WTB Binder C02-Inkubator
Objektnummer | B00014105 |
---|---|
ID-number | 014105 |
Object name | WTB Binder C02-Inkubator |
Status | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
For deliveries to countries outside the EU, an additional fee of 70,00€ for export is charged. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by credit card, letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Firma: WTB Binder
The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representing the delivery volume. The exact delivery content you will find only in the offering text.
1 Instrument box
2 Keys-operating with LED-display
3 Main switch
4 Operation and alarm box
5 Door contact switch
6 Outer unit door
7 Door heating
8 Inner glass door with closable measuring port
9 Inner chamber
10 Shelf holder (not visible)
11 Setting screws at the front part
12 Shelves
13 Homogenisation fan
14 CO 2 sensor
15 Permadry water basins
The inner chamber made of stainless steel 1.4571 (V4A) is deep-drawn from one piece, electropolished (suitable for pharmacy) and has no welds or inaccessible corners. The hinges and the seal of the inner glass door are glued form the outside, which is also very hepful for an easy cleaning of the inner chamber. The complete shelf holder and the perforated shelves are also made of stainless steel and are electropolished. The shelf holder offers 11 different positions for the shelves to be adjusted.
The complete shelf holder with the shelves can be adjusted via four setting screws this allows the uniform distribution of the nutrient solution in the flasks and dishes. The basins can be removed and, therefore, they can easily be cleaned.
With the heating system of the CO 2 -incubator CB 210 a hot-air auto-sterilization at 185°C can be effected, thus the whole inner chamber can be decontaminated.
Thanks to the standard safety device (TWW cl. 3.3 according to DIN 12880) the set temperature is kept in case of failure.
A highliy precise, drift-free CO 2 infrared measuring system with a homogenisation system guarantees precise and constant CO 2 concentrations via long periods of time. This creates optimum growth conditions for cultures. Both, the CO 2 measuring system and the homogenisation system can be removed from the inner chamber by hand and can be cleaned with suitable detergents.
The CO 2 gas is fed into the chamber via a fine filter (aseptic filter) with a high filtration efficiency, which also filters the smallest particels. This fine filter is easily accessible and can be exchanged by hand.
Technical data CB 210
Dimensions housing |
|
Width B mm |
740 |
Height H mm, |
1056 |
Depth T mm |
695 |
Height feet HF |
22 |
Wall clearance mm |
40 |
Depth of the instrument panel |
50 20 |
Inner chamber |
|
Width B1 m |
560 |
Height H1 mm |
750 |
Depth T1 mm |
495 |
Volume inner chamber I |
210 |
|
|
Shelves |
|
Number standard/max. |
3/11 |
Dimensions Width |
510 |
Depth |
460 |
Max. load kg per shelf inserted/pulled out |
20/10 |
|
|
Weight of unit (empty) kg |
128 |
|
|
Temperature |
|
Temperature range °C At least 8°C above ambient temperature |
60 |
Temperature variation ± °C at 37°C DIN 58945 |
0,4 |
Temperature fluctuation ≤±°C |
0,1 |
Recovery time in minutes after door was opened for 30 sec. at 37°C (Up to 96% of the end value) |
8 |
|
|
CO 2 control |
|
Setting range %CO 2 |
0-20 |
Setting accuracy %CO 2 |
0,1 |
Recovery time to 5 % CO2 in minutes after door was opened for 30 sec (Up to 98% of the end value) |
6 |
C0 2 measurement |
IR |
Connection hose nozzle for C0 2 mm |
8 |
|
|
Hummidity |
|
Relative humidity (constant % rH) |
95 |
*Recovery time to 95% rH in minutes after door was opened for 30 sec |
24 |
|
|
Nominal voltage 50/60 Hz |
|
Excess voltage category |
2 |
Nominal voltage V (± 10 %) |
230 |
Nominal power Watt |
1600 |
Energy consumption Wh/h |
140 |
Unit checked acc. to VDE 0113 |
* |
|
|
Ambient conditions |
|
Contamination level |
II |
Temperature °C Operation (min/max.) |
18/40 |
Stocking (min/max.) |
-10/50 |
Max. humidity r.H. % |
60 |