Vacuubrand MZ2C
Objektnummer | B00018087 |
---|---|
ID-number | 018087 |
Object name | Vacuubrand MZ2C |
Status | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, paymet by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
Firma: Vacuubrand
Nachfolgende Abbildungen und Beschreibungen sind modellbezogen und aus Prospekten entnommen.
Sie geben nicht den Lieferumfang des Systems wieder.
Den exakten Lieferumfang entnehmen Sie bitte aus dem Angebotstext.
Technische Daten
Max. Saugvermögen 50/60 Hz m²/h 1,7/2,0
(nach DIN 28432)
Enddruck (absolut) mbar 9
Enddruck (absolut) mit Gasballast mbar 20
Max. zulässiger Druck am Auslass bar 2
(absolut)
Max. Druckdifferenz zwischen bar 1,1
Ein- und Auslass
Zulässige Umgebungstemperatur bei °C -10 bis +60/+10 bis +40
Lagerung / Betrieb
Zulässige relative % 30 bis 80
Umgebungsluftfeuchte bei Betrieb
(nicht betauend )
Max. Strom / Antriebsleistung
100 V~ A/W 3,4/180
120 V~ A/W 3,0/180
230 V~ A/W 1,6/180
400 V~ Mot. EEx e II A/W 0,59/180
Max. Leistungsaufnahme
100 V~ VA 340
120 V~ VA 360
230 V~ VA 370
400 V~ Mot. EEx e II VA 410
Max. zulässiger Bereich 100 V~ +10%/-10% 50/60 Hz
der Versorgungsspannung 120 V~ + 5%/-10% 60 Hz
230 V~ +10%/-10% 50/60Hz
400 V~+5%/-5% 50Hz Mot.EEx eII
Nenndrehzahl min -1 1500/1800
Motorschutz thermischer Wicklungsschutz
Schutzart nach IEC 529 IP 54
Messflächenschalldruckpegel dBA 42
(Abstand lm)
Einlass / Auslass Schlauchwelle DN 10 NPT 1/8-27
Abmessungen über alles L x B x H mm 266 x 241 x 170
Gewicht betriebsfertig kg 10,5
Komponenten Medienberührte Werkstoffe
---------------------------------------------------------------------------------------------------------------------------
Gehäusedeckelinnenteil PTFE kohlenstoffverstärkt
Kopfdeckel ETFE
Membranspannscheibe ETFE
Ventil MZ 2C: FFKM (z. B. Kalrez ® '*)
Membrane PTFE-NBR (z. B. Perbunan ® **) Sandwich
Einlass/Auslass/Verschraubung ETFE
Schlauch PTFE
* eingetragenes Warnzeichen Du Pont
**eingetragenes Warenzeichen Bayer AG
Technische Änderungen vorbehalten!
Beim Einbau in ein Vakuumsystem
- Drosselverluste vermeiden durch kürzest mögliche Vakuumverbindungsleitungen großer Nennweite.
- Übertragung mechanischer Kräfte durch starre Verbindungsleitungen vermeiden und elastische Schlauchstücke oder Federungskörper zwischenschalten. Achtung: Elastische Elemente können sich beim Evakuieren zusammenziehen.
- Vorteilhaft: Ventil auf Saugstutzen zum Warmlaufen/Nachlaufen aufbauen.
- Kondensatrückfluss aus der Auspuffleitung in die Pumpe verhindern, Auspuffleitungen stets fallend verlegen.
Vor Inbetriebnahme
- Max. Umgebungstemperatur: 40 °C.
- Beim Einbau in ein Gehäuse oder bei hoher Umgebungstemperatur für gute Belüftung.
- sorgen. Mindestabstand von 20 cm zwischen Lüftern und angrenzenden Teilen einhalten. Bei einem Aufstellungsort über 1000 m über NN (Gefahr von unzureichender Kühlung) sind Maßnahmen gemäß DIN VDE 0530 zu ergreifen.
- Bei Stromausfall kann es bei geöffnetem Gasballastventil der Pumpe zu unbeabsichtigtem Belüften kommen. Kann dies zu Gefahren führen, geeignete Sicherheitsvorkehrungen treffen.
- Drehrichtung des Motors prüfen (nur Drehstrommotoren): Lüfter durch die Lüfterhaube beobachten. Pumpe einige Sekunden einschalten und überprüfen, ob die Drehrichtung des Lüfters mit der angegebenen Drehrichtung übereinstimmt. Falls die Drehrichtung nicht übereinstimmt, sofort abschalten. Händler bzgl. des weiteren Vorgehens kontaktieren.
- Leckagen bei der Installation zuverlässig verhindern. Nach der Installation Anlage auf Leckagen überprüfen.
Beim Betrieb
Die Pumpe darf nur gegen max. 1 bar Differenzdruck zwischen Einlass und Auslass gestartet werden , da sonst eventuell der Motor blockiert und Schaden nimmt.
- Höchstzulässigen Druck am Auslass und max. zulässigen Differenzdruck zwischen Einlass und Auslass beachten.
Das hohe Verdichtungsverhältnis der Pumpen führt dazu, dass sich am Auslass ein höherer Druck ergeben kann, als die mechanische Stabilität zulässt.
Die Pumpe erreicht die angegebenen Werte für Saugvermögen und Enddruck sowie die Dampfverträglichkeit erst bei erreichter Betriebstemperatur (nach ca. 15 Minuten).
- Kondensation in der Pumpe, sowie Flüssigkeitsschläge und Staub vermeiden, da eine Dauerförderung von Flüssigkeiten oder Staub Membrane und Ventile schädigt.
- Die Pumpe mit Gasballast betreiben, um die Kondensation von gepumpten Stoffen (Wasserdampf, Lösemittel ) in der Pumpe zu verringern.
Ein thermischer Wicklungsschutz schaltet den Motor bei Übertemperatur ab.
- Achtung: Nur manuelle Rückstellung möglich. Pumpe ausschalten oder Netzstecker ziehen. Vor dem Wiedereinschalten ca. fünf Minuten warten.