Dohrmann DX 2001
Objektnummer | B00006781 |
---|---|
ID-number | 006781 |
Object name | Dohrmann DX 2001 |
Status | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance, paymet by credit card |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
DX 2000
The DX-2001 uses a microcoulometric detector which consists of an electrochemical cell and associated coulometer circuitry. The cell has four electrodes immersed in an electrolyte which is 70% acetic acid in water. Three of the electrodes are silver with a silver chloride coating. The fourth electrode is platinum. The electronic circuitry to sense the cell voltage and deliver the appropriate amount of current is located on a circuit board in the furnace module. The sensor and reference electrode pair is used to measure the cell voltage. The generator and auxiliary electrode pair work together to add or remove Ag + ions from the electrolyte solution.
Specifications |
Characteristics |
||
Performance: |
Range |
Precision 1 |
Sample Size |
TOX |
4 -1000 µg TOX/L |
±2 µg/L or ±2% 2 |
100 mL |
|
40 -10,000 µg TOX/L |
±20 µg/L or ±2% 2 |
10 mL |
AOX |
4 -1000 µg AOX/L |
±2 µg/L or ±2% 2 |
100 mL |
EOX/TX |
0.5 - 50,000 ng CI/µL |
±2 µg CI or ±2% 2 |
2-30 µL |
System |
15 -100,000 ng CI |
±6 µg CI or ±2% 2 |
n/a |
Analysis Time |
Combustion and Detection Rate: 3-10 min/injection |
||
|
Adsorption Rate: 0.5-5 mL/min (with AD-2000) |
||
|
Use 5 mL/min onlv with 3mm ID columns |
||
|
Fill/Flush Rate: 33-66 mL/min (with AD-2000) |
||
Combustion Temp. |
25 to 1100°C |
||
Titration Cell Electrodes |
3 solid silver and 1 platinum Independent reference electrode |
||
Power |
90-130 VAC, 50/60 Hz or 180-260 VAC, 50/60 Hz |
||
Gas |
C0 2 or He or Ar; 99%, 45 PSIG, 3.1 bar and 0 2 : 99%, 30 PSIG, 2.1 bar |
||
Dimensions 3 |
Furnace Module (including Cell) |
||
|
61 cm (W), 54 cm (D), 41 cm (H) |
||
|
24" (W), 21" (D), 16 (H) |
||
|
Boat Module |
||
|
51 cm (W), 18 cm (D), 21 cm (H) |
||
|
20" (W), 7" (D), 8" (H) |
||
|
Auto lnjector (optional) |
||
|
23 cm (W),13 cm (D), 51 cm (H) |
||
|
9" (W), 5 (D), 20" (H) |
||
Weight |
Furnace Module (including cell): 19 kg (42.5 Ibs) |
||
|
Boat Module: 5 kg (12 Ibs) |
||
|
Auto Injector (option): 3 kg (6 Ibs) |
Notes: 1 One standard deviation, instrument contribution
2 Whichever is greater
3 Allow 3 clearance at rear of system for system cables and plumbing