Varian CP-3800
| Objektnummer | B00021211 |
|---|---|
| ID-number | 021211 |
| Object name | Varian CP-3800 |
| Status | Stock unit |
Product group: GC / MS
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
|
Country |
Possible payment methods |
Comment |
|
DE, AT, CH |
Payment by invoice, payment in advance, paymet by credit card |
Payment by invoice is only possible for corporate clients. |
|
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
|
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
|
Status |
Condition |
Comment |
|
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
| In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
|
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
|
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
|
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |

Outstanding Flexibility with Proven Performance
The Varian CP-3800 GG for any application
Flexible platform single, dual, or three-channel configurations for maximum productivity.
Easy upgrades quickly add additional capabilities as your requirements change.
Simplified PC control password security, Star Chromatography Workstation Quick Start Applications, and Custom Report. Writer templates streamline operation
The Varian CP-3800 GC meets all your analytical requirements with practical solutions for simple and complex applications.
In a standard configuration, the CP-3800 accommodates up to three injectors and three detectors, all operating simultaneously. In addition, the CP-3800 is easily configured to perform on-line analyses for monitoring critical gas and liquid process streams.
The CP-3800's flexible platform lets you start with a basic configuration and add a wide variety of options in the field. Options include additional channels, gas and liquid sampling valves, and external sampling devices. Future upgrades guarantee that you can meet changing requirements at a moment's notice.
Star Chromatography Workstation combined with Quick Start Applications provide a powerful yet simple interface for controlling the CP-3800 GC. Routine applications can be launched from the desktop with minimal data entry. Reports specific to your application are easily created with Star Custom Report Writer. Report generation can be automated for around-the-clock monitoring.
The CP-3800 GC from Varian-unlimited possibilities!
Simple and easy to operate
Flexible custom configurations
Stand-alone, PC, and client/server control
Injectors
CP-1177 Split/Splitless Injector with optimized
1079 PTV (Programmable Temperature Vaporizing) injector for cold on-column, LVI (Large Volume Injection), and solid sample introduction with ChromatoProbe
1093 SPI (Septum-Equipped Programmable Injector) for high temperature capillary column applications
1041 On-Column/Packed Injector for 0.53 mm ID capillary and packed column applications
1061 Flash Injector for simplified direct injections
Valve Systems for gas and high pressure liquid samples
Integrated SPT (Sample Preconcentration Trap) for concentrating gas samples
Column oven
Spacious oven accommodates three or more columns
Varian Quick Connects provide easy column Installation
Rapid oven cool down increases sample throughput
Cryogenic cooling extends application range
Detectors
Flame lonization (FID) with a patented, ceramic flame tip for ultimate peak shape and sensitivity
Thermal Conductivity (TCD) with automatic filament protection for lowest maintenance costs
Electron Capture (ECD) for ultra-high sensitivity and selectivity of halogenated compounds
Thermionic Specific (TSD) for selective detection of nitrogen/phosphorus compounds
Pulsed Flame Photometric (PFPD) for selective detection of sulfur, phosphorus, nitrogen, and 25 additional elements
Galaxie Chromatography Data System
Simple Interface for easy method building and data acquisition
Password protection eliminates unauthorized data access and method editing
Easy for occasional users with customized screens showing just the Information needed
Simpfified user interface
Large display with 11 lines and 35 characters/line for easy viewing
Easy method navigation for greater productivity
Access to all GC parameters for complete method editing and System control
AutoSamplers
CP-8400 handles up to 100 samples with liquid, ambient headspace, or optional SPME sampling modes
CP-8410 with a fixed Sample trag Supports multiple vial sizes
Combi PAL offers maximum flexibility for liquid, heated headspace, and SPME applications
Archon with Tekmar Vetocity XPTTm for automated Purge and Trap sampling
CP-3800 GC Specifications
Dimensions and Weights
Size: 53 cm (h) x 66 cm (w) x 56 cm (d)
Weight: 43 kg
Environmental Conditions
Operating temperatures: 10°C to 40°C
Operating humidity (relative): 5% to 95%
Line voltage requirements: 101V, 120V, 230V (±10% nominal)
Column Oven
Dimensions: 28 cm (w) x 20 cm (d) x 28 cm (h)
