Sold

Merck-Hitachi L-6200

ID-number: 005566
Status: Stock unit
Preis auf Anfrage
Excl. 19% VAT , excl. Shipping Cost
siehe Storeview
Description:
Merck intelligent Gradient Pump LC-6200. With Low Pressure GradientAccessory and LC- Organizer.
Quotation
siehe Storeview
More Information
Objektnummer B00005566
ID-number 005566
Object name Merck-Hitachi L-6200
Status Stock unit

Product group: HPLC-Pumps (gradient)

We can't find products matching the selection.

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance, paymet by credit card

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance, payment by credit card

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance, payment by credit card

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Labprocure

Used

Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen.

Firma:                Merck-Hitachi

Modell:              L-6200

Kommentar:      Dokumente engl.

The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representating the delivery volume. The exact delivery content you will find only in the offering text.

The Model L-6200 is an intelligent pump which plays a major role in the Hitachi high speed liquid chromatography system. Its functions are detailed below.

Front View of Model L-6200 Intelligent Pump

The Model L-6200 Intelligent Pump is included in the Hitachi high Speed liquid chromatography system and has a controlling function.

The liquid chromatography system is available in three types. This manual gives Information an assembly and Operation of each type of system.

Type of Liquid chromatography System

Manual Pump System

The Model L-6200 is used as a single function pump like the Model L-6000.

Low Pressure Gradient System

The eluents are mixed before being pressurized by the pump. Accordingly this mixing method is called a "low pressure gradient elution". , It employs a proportioning valve for changing over the three eluent flow paths.

High Pressure Gradient System

The eluents are mixed after being pressurized by the pump. Accordingly this mixing method is called a "high pressure gradient elution". It uses two or three pumps for pressuriz­ing each eluent independently.

SPECIFICATIONS

Pump Unit

(1) Liquid feed System                            Dual plunger type reciprocating pump

(2) Maximum discharge pressure           400 kgf/cm 2

(3) Discharge rate range                         1 to 9999 µl/min (in increments of 1 µl,/min)

(4) Flow rate setting                                 Ten-key input

(5) Pressure display                                       Digital display (unit selectable among bar, psi, kgf/cm 2 and MPa)

(6) Pressure limit                                     0 to 412 bar, 0 to 5983 psi,

                                                                0 to 420 kgf/cm 2 , 0 to 41.2 MPa

(7) Discharge volume of                          100 µl/stroke cylinder no.1

(8) Flow rate setting                                 ±2 µl/min

accuracy (at constant                              (10 to 100 µl/min at 5 to 400 kgf/cm 2 )

room temperature of 20°C                     

for distilled water)                                    ±2 % (101 to 5000 µl/min at 5 to 400 kgf/cm 2 )

                                                                ±2 % (5001 to 9999 µl,/min at 5 to 20 kgf/cm 2 )

+2 % to -0.1P % where P devotes discharge pressure in kgf/cm 2

(5001 to 9999 µl/min at 21 to 400 kgf/cm 2 )

(9) Flow rate stability                               ±0.3 % (10 to 5000 µl/min,

(at constant room                                    5 to 400 kgf/cm 2 )

temperature of 20°C for

distilled water)                                          ±1.0 % (5001 to 9999 µl/min,5 to 400 kgf/cm 2 )

(10) Compensation of solvent                 Automatic compensator built in

compression ratio

(11) Pressure display accuracy:            ±5 %

(12) constant pressure liquid feed           20 to 400 kgf/cm 2 (for column packing)

(13) Material of wetted parts                           SUS 316, rubby, sapphire and teflon containing carbon

Gradient System

A. Low Pressure Gradient System

(1) Number of eluents                              Up to 3 eluents

                                                                (4 eluents with ACC pump)

(2) Mixing                                                 Valve-time control

(3) Mixing ratio range                               0 to 100 %

                                                                (in increments of 1%)

(4) Flow rate range                                  1 to 9999 µl/min in increments of 1 µl/min

