HPLC-Systems

Perkin Elmer 200

€7,180.00
Excl. 19% VAT , excl. Shipping Cost
€8,544.20 incl. Vat
ID-number: 37639
Status: Stock unit?The device is in our stock, however, has to be inspected by our technician, in order to be able to guarantee complete function.
Shipping: 10-14 working days
Description:
Perkin Elmer HPLC 200 Series. Consisting of:
Autosampler Series 225.
Degasser Series 200.
Series 200 pump.
Antek Electrochemical Detector Decade II.
PE NCI 900.
230 V. 50/60 Hz. 580 W.
Installation
We execute the installation of this laboratory device professionally. We would be pleased to offer you the installation of this laboratory device for 790€, excluding travel expenses.

Usually, laboratory and analysis devices are very exigent system with a very own specifications and requirements. Therefore, the startup of a laboratory device often rises the question how to assemble, adjust and to launch it.


We would be pleased to take over the professional installation of your laboratory devices!

Please contact us.


Benefit from the installation by us in more than one way:

  • Immediate and smooth startup of your laboratory device
  • A professional installation extends the lifetime of your laboratory device
  • Guarantee extends to 3 months
  • Freight charges do not apply*
  • You are entitled to benefit from the Labexchange Repurchase Guarantee.

* products of the product groups laboratory furniture and laminar flow are excluded. In individual cases, other product groups can be excluded.


Add to Offer List
More Information
Objektnummer B00037639
ID-number 37639
Object name Perkin Elmer 200
Status Lagergerät
Manufacture Perkin Elmer
Shipping 10-14 working days

Status, terms of delivery and payment

Verification of devices

The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.

Dispatch time

The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.

As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.

 

Shipping methods

Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.

 

Delivery information

Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.

For deliveries to countries outside the EU, an additional fee of 70,00€ for export is charged. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.

 

Terms of payment

We do not accept payment by credit card, letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.

Country

Possible payment methods

Comment

DE, AT, CH

Payment by invoice, payment in advance

Payment by invoice is only possible for corporate clients.

NL, BE, LU

Payment by invoice, payment in advance

Payment by invoice is only possible for corporate clients

Other countries

Payment in advance

 

 

Our General Terms of Sale, Delivery and Payment are valid and are available for download here. 

The goods are offered subject to prior sale.

 

Definition of status

All articles are used articles, except an article is listed especially as a new device.

Status

Condition

Comment

Immediately available

Used The article is fully functional and in impeccable condition. It can be shipped immediately.
In stock

Used

The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article.

Published

Used

The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery.

New device

new

The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply.

Firma:              Perkin Elmer

Modell:            200

Kommentar:    Dokumente engl.

The following illustrations and descriptions are referring to the instrument model and are drawn from brochures. They are not representating the delivery volume. The exact delivery content you will find only in the offering text.

Series 200 LC-Pump

The series 200 LC Pump

The Perkin-Elmer Series 200 LC Pump can be used with virtually every LC system or appli­cation. The Series 200 LC Pump is upgradable from a single solvent (isocratic) pump, to mix­ing two solvents (binary) or to mixing four solvents (quaternary). In addition, each version is available in a biocompatable version. The Series 200 LC Pump provides you with the inter­changeable technology so that you do not have to buy a new pump and learn new user inter­faces every time your needs change.

Series 200 LC Pump Features and Benefits

•     Upgradable isocratic, binary, and quaternary versions as well as biocompatible versions for maximum flexibility.

•     Small, compact design (12 inches (30.5 cm) wide) saves valuable bench space.

•     Directory of up to 20 methods, 20 steps each with built-in method linking increases sam­ple throughput.

•     Edit the active method or sequence in real-time with easy access function keys.

•     Large backlit liquid crystal display allows clear viewing of pump method and operating parameters at all times.

•     Unique function allows automatic shutdown of the pump method at final gradient method.

•     Graphic display of solvent profile on-screen shows real-time status of the gradient method.

•     Hold gradients or advance to the next step with a simple keystroke.

•     Solvent-resistant front panel with rounded ergonomic design for easier interaction with the keypad.

•     Tactile keyboard with unique "key click" so that you know your entry has been made.

•     Phone-type contact closure connections and finger tight plumbing Fittings make it simple to connect or disconnect your system.

•     Full method documentation including modification with method name, number, date and time stamp.

•     Visual access front-door panel allows quick plumbing inspection and easy servicing.

•     Maintenance log coded by the serial number of the pump, including pump and seal cycles.

•     Instrument log records history of operation for quick diagnosis and servicing to get you up and running quickly.

•     Extensive service diagnostics.

•     Pump module can be completely removed to lab bench for maintenance without having to unstack modules.

