Spettrometri

Agilent 4100 MP-AES

16.800,00 €
Excl. 19% VAT , excl. Shipping Cost
19.992,00 € incl. Vat
Seriennummer: 42368
Stato: Stock unit?The device is in our stock, however, has to be inspected by our technician, in order to be able to guarantee complete function.
Spedizione: 10-14 working days
Descrizione:
Agilent Spettrometro 4100 MP-AES.
Con nebulizzatore.
Camera di nebulizzazione a singolo e doppio passaggio.
Torcia. Finestre pre-ottiche.
PC, schermo, tastiera e mouse.
Accessori: tubi.
230 V. 50-60 Hz. 2100 W.
Add to Offer list
Maggiori Informazioni
Objektnummer B00042368
Seriennummer 42368
Nome oggetto Agilent 4100 MP-AES
Stato Lagergerät
Fabbricante Agilent
Spedizione 10-14 working days

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

For deliveries to countries outside the EU, an additional fee of 70,00€ for export is charged. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by credit card, letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

 

manufacturer:     Agilent

model:                  4100 MP-AES

 

 

The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.

 

Specifications

 

Power supply

Single-phase AC, 10 A, 3-wire system, 200–240 V AC, 50/60 Hz

Dimensions

960 mm (37.8 in) x 660 mm (26.0 in) x 660 mm (26.0 in)

Weight 73 kg (161 lb)

Wavelength range  213.857 to 614.171 nm (depending on element)

 

Optical system

Fast-scanning, high-resolution optical system for fast measurement using a compact design. Czerny-Turner design monochromator with 600 mm focal length and fi xed entrance slit for ease of operation. Holographic diffraction grating with 2400 lines/mm is blazed at 250 nm for optimum UV performance. Wavelength range 178–780 nm. The optical system can be purged with air for protection in dusty or corrosive environments, or nitrogen at 10 L/min for sulfur determinations.

Detector

Hermetically-sealed, UV-sensitive, back-thinned solidstate CCD detector (532 x 128 pixels) designed specially for low-light level detection with > 90% quantum effi ciency at peak sensitivity. It is directly cooled to 0 °C using a thermoelectric Peltier device for low dark current and reduced baseline noise. High dynamic range and blooming resistance allow fl exible measurement conditions. The CCD array detector collects the analyte spectra and neighbouring background spectra, allowing simultaneous background correction for enhanced stability and precision. Standby and startup When powered off, there is no standby gas or power consumption. With the nitrogen generator, a purge time of up to 20 minutes is required from power on, depending on the prior usage history, to allow the nitrogen generator to achieve required nitrogen purity. For optimum performance and stability of the MP-AES instrument, a warm up time of 30 minutes is recommended from standby

Gas controls

Computer-controlled solenoid valves are used to automatically enable and regulate the fl ow. The plasma gas is fi xed at 20 L/min and the auxiliary gas is fi xed at 1.5 L/min for ease of operation. Nebulizer gas fl ow is computer controlled using mass fl ow control to provide accurate fl ow control in the range 0.3–1.0 L/min

Plasma generator

The microwave excitation assembly features a solidstate HV power supply and an industrial-grade, aircooled magnetron operating at 2450 MHz. Fixed plasma power of 1 kW for ease of operation. Requires no cooling water supply, as the plasma generator is aircooled. Computer-controlled plasma ignition using a momentary fl ow of auxiliary argon. Once the plasma is operational, it automatically switches to nitrogen for routine operation

Signal stability

< 2% RSD over 2 hours without internal standardization or any form of drift correction.

Resolution

< 0.050 nm (measured as full width at half maximum).

Detection limits 3 sigma detection limits (µg/L) using a 10 second integration time. Pure nitrogen and nitrogen generator IDLs Element

Full environmental range*

Mn 257.610 nm 5.00 ppb Ba 614.171 nm 1.5 ppb * Full environmental range Altitude Operating temperature 0–3000 m (0–9800 ft) 15–30 °C (59–86 °F) 3000–4000 m (9800–13100 ft) 15–25 °C (59–77 °F)

 

Back to Top
part-of-labexchange-group
Search engine powered by ElasticSuite
Labprocure