The guaranteed range for gradient elution analysis is 0.3 to 3 ml/min

(5) Pressure limit                                     0 to 412 bar, 0 to 5983 psi,

                                                                0 to 420 kgf /cm 2 , 0 to 41.2 MPa

(6) Gradient elution profile                       Linear/multi-step

(7) Number of programming                    100 steps max. per program

steps                                                        100 steps max. for total of 9 programs

(8) Programming time                             Up to 600 min

                                                                (in increments of 0.1 min)

(9) Programming items                           1) B/C mixing ratio

                                                                2) Flow rate (main pump and ACC pump)

3) Wavelength (when L-3000 is connected via PAN interface)

                                                                4) Timer (marker) output

                                                                5) Pump stop, buzzer

(10) Number of programs                        9

(11) Display                                             Liquid crystal

                                                                (80 characters/2 lines)

(12) Contents of display                          1) Input parameters

                                                                2) Monitor

(time progress, mixing ratio, flow rate, pressure, timen, etc.)

High Pressure Gradient System

(1) Number of eluents                              2 or 3 eluents (with L-6000 combined)

(2) Mixing                                                 Double or triele pump flow rate control

(3) Mixing ratio range                               a) With L-6000 combined 0 to 100 %

                                                                b) With 655A-12 combined 0 to 100 %

                                                                c) With 655A-11 combined

                                                                      

                                                                10 to 90 % (in increments of 1 %, at 1 ml/min)

(4) Flow rate range                                  a) 1 to 9999 µl/min (in increments of 1 µl/min)

                                                                (with L-6000 combined)

                                                                b) 1 to 9990 µl/min (in increments of 1 µl/min)

                                                                (with 655A-12 combined)

                                                                c) 10 to 9900 µl/min (in increments of 10 µl/min)

                                                                (with 655A-11 combined)

(5) Guaranteed range for                         a) 0.3 to 3.0 ml/min

gradient elution analysis

                                                                b) 0.05 to 0.3 ml/min (when used as semi-micro LC)

(6) Pressure limit                                     0 to 412 bar, 0 to 5983 psi,

                                                                0 to 420 kg/cm 2 , 0 to 41.2 MPa

(7) Gradient elution profile                       Linear/multi-step

(8) Number of programming steps          100 steps max. per program

                                                                100 steps max. for total of 9 programs

(9) Programming time                             Up to 600 min

                                                                (in increments of 0.1 min)

(10) Programming items                         1) Mixing ratio of eluent B or eluents B and C

                                                                2) Flow rate

3) Wavelength (when L-3000 is connected via PAN Interface)

                                                                4) Timer (marker) Output

                                                                5) Pump stop, buzzer

(11) Number of programs                        9

(12) Display                                             Liquid crystal

                                                                (80 characters/2 lines)

(13) Contents of display                          1) Input parameters

                                                                2) Monitor

(time progress, mixing ratio, flow rate, pressure, timer, etc.)

Other Common Items

(1) Contact input/output terminals           Output: 1 Busy, 2 Start, 3 Stop

                                                                Input:     1 Busy, 2 All Stop, 3 Start, 4 Local/Remote

(2) External communication                    PAN and RS-232C possible

(3) Timer or timer output                         8 contacts

(4) Specification of pump                        1 Manual pump mode

control mode                                            2 Low pressure gradient mode

                                                                3 High pressure gradient mode

                                                                4 Column packing mode

(5) Specification of pressure                   Bar, psi, kgf/cm 2 or MPa unit

(6) Pressure zero adjustment                 Possible

Operating Conditions, Dimensions and Weight

(1) Power supply                                     100 to 115 V or 200 to 240 V AC, 50/60 Hz

100 or 200 V System is selected via the external switch at rear panel of main unit.

(2) Power consumption                           Approx. 110 VA

(3) Ambient temperature                         15 to 35°C

(4) Humidity                                             45 to 85 % RH

(5) Dimensions                                               260(W) x 210(H) x 400(D) mm (protrusions not included)

(6) Weight                                                Approx. 14 kg

Part of labexchange logo
Search engine powered by ElasticSuite