manufacturer: Labexchange
model: ECS03
The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
Technical specifications
ECD2600 UV-VIS Detector
Detector with continuously variable wavelength in the range of 190-600 nm and a noise level ±3·10-6 AU.
Unit offers high-speed sampling up to 100 Hz and wavelength autocalibration. AC 05 analytical cell is
CLARITY Chromatography Software
This system is fully controlled by Clarity Chromatography software, which is also a part of delivery. Clarity software complies CFR21GLP requirements.
ECP2010 Analytical HPLC Pump
This isocratic pump works with Gradient Box ECB2004 as a quaternary gradient pump. Flow rate range is 0.01-10 ml/min and pressure limit at 40 MPa.
Unit software includes new learning algorithm for pulsation suppression and many testing and diagnostic functions
ECB2006 Gradient Box with Degasser
Box is suited for secure mobile phase bottles installation onto isocratic systems
Detector ECD2600 ECD2800
Wavelength range 190-600 nm 190 - 800 nm
Spectral half width 6 nm
Adjustment accuracy ± 1 nm
Reproducibility ± 0.5 nm
Light source Deuterium lamp Deuterium + halogen lamps
Noise (test cell, 254 nm, TC 2s, 10 Hz) ± 3 · 10 -6 AU
Drift (test cell, 254 nm) 1 x 10 & sup4; AU / h.
Time constant 20 - 10 000 ms
Scanning speed up to 100 Hz
Digital output 1 V / AU
Analog output 1 1x configurable
Wetted materials are dependent
from one cell - quartz glass, PTFE, stainless steel, PEEK
Communication RS232, Ethernet (LAN)
Display, keyboard VFD 140x32 pixels, 10 buttons
Power supply 100-240V 50 / 60Hz 100VA
Dimensions (W x H x D) 280 x 135 x 498 mm (11.02 x 5.12 x 18.23 in)
Weight 9 kg
Pump ECP2010
|
flow rate
|
0.01-10.00 ml / min.
|
|
pumping system
|
two pistons dia. 1/8 "connected in series
|
|
Maximum operating pressure
|
40 MPa (5800 PSI, 400 bar)
|
|
Accuracy of flow rate (1 ml / min, 12 MPa, H 2 O)
|
± 2%
|
|
Repeatability of the flow rate (1 ml / min, 12 MPa, H 2 O)
|
± 0.5%
|
|
Accuracy of pressure measurement
|
± 2%
|
|
Adjustable lower pressure limit
|
0.0-39.0 MPa
|
|
Adjustable upper pressure limit
|
1.0 - 40.0 MPa
|
|
Number of valves *
|
4 (A, B, C, D)
|
|
Setting of component concentrations *
|
0.0 - 100.0%
|
|
Wetted materials
|
Stainless steel, sapphire, KEL-F, gaskets **
|
|
communication
|
RS232, Ethernet (LAN), USB
|
|
Display, keyboard
|
VFD 140x32 pixels, 10 pushbuttons
|
|
power supply
|
100-240 V 50/60 Hz 100 VA
|
|
Dimensions (W x H x D)
|
280 x 135 x 498 mm
|
|
Weight
|
10 kg
|