Manufacturer: Labexchange
Model: ECS05
The following illustrations and descriptions refer to the device model and come from brochures.
They are not the scope of delivery. For the exact scope of delivery please refer to the offer text.
Technical specifications
ECDA2800 UV-VIS-PDA DETECTOR
Detector with a wavelength of 200-800nm âÂÂâÂÂand a noise level of ± 5 · 10-6 AU.
ECO2080 Column oven with heating and cooling
The Peltier heating / cooling column oven with large storage space can accommodate up to 3 columns with a length of up to 25 cm.
Overheating - The safety fuse prevents the destruction of temperature-sensitive columns. The temperature range is between 0 and 80 ° C.
CLARITY CHROMATOGRAPHY SOFTWARE
This system is fully controlled by the Clarity Chromatography Software, which is also part of the delivery.
The Clarity software meets the CFR21GLP requirements.
ECP2010 ANALYTICAL HPLC PUMP
This isocratic pump works with the ECB2004 gradient box as a quaternary gradient pump. The flow rate range is 0.01-10 ml / min and the pressure limit is 40 MPa.
The device software includes a new learning algorithm for pulsation suppression and many test and diagnostic functions.
ECB2004 GRADIENT BOX WITH DEGASSER
Box is suitable for handling liquids in ECOM gradient systems. It houses a container for solvent bottles, a built-in high-performance vacuum degasser and a four-way gradient valve.
Up to four solvent lines can be simultaneously degassed by one unit. Four-way gradient valve is controlled by the ECP2010 pump.
Autosampler L-3320
Automatic autosampler ensures excellent accuracy and linearity of sample injection.
Very low sample residues, achieved μl pick-up mode, zero sample loss. Automatic self-examination.
Rheodyne valve ensures reliable operation.
Detector ECDA2800
Wavelength range 200 - 800 nm
Spectral half width 10 nm
Adjustment accuracy ± 1 nm
Reproducibility ± 0.5 nm
Light source Deuterium + halogen lamps
Noise (test cell, 254 nm, TC 2s, 10 Hz) ± 5 · 10 -6 AU
Drift (test cell, 254 nm) 1 x 10 -4AU / h.
Time constant 10 - 10 000 ms
Scanning speed up to 100 Hz
Digital output 1 V / AU
Analogue output 1 4 x configurable
Wetted materials quartz glass, stainless steel, PEEK
Communication RS232, Ethernet (LAN)
Display, keyboard VFD 140x32 pixels, 10 buttons
Power supply 100-240V 50 / 60Hz 100VA
Dimensions (W x H x D) 280 x 135 x 498 mm (11.02 x 5.12 x 18.23 in)
Weight 9 kg
ECO 2080 column oven
|
Operating temperature range
|
0-80 ° C
|
|
temperature stability
|
± 0.2 ° C
|
|
Reproducibility of temperature
|
± 0.2 ° C
|
|
temperature accuracy
|
± 2 ° C
|
|
communication
|
RS232, Ethernet (LAN)
|
|
Display, keyboard
|
VFD 140x32 pixels, 10 pushbuttons
|
|
power supply
|
100-240 V 50/60 Hz 250 VA
|
|
Heated / cooled interior dimensions (W x H x D)
|
328 x 54 x 42 mm
|
|
Dimensions (W x H x D)
|
280 x 135 x 498 mm
|
|
mass
|
14 kg
|
Pump ECP2010
|
flow rate
|
0.01-10.00 ml / min.
|
|
pumping system
|
two pistons dia. 1/8 "connected in series
|
|
Maximum operating pressure
|
40 MPa (5800 PSI, 400 bar)
|
|
Accuracy of flow rate (1 ml / min, 12 MPa, H 2 O)
|
± 2%
|
|
Repeatability of the flow rate (1 ml / min, 12 MPa, H 2 O)
|
± 0.5%
|
|
Accuracy of pressure measurement
|
± 2%
|
|
Adjustable lower pressure limit
|
0.0-39.0 MPa
|
|
Adjustable upper pressure limit
|
1.0 - 40.0 MPa
|
|
Number of valves *
|
4 (A, B, C, D)
|
|
Adjustment of component concentrations *
|
0.0 - 100.0%
|
|
Wetted materials
|
Stainless steel, sapphire, KEL-F, gaskets **
|
|
communication
|
RS232, Ethernet (LAN), USB
|
|
Display, keyboard
|
VFD 140x32 pixels, 10 pushbuttons
|
|
power supply
|
100-240 V 50/60 Hz 100 VA
|
|
Dimensions (W x H x D)
|
280 x 135 x 498 mm
|
|
mass
|
10 kg
|
ECB2004 Gradient Box with degasser
|
degasser
|
|
|
Degasser channels
|
5 independent
|
|
degassing
|
Gas permeation through a fluoropolymer membrane
|
|
Maximum flow rate
|
10 ml / min.
|
|
degassing
|
~ 2 ppm at 1 ml / min.
|
|
Dead volume
|
~ 480 μl per channel
|
|
Wetted materials
|
PEEK, glass fiber reinforced PTFE, Teflon AF®
|
|
Gradientenventil
|
|
|
Number of gradient valves
|
4 (A, B, C, D)
|
|
Maximum operating pressure
|
1.38 bar (20 PSI)
|
|
opening
|
0.062 "(1.6 mm)
|
|
Wetted materials
|
PTFE
|
|
Common parameters
|
|
|
connecting thread
|
1/4 "-28 flat bottom
|
|
power supply
|
100-240 V 50/60 Hz 20 VA
|
|
Within the box Dimensions (W x H x D)
|
267 x 56 x 305 mm
|
|
Dimensions (W x H x D)
|
280 x 135 x 498 mm
|
|
mass
|
7 kg
|
Autosampler L-3320
Sample Capacity 2 standard 48-vial vials (1.5ml vial)
Loop volume standard 100μl; 10μl, 20μl and 50μl optional
Syringe volume standard 250μl; 100μl optional
Reproducibility Full Loop Injection: RSD≤0.3%
Partial loopfill Injection: RSD≤0.5% μl
Pickup Injection: RSD≤1%
Cross-contamination <0.02% after programmed needle wash
Linearity 0.9999
Pressure limit 6000 PSI
Communication RS232, USB
Control PC (Clarity SW)
Power supply 100 - 240VAC, 50 - 60 Hz, 110W
Dimensions (L x W x H) 453 x 359 x 244 mm