Specification

General Specifications

Flow Rate Range

0.01 to 10.0 mL/min (maximum pressure of 6 200 psi (420 bar) over the entire range

Flow Rate Increments

0.01 from 0 to 0.99 mL/min and 0.1 from 1.0 to 10.0 mL/min

Flow Precision

0.3% relative standard deviation (typically <0.1%) at 1 mL/min and 1 000 psi (70) bar with water as the mobile phase

Flow Accuracy

±1% of the setting at 1 mL/min and 1 000 psi (70 bar) with water as the mobile phase

Retention Time Reproducibility

<0.3% RSD (typically <0.1% RSD)

Pressure Range

0 to 6 200 psi or 0 to 420 bar from the entire flow-rate range

Pressure Limits

High and low limit user-settable to 10 psi or 1 bar increments

Solvent Compressibility

Keyboard settable flow calibration

Composition Range

0 to 100% solvent A - D (depending on your pump configuration)

Composition Settability

Settable to 0.1%

Composition Accuracy

±1.0% absolute from 3 to 97% at 5 mL/min

Composition Precision

Typically less than 0.2% variation

Gradient Curves

Linear, exponential 1 to 9.9 (positive and negative), or step (89 profile selections)

Linearity

Within ±1% from 10 to 90% at 5 mL/min

Program Steps

Step 0 plus up to 20 solvent program steps. Battery-backed storage of up to 20 methods plus one default method.

Step Time

0 to 999 min

Step Time Increments

0.1 min, 0 to 9.9 min; 1.0 min, 10 to 99 min

Programmed Timed Events

Two built-in independent momentary (two second) contact closures

Events/Methods

Maximum of 20

Event Increments

0.1 min, 0 to 9.9 min; 1.0 min, 10 to 99 min

Dimensions (HxWxD)

14 cm x 30.5 cm x 43 cm

Weight

15.9 kg (35 lbs.)

Operating Specifications

Operating Conditions:

Environment

This instrument is designed for indoor use. It is not designed to operate in an explosive environment.

Ambient temperature

10 - 35°C

Ambient relative humidity

20 - 80% (without condensation)

Altitude

0 to 2000 meters

Installation Category

This instrument is able to withstand transient overvoltage according to Installation Category II as defined in IEC 1010-1.

Pollution Degree

This instrument will operate safely in environments that contain nonconductive foreign matter to Pollution Degree 2 in IEC 1010-1.

Storage Conditions:

Ambient temperature

-20 to 60°C (-4 to 140°F)

Ambient relative humidity

20 - 80% (without condensation)

Altitude

0 to 12,000 meters

Electrical Specifications

AC Power Requirements:

AC power

70 Watts

AC voltage range

Grounded 100, 120, 220, 230, VAC (±10%)

Grounded 240 VAC (+6%/-10%)

Frequency

50/60 Hz (±1%)

Fuse rating characteristics:

100/120 VAC Single Fuse

1.6 A 3 AG Slow Blow, 250 V

220-240 VAC Single/Dual Fuse

0.8 A TT (5x20 mm), 250 V


Series 200 Autosampler

Figure 1-1. The Series 200 LC Autosampler.

The Perkin-Elmer Series 200 LC Autosampler is an advanced sample processor that can be used with virtually every liquid chromatography system or application. You configure the Autosampler and define all of your methods and operations using function keys on a key­board located on the front panel of the instrument. A backlit Liquid Crystal Display (LCD) located directly above the keyboard shows both status information about the instrument and provides access to the screens you use to define methods and set configuration parameters. If you use the Perkin-Elmer Series 200 LC Pump, you will find that the Autosampler front panel is almost identical to the one on the Pump. In addition, the Pump and the Autosam­pler share many common operating features that will allow you to get your system up and running quickly. The Series 200 LC Autosampler provides you with interchangeable technol­ogy so that you do not have to buy a new Autosampler and learn new user interfaces every time your needs change.

Series 200 LC Autosampler Features and Benefits

•     Upgradable to biocompatible versions for maximum flexibility.

•     Small, compact design (12.5 inches (31.8 cm) wide) saves valuable bench space.

•     Directory of up to 20 methods.

•     Ability to link up to 10 methods together in a sequence.

•     Priority method which allows you to interrupt a method or sequence to run critical sam­ples. The method or sequence resumes once the priority samples are analyzed.

•     Edit the active method or sequence in real-time with easy access function keys.

•     Large backlit liquid crystal display allows clear viewing of autosampler method and operating parameters at all times.

•     Tactile keyboard with unique "key click" so that you know your entry has been made.

•     Phone type contact closure connections and finger tight plumbing fittings make it simple to connect or disconnect your system.

•     Full method documentation including modification with method name, number, date and time stamp.

•     Visual access front-door panel allows quick plumbing inspection and easy servicing.

•     Maintenance log coded by the serial number of the autosampler, including autosampler injection cycles.

•     Instrument log records history of operation for quick diagnosis and servicing to get you up and running quickly.

•     Extensive service diagnostics.

•     Sampler module can be completely removed to lab bench for maintenance without having to unstack modules.

Specification

Injection range

Programmable from 0.1 - 1.0 µL at 0.1 µL increments, 1 µL to 2.5mL at 1µL increments

Replicate injections

1 - 99 from one vial

Precision

Typically <0.5% RSD of peak areas from 3 µL

Linerarity

>0.999

Minimum sample volume

1 µL in 7 µL

Sample syringe sizes

50, 100, 250, 500, 1000, 2500 µL

Sample waste

None (or user selectable)

Carryover

<0.02%

Flush cycles

User selectable, 1 - 9

Flush volume

0 - 2500 µL

Sample capacity

up to 225 vials with optional tray

Method parameters

Injection volume, sample draw speed, flush volume, flush cycle, 2 external contacts, start and stop time

Methods

Battery-backed storage of up to 20 methods including time programming. Editing of stored methods possible during runs.

