Dionex DX-500
Objektnummer | B00019029 |
---|---|
ID-number | 019029 |
Object name | Dionex DX-500 |
Status | Archiv |
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
For deliveries to countries outside the EU, an additional fee of 70,00€ for export is charged. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by credit card, letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
Country |
Possible payment methods |
Comment |
DE, AT, CH |
Payment by invoice, payment in advance |
Payment by invoice is only possible for corporate clients. |
NL, BE, LU |
Payment by invoice, payment in advance |
Payment by invoice is only possible for corporate clients |
Other countries |
Payment in advance |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
Status |
Condition |
Comment |
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
Manufacturer: Dionex
Model: DX 500
The following illustrations and descriptions refer to the instrument model and are drawn from brochures. They do not represent the scope of delivery. Please refer to the text of the offer for the exact scope of delivery.
Specifications
Signals
Up to nine, each defined by a sample wavelength and bandwidth.
Noise
< ±25 }SAU, 250 nm 4-nm bandwidth
Drift
< 1 mAU/hr, 250 nm, empty cell
Wavelength Range 195 Â 600 nm
Wavelength Accuracy ± 1 nm
Resolution
1 nm Digital
Spectral Acquisition
Selectable, 240, 320, 640, 1280, 2000 ms
Light Source
Deuterium; 1000 hrs
Flow Cells
Stainless Steel: 8-µL volume, 10-mm path, 5 MPa (750 psi) max
Titanium (optional): 8-µL volume, 10-mm path, 5 MPa (750 psi) max
Prep (optional): variable volume and path, 2 MPa (300 psi) max: 0.8-µL volume, 0.1-mm path 1.6-µL volume, 0.2-mm path 4.0-µL volume, 0.5-mm path
High pressure (optional): 3-µL volume, 6-mm path, 40 MPa (6000 psi)
Diode Array Thermostat
3-stage temperature control:
Off, Low (37 °C), High (47 °C) Important for dark-current control
Operating Temperature Range
10Â35°C
Dimensions (h x w x d)
20 cm (7.9 in.) x 26 cm (10.2 in.) x 52 cm (20.5 in.)
Weight
18 kg (401b)
Power Requirements
120 V ac, 2 A, 60 Hz;
or 220 V ac, 1 A, 50 Hz;
or 240 V ac, 1A, 50 Hz
Autosampler AS 40
Specifications
Sample Delivery Method:
Displacement
Capacity:
11 cassettes of either six 5-mL or eight 0.5-mL PolyVials
Vial Size:
Nominal volume of 0.5 or 5 mL
Filter Pore Size:
20 µm
Minimum Volume Delivered:
0.17 mL from 0.5-mL vials; 1.7 mL from 5-mL vials
External Control:
Relay closure
Injections Per Vial:
1, 2 or 3 samples may be delivered from each vial
Type of Loading:
Loop: Delivers sample rapidly to sample loop
Concentrator: Delivers sample against back pressure of up to 100 psi (690 kPa) at approximately 0.4 mL/min (0.5-mL vials) or 1 mL/min (5-mL vials)
Mode of Sampling:
Proportional: delivers volume inversely proportional to number of injections per vial
Constant: Delivers fixed volume independent of the the number of injections per vial
Bleed:
On: Displaces excess sample and
any trapped air through bypass valve before sampling
Off: Delivers sample and any trapped air directly to injection valve
Dimension, (H x W x D):
11.3 x 14.3 x 17.5 in.
(28.7 x 36.3 x 44.5 cm)
Amperometry Electronics
Noise: 1 pA (DC amperometry); 20 pC (integrated amperometry)
Potential Ranges: ±2.00 V in 0.01-V increments.
Control Modes: Local or remote via relay closures, TTL, or by PeakNet Software via DX LAN.
Amperometry Flow Cell
Electrodes:
Interchangeable working electrodegold, platinum, silver, or glassy carbon. Counter electrodeÂtitanium.
Reference electrodeÂpH-Ag/AgCI combination.
Wetted Materials: PEEK, Titanium, Kel-F, polyetherimide, EPR, glass, high density polyethylene (HDPE), and working electrode material (gold, platinum, silver, and glassy carbon).
Cell Volume at Working Electrode:
Maximum Operating Pressure: 0.7 MPa (100 psi).
Output Control
Filter: 0 to 10 s response time.
Full-Scale Analog Output Ranges: Conductivity-0.01 gS to 3000 gS Integrated amperometryÂ50 pC to 200 gC. DC amperometryÂ50 pA to 300 p.A.
Offset: To full-scale for each range.
Analog Signal Output Voltage: 0.01, 0.1, or 1 V.