Temperature range: arnbient +4 °C to 450 °C
With liquid N 2 : -99 °C to 450 °C
VVith liquid CO 2 : -65 °C to 450 °C
Temperature program ramps/holds: 7/8
§ Maximum temperature ramp rate: 100 °C /min for all voltages
§ Cool down rate: 400 °C to 50 °C in 4.5 minutes
§ Temperature set-point resolution: 1 °C
Injector Options
Maximum injectors: three operating concurrently
Pneumatics: manual or electronic flow control (EFC)
Columns: capillary columns standard with optional packed column adapter
Injectors
CP-1177 Split/Splitless
1079 Programmable Temperature Vaporizing (PTV)
1061 Flash Megabore/Packed
1041 On- Column Megabore/Packed
CP-1177 Split/Splitless
Pressure range: 0-100 psi
Total flow: 500 rrillrnin at 10 psi
Maximum temperature: 450 °C
Split range: 1-10,000 (column dependent)
1079 Programmable Temperature Vaporizing (PTV)
Pressure range: 0-100 psi
Total flow: 500 mL/min at 10 psi
Temperature range
-99 °C to 450 °C liquid N 2
-65 °C to 450 °C liquid CO 2
Maximum temperature ramp rate: 180 °C/min
Temperature ramps: 4
Split range: 1-10,000 (column dependent)
Operational modes
Temperature ramped splitless
Cold on-column
Large Volurne Injection
Split and splitless
ChromatoProbe solid sample introduction:
optional
1061 Flash Megabore/Packed
Pressure range: 0-100 psi
Total Flow: 50 mL/min (Type 3 EFC)
500 mL/min (Type 4 EFC)
Maximum temperature: 450 °C
1041 On-Column Megabore/Packed
Pressure range: 0-100 psi
Total Flow: 50 mL/min (Type 3 EFC)
500 mL/min (Type 4 EFC)
Maximum temperature: 450 °C
Quick-Switch Valve Option
Instantly switch between injectors/columns and cletectors
Configurations: automated or manual, factory or field installed
Detector Options
Maximum detectors: three operating concurrently
Pneumatics: manual or electronic flow control (EFC)
Columns: capillary columns standard with optional packed column adapter
Detectors
Flame lonization Detector (FID)
Thermal Conductivity Detector (TCD)
Electron Capture Detector (ECD)
Thermionic Specific Detector (TSD or NPD)
Pulsed Flame Photometrie Detector (PFPD)
FID
Maximum temperature: 450 °C
Detectivity: 2 pg C/see
Linear dynamic range: 10 7
Patented ceramic -Flame tip: standard
Flame-out detection/auto re-ignition: standard
TCD
Maximum temperature: 450 °C
Detectivity: 300 pg/sec (C4)
Linear dynarnic range: 106
Filament proteefirm: standard
Automatic bridge balancing: standard
ECD
Maximum temperature: 450 °C
Detectivity: 7 fg/sec Lindane
Linear dynamic range: 10 4
TSD
Maximum temperature: 450 °C
Detectivity:
N: 100 fg N/sec (Azobenzene)
P: 100 fg P/sec (Malathion)
Maximum temperature: 450 °C
Linear dynamic range:
N: 10 5
P: 10 4
Self-aligning bead: standard
PFPD
Maximum temperature: 450 °C
Detectivity: using 5/P Photomultiplier Tube
S: 1 pg S/sec
P: 100 fg P/sec
N: 20 pg N/sec
Linear dynamic range:
S: 10
3
P: 10
4
N: 10
2
Electronic Flow Control: injectors
Module types: 4 injector-specifie modules
Accuracy: ± 0.1 psi, 5% full scale flow
Resolution: 0.1 psi or 0.1 mL/rnin
Electronic Flow Control: Detectors
Module types: 6 detector-specific modules
Accuracy: ± 7% set point flow
Resolution: 1 mL/min
Communications
Three channel analog output time prograrnmable (0-1V or 0-10V): standard
Three channel digital output and control with TCP/IP Ethernet and LAN connectivity: optional
Synchronization signals: for 'ready" and 'start" with other devices and data systems
Data Handling
Star Integrator: optional
Star Workstation (PC) control: optional
Galaxie Workstation (PC) control: optional
Galaxie Chromatography Data System (CDS) client/server control: optional
Other Specifications
Powered external events: 7
LCD display: 11 lines, 35 characters/line
Maximum stored methods with battery protection: 8
Internal autosampier control: standard
On-line hell): standard
Complete local control: standard
Automation Options
CP-8410 Autoinjector
Sample eapacity: 10 x 2 mL, 6 x 5 mL, and 5 x 10 mL vials
Dual and duplicate mode: standard Internal standard addition: standard
Modes of Operation: liquid, standard; ambient headspace, optional; and SPME (Solid Phase Microextraction), optional
Pre-programmed modes of injection: standard
CP-8400 AutoSampler
Sample capacity: 100 x 2 mL vials
Dual and duplicate mode: standard
Internal standard addition: standard
Modes of operation: liquid, standard; ambient headspace, optional; and SPME, optional
Sample heating and cooling: optional
Pre-programmed müdes of injection: standard
Combi PAL AutoSampler
Sample trays: two standard and expandable to four
Tray typcs: 98 x 2 mL vials; 200 x 1 mL vials;
32 x 10 m1120 mL vials; and 96-well plates
Dual and duplicate mode: standard
Internal standard addtion: standard
Modes of Operation: liquid, standard; heated headspace, optional; and SPME, optional
Sample heating and cooling: standard
Optional modules: additional sample trays, micro-well plate holders, wach station, SPME fiber bake-out station, dilutor, barcode readers, and flowcell