Injection cycle time

2 - 3 samples per minute

Valve materials in contact with

sample

With Rheodyne 7725 for Analytical - stainless steel, Vespel, ceramic, PEEK

With Rheodyne 9725 for Bio Applications - PEEK, Tefzel,

ceramic

With Rheodyne 8125 for Microbore - stainless steel, Vespel, ceramic, PEEK

Display

7-line by 40-character backlit liquid crystal display

Communications

Integrated communications for derivatization and dilution options

Serial communications - standard

BCD interface - standard

5 relay outputs; 2 timed events, 2 inject, 1 start

AC Power Requirements:

AC power

70 Watts

Specification

AC voltage range

Grounded 100, 120, 220, 230, VAC (±10%)

Grounded 240 VAC (+6%/-10%)

Frequency

50/60 Hz (±1%)

Fuse rating characteristics:

100/120 VAC Single Fuse

1.6 A 3 AG Slow Blow, 250 V

220-240 VAC Single/Dual Fuse

0.8 A TT (5x20 mm), 250 V

Operating Conditions:

Environment

This instrument is designed for indoor use. It is not designed to operate in an explosive environment.

Ambient temperature

10 - 30°C

Ambient relative humidity

10 - 80% (without condensation)

Altitude

0 to 2000 meters

Installation Category

This instrument is able to withstand transient overvoltage according to Installation Category II as defined in IEC 1010-1.

Pollution Degree

This instrument will operate safely in environments that contain nonconductive foreign matter to Pollution Degree 2 in IEC 1010-1.

Peltier sample cooling tempera-

ture range

4 - 50°C, ± 2°C

Storage Conditions:

Ambient temperature

-20 to 60°C (-4 to 140°F)

Ambient relative humidity

10 - 80% (without condensation)

Altitude

0 to 12,000 meters

Dimensions

Height: 400 mm (15.8 in)

Width: 318 mm (12.5 in)

Depth: 483 mm (19 in)

Weight

21 kg (461bs)


DAD-Detector LC-235C

The LC-135C and LC-235C multiple wavelength diode array LC detectors contain the following features:

•     Individual signal amplification of each pixel for a significantly improved signal-to-noise ratio.

•     Automatically determines peak purity by an absorbance index technique, for dependable data, even on trace levels of sample.

• A dual beam design using two 35-element diode arrays pro­vides excellent baseline stability at all wavelengths and all sol­vent systems.

•     History File reports (records of detector conditions for the run) are printed at the end of each run.

•     Automatically annotates the wavelength of maximum absorbance on each peak.

•     Biocompatible solvent path.

•     Instantaneous wavelength change.

•     All wavelengths monitored at the same time.

•     Computerized diagnostics.

     In addition, the LC-235C Diode Array Detector contains the following features:

•     Produces instantaneous spectra.

•     Plots the spectra.

•     Overlays the spectra.

•     Normalizes the spectra.

•     Constructs an Absorbance Profile Map.

•     Contains the option of selectable modes of spectral acquisition.

•     Produces automated background correction of spectra.

Specifications

Optical System                                 Dual beam polychromator.

Source                                              Deuterium lamp.

Spectral Bandwidth                          5 to 175 nm.

Wavelength Range                           195 to 365 nm.

Wavelength Accuracy                      < 1 nm.

Noise                                                 < 5 x 10 -5 AU P/P at 240 nm, and 15 nm bandwidth at          peak width setting of 16 sec.

Drift                                                   < 2 x 10 -4 AU/hour at 240 nm, after warmup.

Absorbance Range                           -0.3 to 1.0 AU.

(on the optional plotter)

Linear Absorbance Range                ≤1% deviation to 1.5 AU.

Flowcell                                             Standard 8-µL volume, 10-mm pathlength, titanium. Dispersion:      20 µL, four sigma. Built-in holmium oxide filter.

Flowcell Pressure Rating                 2200 psi

Analog Signal Outputs                      Two outputs, for wavelengths A and B,

(integrator)                                        with software-selectable output voltage and filtering.

Number of Methods                          Nine user-programmable methods plus Method 0.    Protected memory.

                                                         Nine steps per method.

Chromatograms and Spectra*         Printer/plotter output.

Spectral Storage*                             1000

Spectral Library*                               21 in protected memory. Dimensions:

Height                                                28 cm (11 in.)

Width                                                36 cm (14 in.)

Depth                                                46 cm (18 in.)

Weight                                              25 kg (55 lb.)

Electrical Requirements                   120/220 V, 50/60 Hz

Power Consumption                         100 watts

Product group: HPLC

Back to Top
part-of-labexchange-group
Search engine powered by ElasticSuite
lazyOwllazyOwl
Labprocure