

BIACORE®probe consists of the following main components (Figure 2-1):

•            the detection unit wich Couch screen display

•            the sensor probe

•            the optocable connecting the sensor probe to the detection unit

•            the pipette which serves to bring the Sample to the sensor probe

The detection unit

The main body of the detection unit houses the light source (a low temperature halogen lamp), detector and software for registering and interpreting the Signal from the Sensor probe, and a 3.5 " diskette drive for copying data to PC-formatted computer diskettes (See Section 4.7). A green lamp on the front panel indicates that the power is on.

The rear panel of the detection unit has the mains power switch and connectors for mains power and a printer (25-pin computer connector). There is also a 9-pin computer connector which is used for Service purposes.

The touch screen provides the Interface for entering instructions and data to the Instrument, as well as displaying the results of measurements. Buttons are displayed on the screen for the functions available at any given time. To issue an instruction, touch the screen lightly on the appropriate display button.

The screen can be turned on its supporting column and tilted from the vertical for adjustment to Standing or sitting working positions (Figure 2-4). The red buttons on the top and side of the screen column release the screen for turning and tilting adjustment respectively. The screen

locks automatically in the Chosen position. Take care not to trag your hand between the screen and the supporting column when you adjust the screen angle.

Pipette holder

A holder for the pipette is fitted on either the left or right side of the screen, according to which side you prefer to work (See Figure 3-1). The pipette should hang on the holder during measurements as well as when not in use.

The sensor probe

The sensor probe, where the actual detection takes place, is an optical fibre coated at the tip wich a gold layer (Figure 2-5). The detecting molecule is immobilized directly or indirectly an the surface to provide a biospecific sensor probe. A description of the surface characteristics and full instructions for immobilizing the detecting molecule and using the sensor probe are provided in a separate Sensor Probe Handbook.

The sensor probe is supplied in a protective tube. The tube is used for handling the sensor probe during fitting to and removal from the pipette, and for storing the sensor probe when it is removed from the pipette (see Section 4.4).

The optocable

The optocable is an optical fibre cable that transmits both the incident and reflected light between the detection unit and the sensor probe, The optocable is flexible, but should not be bent tightly since this may damage the optical fibre in the cable. The minimum diameter of curvature for bending the cable is about 13 cm.

The pipette

BIACORE@probe uses a specially designed pipette to hold the sensor probe. The sensor probe protrudes into the pipette tip. Measurements are made on sample aspirated into the pipette tip.

The red slider on the pipette has three positions (see Figure 2-6):

•            The top position is used when fitting or removing the sensor probe.

•            The middle position (locked by turning the slider) holds the tip of the sensor probe in the protective barrel, and is used when fitting or changing the pipette tip.

•            In the bottom position (locked by turning the slider), the sensor probe tip is extended into the pipette tip. This position is used when making measurements.

With the slider in the bottom position, the plunger at the top of the pipette aspirates and dispenses a fixed volume of about 100 µl. This is sufficient to cover the tip of the sensor probe when it is extended in the pipette tip. The pipette will only aspirate and dispense liquids when the slider is in the bottom position.

With the slider in the middle position, the plunger releases the pipette tip from the pipette.

Recommended pipette tips are 200 µl micropipette tips from Gilson.

Construction

Sensor Probe CM5 (Figure 2-1) consists of an optical fibre (diameter at the Sensor tip about 400 }.m) with a screw attachment for fitting to BIACORE®probe at one end. The fibre is covered with a protective cladding, except for 10 mm at the tip which is exposed to the solution being measured. This tip is coated with a thin layer of gold (— 50 nm thick) to provide conditions for SPR. The sensor probe is supplied in a capped protective tube which also serves as a holder when fitting the sensor probe to the probe holder an the Instrument.

Surface properties

The gold film an the tip of Sensor Probe CM5 is covered with a layer of carboxymethylated dextran, covalently attached to the gold through an inert linker layer. The dextran is swollen in aqueous media, providing an extensively solvated hydrogel with a thickness of about 100 nm (Figure 2-2). The detecting molecule is attached to the dextran matrix either directly or through binding to a covalently immobilized capturing molecule. Carboxymethyl groups an the dextran supply convenient sites for attachment of proteins using well-defined chemistry.

Attachment to the flexible dextran chains preserves a hydrophilic environment around the detecting molecules and provides a high degree of accessibility to the molecular surfaces. As a result, the biological activity of attached detecting molecules is retained to a high degree. This is a significant advantage of BIACORE°probe technology in comparison with techniques such as ELISA that rely an adsorption of proteins to plastic surfaces which can impair biological activity.

The dextran layer an Sensor Probe CM5 is chemically stable in most buffers used for preparing and studying biomolecules, and can be exposed to extremes of pH for short periods (2-3 minutes) without significant deterioration. In general, the stability of a functionalized sensor probe will be determined by the properties of the detecting molecule rather than of the dextran layer.


System specifications

Detection unit

Power supply                                  100-120/220-240 VAC auto-range 50-60 Hz

Maximum power consumption          160 W

Mains fuses                                    2 x T2.5AH

Light source                                   Halogen lamp (expected lifetime at least 7000 hours)

Connection to sensor probe             Optical fibre

Dimensions (1 x w x h)                    420 x 340 x 440 mm

Weight                                           13.5 kg

Sample measurement

Detection principle                          Affinity-based biosensor using surface plasmon resonance

Dynamic range                               Refractive Index 1.33-1.40 (0-70 kRU)

Sample volume required                   ~100 µl (sample is not consumed)

Sample temperature                        4-40°C

Analyse molecular weight                > 10,000 daltons

Typical detection Limit                     1 µg/ml using direct coupling*

                                                    2-5 µg/ml using capturing*

*Determined for myoglobin (molecular weight 17,000 daltons) using anti-myoglobin as detecting molecule.

Data storage

Internal data memory                       Flash disk

                                                    Sufficient for >1000 typical analyses

Working memory                            Sufficient for >1000 measurements per analysis

Diskette drive                                  3.5" 1.44 Mb PC-compatible

Accessories

Printer                                            A Hewlett-Packard DeskJet printer is recommended for graphical printouts. For text-only printouts, an IBM ProPrinter or compatible may be used.

For conformance with European Standards, the printer muss be CE­marked in accordance with the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC.

Environment

Instrument Operation

   Temperature                                15-35°C

   Relative humidity                          20-85% (non-condensing)

Noise level (LEQ)                            <40 dB(A)