Merck-Hitachi D-7000 HPLC
Objektnummer | B00017618 |
---|---|
ID-number | 017618 |
Object name | Merck-Hitachi D-7000 HPLC |
Status | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, paymet by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
manufacturer : Merck-Hitachi
model : HPLC
annotation : Dokumente engl.
The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
The systems are modular designed.
Please finde the modules you are looking for.
L-7100 HPLC-Pump
Precision and Reliability
The L-7100 has been designed along the lines of the proven serial dualpiston principle, characterised by an extremely constant flow rate, low pulsation and reliable delivery of all commonly used solvents. Moreover, the L-7100 is able to form solvent gradients with high precision and accuracy. Each individual component and the finished instrument are subject to stringent quality control measures. A validation certificate and the manufacturer's test report documents that all functions and specifications have been fulfilled.
Versatile in Application
User-friendly programming via keyboard and LCD dialogue enable the L-7100 to be used as
n Isocratic HPLC pump
n Quaternary low-pressure gradient pump in conjunction with a gradient accessory
n High-pressure gradient pump in conjunction with one or two further L-7100 pumps and as
n Column packing device at constant pressure.
The L-7100 pump can store
n 9 time programmes, in which mobile phase composition, flow rate and four events can be programmed and varied over a period of time.
n Via a Link function, programmes for multi-method applications can be linked.
A shut-down routine can be programmed to be carried out at the end of an analysis series by
n Delayed pump stop and a
n System flush programme.
Due to minimised internal volume, a highly constant flow-rate and excellent gradient reproducibility at low flow-rates, the L-7100 pump is
n Optimally suited for narrow-bore
applications.
For applications involving sensitive materials requiring operation at low temperatures
n The cold room compatibility of the L-7100 is advantageous.
Premature wear caused by saltcontaining eluents is prevented by the fitting of a
n Piston cleaning facility, whereby the piston wash procedure can be carried out automatically in combination with the LaChrom autosamplers.
GLP Functions
In view of the requirements of quality assurance systems, all modules of the LaChrom system have been fitted with comprehensive facilities for status display, reliability monitoring and the provision of corresponding documentation and traceability.
n All parameters, programmes and gradient curves of the L-7100 pump can be printed out by the La Chrom data systems.
n The pressure signal of the
L-7100 pump is constantly monitored as part of the reliability check. Any deviations from the normal pressure profile, caused for example by air bubbles or malfunction of pump valves, are registered during the course of the analysis.
n The Pump pressure profile can be printed out in the analysis report and hence serve in the assessment of the quality of the analytical results.
Maintenance is facilitated by the
n Maintenance logbook which automatically registers the total volume delivered. This provides an indication as to when the next replacement of the pump seal or the next cleaning of the pump valves
should take place.
Controller Function
In stand-alone operation, the system control function of the
L-7100 enables external instruments to be controlled by providing
n Four event and other control signals.
Technical Data L-7100
Delivery system Series configured double piston pump with electronic compensation of solvent compressibility and elimination of residual pulsation
Maximum pressure 400 bar
Flow rate 0.001 9.999 ml/min
Flow stability ≤ 0,2% RSD
Gradient modes Quaternary low-pressure gradient, binary or ternary high-pressure gradient
Special functions Programme link, column pack programme, delayed pump stop,
system flush programme
GLP functions Pump logbook, display of pressure profile deviations, key-board lock
Documentation Print-out of pump parameters, gradient programmes, gradient curves, pump logbook via LaChrom data systems
Outputs 4 time-progammable event contacts, analogue pressure signal,
busy/ready , D-line
I nputs Start, stop, pump on, D-line
Materials Stainless steel, ruby, sapphire, ärconium oxide-ceramic, PTFE, carbon
L-7400 Programmable UV-Detektor
The detectors is equipped with a high-quality reference beam system which compensates for fluctuations in the light beam instability and ageing of the lamp. Newly developed deuterium lamps guarantee longer life and increase the already proven operational reliability of the detectors. The measuring cells are so designed that any refractive index effects brought about by alteration of chromatographic parameters are minimised. This results in the highest degree of detection sensitivity throughout the entire measuring range of the detectors.
Wavelength Programming and Spectral Recording
Detection at the absorption maximum of an analyte improves not only the detection sensitivity but also the linearity and reproducibility of an analysis. The wavelength-programmable detectors are capable of adjusting automatically to the optimal wavelength prior to each peak. This occurs so quickly that each component of a sample is detected at its maximum sensitivity. During wavelength switching, the autozero function can be activated; this prevents baseline steps occurring and increases reliability when integrating peaks. In order to enable a particular sample component to be specifically identified, a complete spectrum can be stored and printed out on a recorder or integrator when the chromatogram has been completed.
High Operational Reliability and Ease of Use
For GLP work, information on lamp energy, lamp light-on time and an automatic test of wavelength accuracy can be recalled. The easy-toread display and simple operation enables the detectors to be optimally set. Numerous programming possibilities are available for wavelength, baseline parameters and spectral files. A total of nine time programmes can be entered using a dialogue process. Once entered, programmes remain intact even after the instrument has been switched off.
Digital Communication and System Advantages
The new LaChrom detectors have been fitted with an analogue and a digital data output. They can be used either as stand-alone instruments or be incorporated into the LaChrom system. When combined with other LaChrom modules, digital data communication ensures maximum data security and interference-free data processing.
The new high-performance data communication network D-Line also facilitates the intelligent and interference-free transfer of digital information between the LaChrom modules as well as the control and display of all parameters, files and spectra via an integrator or PC. Functions such as the "event logbook", the "reliability report" and the automatically compiled service report facilitate GLP work and simplify both maintenance and checking of the instruments.
Technical Data
Optics Double beam system
Light source Deuterium lamp
UV cell path length 10 mm
Wavelength range 190600 nm
Wavelength accuracy ± 1 nm
Spectral band width 10 nm
Time constants Selectable, 0.1 s (FAST), 0.5 s (MED), 2.0 s (SLOW)
Noise ≤ 0.8 x 10 5 AU (250 nm)
Drift ≤ 3 x 10 4 AU/h (250 nm)
Detector range 0.001 to 2.5 AUFS
Time-programmable parameters Wavelength, auto zero, hold for baseline
Working parameters Time constants, recorder range, offset
Digital communication D-Line
Analogue output Data processor 1 V/AU, recorder 10 mV/ AU
Inputs Marker in, spectrum start in, start in,
lamp off in, autozero in
Diagnostic functions Lamp energy, lamp light-on time, lamp switching, wavelength accuracy
Optional accessories Semi-micro, micro and preparative flow cell