Soldé

Shimadzu AA 6200

Seriennummer: 007890
Status: Archived Product
Preis auf Anfrage
Excl. 19% VAT , excl. Shipping Cost
siehe Storeview
Descrizione:
Shimadzu AAS AA 6200. Con Calculator. Tenere Sotto Controllo. Stampante. Software Spectro. Alta Temp Burner Head. TI. Dispositivo da banco. Anno: 1999.Neupreis: 18.407, - Euro.
Quotation
siehe Storeview
Maggiori Informazioni
Objektnummer B00007890
Seriennummer 007890
Nome oggetto Shimadzu AA 6200
Stato Archived Product

Gruppo prodotti: Assorbimento Atomico (Fiamma)

Non è stato trovato alcun prodotto corrispondente alla selezione.

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

Firma:              Shimadzu

Modell:            AA 6200

Kommentar:   Dokumente engl.

 

Nachfolgende Abbildungen und Beschreibungen sind modellbezogen und aus Prospekten entnommen.

Sie geben nicht den Lieferumfang des Systems wieder.

Den exakten Lieferumfang entnehmen Sie bitte aus dem Angebotstext.

 

The Shimadzu Atomic Absorption Spectrophotometer AA-6200 is used in the flame atomic absorption analysis and flame emission analysis. It is compact and can be operated easily, provid­ed with double beam optical system and D2 (deuterium lamp) background correction method.

Software for control of the AA-6200 is operated in MS-Windows TM environment, which allows to create measurement conditions easily even by person who first time operates the atomic absorption spectrophotometer, using the parameter setting Wizard.

When the AA-6200 is combined with the auto sampler ASC-6100, automatic analysis becomes possible for every elements . This function is useful to save energy.

 

Specifications

Hardware

Optics

Double Beam (chopper mirror)

Monochromator

aberration corrected Czerny-Turner monochromator Holographic grating (1,600 lines/mm)

Wavelength range

190-900nm Automated wavelength selection

Slit

0.2nm~0.7nm Manual setting

Background Correction

D 2 -Lamp method

Lamp Tunet

2-lamps simultaneously lit (manual turrent)

Lamp Mode

EMISSION, NON-BGC, BGC-D 2

Frequency

100Hz

Nebulizer

Nebulizer integrated impact bead and jacket tip Pt/Ir capillary

Chamber

Polypropylene

Burner

Fixed back/forward position and burner height

(Simply switching of Air/C 2 H 2 and N 2 0/C 2 H 2 burner)

Titanium 10 cm slot burner (Optional high temperature burner)

Gas Control

Manual setting of flow rate Automatic Air/N 2 0 switching system

Safety

Gas pressure monitoring to prevent flashback

Automatic flame monitoring

Safety interlock for misuse burner

Automatic flame extinghish when power failure

Ignition

Push ignite button

Software

Software based an MS-Windows TM

Dimension and weight

W690 x D425 x H370mm, 38kg

Power requirements

AC220V, 230V ± 10%, 50/60Hz, 300VA

(Certification of CE marking)

Ambient temperature and humidity range

Temperature : 10~35°C

Humidity : 45~80% (but less than 70% if temperature is over 30°C)

 

Software

Concentration conversion

Calibration method, standard addition method, simplified standard addition method

Measurement type

AUTOZERO, BLK, STD, UNK, MSA, SMAS, WAIT, PAUSE, FILESAVE

determines measurement type by designating above actions

Data processing

Linear regression analysis by least square method,

Abs=f (Conc.)

Calibration curve

Dimension : first order, second order, third order, ON/OFF of zero intercept selectable calibration curve by Curve ID

Repetitive analysis

Maximum 10 analysis per sample

Accuracy enhancement

Usage of average value, cancel of abnormal data by

the limitation of CV or SD values.

Sensitivity enhancement

Automatic calculation by sensitivity correction

Actual concentration calcuation

Automatic calculation of actual concentration by sample volume, constant volume, and dilution factor upon data acquisition

Signal process

 

Measurement time

Pre-spray: 99 seconds, Accumulation time: 99 seconds

Response

4 response time selectable

Measurement procedure setting

Easy setting by Wizard function

Data display MRT Worksheet

Measurement procedure, analysis results, actual concentration, calcuation factor of actual concentration, combined display of data and time

Signal display

display of multiple elements by tab sheet

Signal profile display area : 4

Current signal profile display area : 1

Overlay display of signal profiles by designating the row number of MRT

Retrieval and storage of data

Automatic loading of template parameter file upon software activation

Automatic data storage by "FILESAVE" command

Raw profile data can be stored

Import/Export of data

Import sample name from ASCII file

Export data file by ASCII file

Report

Summary report : display and output of results by sample

Output after analysis : Parameter, Raw data, signal, sequence

Selectable format and items for out put

 

 

 

ASC-1600 Autosampler

By connecting to the Shimadzu Atomic Absorption Spectrophotometer, the ASC-6100 per­forms the functions as the autosampler for flame absorption measurement. Also, when G attach­ment ASK-6100 is mounted, the ASC-6100 can perform a wide variety of following functions automatically in the furnace quantitative measurement by linking wich the Shimadzu Atomic Absorption Spectrophotometer main unit.

