Leybold GT 2
| Objektnummer | B00015443 |
|---|---|
| ID-number | 015443 |
| Object name | Leybold GT 2 |
| Status | Stock unit |
Product group: Freeze drying systems
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
|
Country |
Possible payment methods |
Comment |
|
DE, AT, CH |
Payment by invoice, payment in advance, paymet by credit card |
Payment by invoice is only possible for corporate clients. |
|
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
|
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
|
Status |
Condition |
Comment |
|
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
| In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
|
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
|
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
|
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
manufacturer : Leybold
model : GT 2
annotation : Dokumente engl.
The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
Description of the basic plant GT 2
The GT 2 freeze drier is placed on a fable er a trolley. The vacuum pump is set on the fioor or on the lower sheft of the trolley (1) and is connected to the plant by PVC tubing (13).
The vacuum manifold or base plate (6) with the drying bell jar (8) is placed on the vacuum intake port. The various accessories see Technical Data and Seeäon 4 allow multiple use of the freeze drier.
The plant is evacuated with the two-stage vacuum pump (2). When the pump is switched off, the built-in intake port valve closes automatically, the vacuum in the plant is maintained, the vacuum pump itself is vented.
The ice condenser is cooled to the final temperature as indicated in the Technical Data by a one-stage refrigerator. its condensation area is designed as to take up large quantities of water at the beginning of the process and net to exceed ice layers higher than 1 to 2 cm at maximum load. Defrosting is by a radiation heater in the center.
The refrigerator for the ice condenser is air-cooled. lt operates with Freon R 22 and is equipped with a fully enclosed compressor (20). The liquid coolant is supplied from the air-cooled condenser (22) via a drying cartridge (23) to the expansion valve (24). The coolant evaporating behind the expansion valve (24) in the ice condenser (evaporator) (25) is sucked on by the compressor and returned to the condenser. The fan (21) supplies cold air to the condenser.
A shut-off valve is fitted between ice condenser and drying chamber er vacuum manifold respectively. lt allows pressure control and pressure rise measurements to be carried out in the drying chamber.
The electrica equipment of the freeze drier comprises of: power supply and control module with vacuum gauge and programme switch. The basic equipment Cat. No. 042 50 B 1 comprises in addition of the controller for electrically heated product shelves and of a temperature controller (9c).
The individual operational steps are solected by means of the programme switch (9a).
Programme switch positions:
"OFF": all assembly units switched off.
"CONDENSER PRE-COOLING": The refrigerator is switched an to pre-cool the condenser.
"DRYING": Refrigerator operating, vacuum pump and vacuum gauge are switched on. In freeze driers with electrically heated product shelves also the heater is switched on.
"DEFROSTING": Refrigerator, vacuum pump, vacuum gauge and heating of product shelves if any are switched off. The defroster is switched in.
Technical Data
Vacuum pump Typ GT 2
Filling of special oil N 62
min./max. I 0,5/0,7
Ice condenser
Final temperature °C <-45
Capacity, approx. kg 2
Defroster kW 0,3
Refrigerator
System with one stage Typ Danfoss-Kapsel
Compressor PW 11 x 22 T 2
Refrigerant Frigen R22
Expansion valve on:
Ice condenser Typ TCE 1 HXxG1 ,8
( evaporator )
Power requirement without
Heater, approx. kVA 1 ,2
Vacuum gauge Typ THERMOVAC TM 201
Dimensions of plant
(Height x width x depth) mm 570 x 560 x 300
Weight, approx. kg 80 (with pump)