Archive

Wave Biotech AG Thermasept II

ID-number: 021950
Status: Archived Product?The unit is no longer available.
Shipping:
Description:
Wave Biotech Welding unit Reewelder Thermasept II. For fully automatic welding of thermoplastic hoses laboratory. Sterile and validated
siehe Storeview
More Information
Objektnummer B00021950
ID-number 021950
Object name Wave Biotech AG Thermasept II
Status Archiv

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, paymet by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

manufacturer:       Wave Biotech AG

model:                   Thermasept II

annotation:           Dokumente engl.

 

 

 

 

 

 

 

Technical information

 

 

The BioWave 200SPS-F contains all features for production. There is only one execution available.

 

 

BioWave 200SPS-F

 

Operation

Fully automated, PLC controlled system, operation by front coloured LCD touch screen display with back-lightning. The display can be rotated. Fully menu-guided operation. With integrated features:

Password:

Viewer, User and Supervisor mode.

Alarm:

Exceeds the actual values, the pre-defined upper and lower alarm values and alarm message will be shown on the display. Alarm outputs can be connected at any customer devices to act respectively.

Controller:

pH can be controlled by connected peristaltic pumps (add base and acid) or CO 2 content in air flow. DO can be controlled by 0 2 content in the air flow.

Stopwatch:

For easier writing of protocols.

Time:

Integrated, battery powered date for data acquisition.

Acceleration mode:

Increasing/decreasing rock rate during pre-adjusted time.

Working hours:

Counting total working hours of BioWave.

Set flow:

Adjustable at display. CO 2 and 0 2 content are predefined in %. Flexible hoses with couplings to connect bags are included with BioWave 200 SPS-F.

 

Quick mounting of Wave Bags

Two large handles at the front rear side allow to fix the Wave Bags immediately and securely without damaging the Wave Bag.

 

Aeration control

Sucks either ambient air for aeration through a connector at the rear side panel or receives pressurised air connected at the pressurised air port. A positive pressure applied at the pressurised air port stops automatically the air pump.

 

Gases to connect

·         Ambient air: Centrifugal air pump built-in, max. pressure 500 mbar.

·         Pressurised air port: Built-in pressure reducer, Input pressure +0.8...+1.0 bar, to connect by user. Safety pressure measurement avoiding to high over pressure in the Wave Bag.

·         CO 2 , +0.8...+0.8 bar

·         O 2 , +0.8...+0.8 bar

 

Air flow rate

·         Using ambient air: 0...4.0 l/min, +/- 0.2 l/min.

·         Using pressurised air: 0...4.0 I/min, +/- 0.2 l/min.

 

CO 2 content in the air flow

1.0...12 % +/- 0.5 %

 

O 2 content in the Air Flow

21... 50 % +/- 1.0 %

 

Measurement CO 2 and O 2

Content of CO 2 and O 2 in air flow are measured by IR sensor.

 

Over pressure protection

Pressure transmitter measuring constantly the positive pressure. Switches off

the air flow in case of too high pressure to avoid too high pressure in the Wave Bag.

 

No Flow Alarm

An alarm will be shown on the display in case of no air flow is dedected.

 

DO and pH measurement

The two light fibre cables from the pH/DO inserted into the respective ports of

the Wave Bags D opt. are measuring the light luminescence decay time response from the sensor caps. The light fibre cables are marked with colours to avoid insert into the wrong port of the Wave Bag. The DO and pH of only one Wave Bag D opt. can be measured at the same time. For principle of measurement see data sheet Wave Bag D opt.

 

Parameters measured

DO content of media: 0...200 % +/- 2% (yellow)

pH of media 5.5... 9.0, +/- 0.2 (red)

 

Calibration

The Wave Bag D opt. is marked with individually calibration codes. This codes have to be entered at the display. There is no periodically calibration required.

 

Temperature influence of media

The actual temperature of the media in the Wave Bag D opt. has to be entered by the operator to compensate the temperature shift.

 

Length of fibre cables

2.5 m.

 

Filter Heater

Heating of Wave Bags generates small amounts of vapour rising into the air outlet filter. The filter heater keeps the vent filter dry and prevents condensation from plugging the filter media. A green LED shows the heater is heating.

 

Temperature range

40...50°C, +/- 5°C

 

Material

Silicon

 

Heating power

2 Watt

 

Installation

The filter heater is put over the outlet filter manually without using any tools.

 

Power Supply

Built-in BioWave 200SPS, 6 VDC, 2m cable length

 

Number included

2

 

 

 

General data

 

Rocking Rate

3...20 rocks/min, +/- 1 rocks/min, adjustable at display.

 

Rocking Angle

4...10°, +/-0.3°, adjustable at display. Actual angle is shown on display.

 

Temperature Control

On PLC programmed PID controller. Heating of left side and/or right side for

heating of one 100L or two 100L or one 200L mounted Wave Bags at different temperatures. Selectable at front LCD display as the set temperature as well.

 

Temperature range

+20...+40 °C, +/- 0.2 °C.

 

Temperature measurement

Resistance thermometer 1x Pt100 class A (IEC751) built-in in button-housing for measurement of Wave Bag content. Two resistance thermometer are included measuring the temperature of two mounted Wave Bags

 

Serial interface

1x RS232/RS485, for either printer EPSON or software PCDAQ.

 

Material housing and Frame

Stainless steel 1.4301 (304), brushed, electro polished.

 

Safety

Emergency switch off, safety rubber bar installed at rocking table. They stop

rocking when touched. Key to lock the main power supply of the BioWave 200SPS.

 

Wheels

Stainless steel 1.4301 (304), polyurethane.

 

Transportation

Tilt position reducing required width for transportation WxDxH 1'930x890x1'200 mm.

 

Dimensions

WxDxH 1'930x1’100x1’100mm.

 

Weight

Complete system, without Wave Bags: 210kg.

 

Power supply

230 V -50 Hz, 10 A

 

Ingress protection

IP40 minimum

 

Ambient temperature range

0...+50°C Note: only for equipment and not for process application

 

 