(1) Sample dilution

(2) Reagenz addition

(3) Mixture of samples and reagents at the mixing port

(4) Combined rinsine, of mixing port

(5) Automatic dilution and re-analysis for off-scale measurements due to excessive concentra­tion of unknown samples

(Dilution rate is automatically set. Possible dilution rate is 300 times at maximum.)

(6) Injection of high accuracy sample. Sampling amount can be changed in the unit of 1µl.

(7) Booting of sample concentration in Furnace measurement

 

For Flame and Furnace

Control

By RS-232C with AA

Function

Origin search

Self diagnosis

Automatic rinsing

Automatic preparation

Random access

Number of sample

8 for Reagent

60 for Sample

Irandom access available for both)

Sample cup

Sample 15ml (testing tube for Flame) or 2m[ (vial for Furnace)

Reagent 20ml

Rinse nozzle

Discharge rinsie liquid from nozzle (for Furnace)

Intake rinse liquid from nozzle (for Flame)

Bottle for rinse liquid

2l

Power

AC 100V ~ 240V (auto sensing). 100VA

Dimension • Weight

For Flame : W330 x H320 x D295mm , 7.5kg

For Flame and Furnace : W430 x H320 x D360mm, 15kg

 

For Furnace only

Function

Automatic dilution

Automatic reagent addition

Syringe

250µl

Injection volume

2 ~100 µl

Reproducibility

1% R.S.D (20µl )

Cary over

Rinsing port : under 0.00001

Mixing port :under 0.00001

Rinsing Mixing port

With rinse liquid

With next sample

Mixing

In the mixing port

Maximum volume for preparation . 0.6ml

Number of reagent

for addition

Maximum 4

Number of liquids for mixing

Calibration method : Maximum 5 (one sample + 4 reagents)

Standard addition method : Maximum 6 (one sample + one

Standard + 4 reagents)

Auto dilute and

re-analyze

When the absorbance exceeds the range of the calibration curve:

The optimum dilution ratio is automatically set.

(When it can not be calculated the dilution ratio is fixed at 10 times.)

 

 

 

 

 

Graphite Furnace Atomizer GFA-EX7

The GFA-EX7/GFA-EX7i is a graphite furnace atomizer which, when connected with the atomic absorption spectrophotometer, allows a high-sensitive quantitative analysis of inorganic elements by the electrothermal atomic absorption spectrometry (furnace method).

The Features of the GFA-EX7/GFA-EX7i are as follows:

(1) A high sensitivity optical temperature sensor is incorporated along with automatic temperature calibration in current control region. These assure accurate temperature repeatability.

(2) All the heating is Computer-controlled with the digital temperature control system to ensure the Optimum heating.

(3) Three graphite tube types, which are high-density graphite tube, pyrolytic-coated graphite tube and platform tube, are prepared according to the measurement element and sample characteristics.

(4) At the sample atomization stage, the high sensitivity mode setting allows trace analysis with even higher sensitivity and higher accuracy.

(5) At each heating stage, the precise control of inner gas flow rate and the automatic switching of gas types allow flexible furnace program composition.

(6) With the two-live gas switching system, air can be used at the ashing stage to accelerate the ashing of organic samples effectively.

Thun the GFA-EX7/GFA-EX7i is extremely effective in analyzing a wide Tange of sample types.

 

Specifications

Heating temperature range

   Room temperature              ~ 3000V

Heating control methods

   Ambient ~ 599 °C                 Current control method

                                               ( wich automatic temperature calibration function)

   600 ~ 3000°C                       Optical temperature control System

Heating parameters setup

   Stage number                      Max. 20

   Heating mode                       RAMP/STEP

   Inner gas type                       2-live switching

   High sensitivity mode setup

   Boost cycle :                       Max. 20 times

   Inner gas flow rate :             0 ~ 1.5 L/min

   Optimum temperature program search function

Safety measures

   Cooling water flow rate monitor

   Gas pressure monitor

   Overcurrent preventive mechanism

   Furnace block Cooling check

Graphite tubes (optional)

   High-density graphite tube

   Pyrolytic-coated graphite tube

   Platform type graphite tube

Cooling water

   Watet temperature               30°C or below

   Supply pressure                   0.08 ~0.15 MPa, Max.0.17 MPa

   Flow rate                              0.6 L/min or more

Gas

   Necessary gas type            Argon

   ALT gas (inner gas)            Argon, dry air

   Supply pressure                   0.32~0.38 MPa, max. 0.4 MPa (Ar, ALT)

Required power

   AC 200 V, 220 V, 230 V, 240 V ±5%, 6000 VA, 50/60 Hz in common

Power unit dimensions

   W 255(mm) X D 450(mm) X H 460(mm), Weight 41 Kg

Part of labexchange logo