Shimadzu AA 6200
| Objektnummer | B00007890 |
|---|---|
| ID-number | 007890 |
| Object name | Shimadzu AA 6200 |
| Status | Stock unit |
Product group: Atomic absorption spectrometers (flame)
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
|
Country |
Possible payment methods |
Comment |
|
DE, AT, CH |
Payment by invoice, payment in advance, paymet by credit card |
Payment by invoice is only possible for corporate clients. |
|
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
|
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
|
Status |
Condition |
Comment |
|
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
| In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
|
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
|
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
|
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
Nachfolgende Abbildungen und Beschreibungen sind modellbezogen und aus Prospekten entnommen.
Sie geben nicht den Lieferumfang des Systems wieder.
Den exakten Lieferumfang entnehmen Sie bitte aus dem Angebotstext.
The Shimadzu Atomic Absorption Spectrophotometer AA-6200 is used in the flame atomic absorption analysis and flame emission analysis. It is compact and can be operated easily, provided with double beam optical system and D2 (deuterium lamp) background correction method.
Software for control of the AA-6200 is operated in MS-Windows TM environment, which allows to create measurement conditions easily even by person who first time operates the atomic absorption spectrophotometer, using the parameter setting Wizard.
When the AA-6200 is combined with the auto sampler ASC-6100, automatic analysis becomes possible for every elements . This function is useful to save energy.
Specifications
Hardware |
|
|
Optics |
Double Beam (chopper mirror) |
|
Monochromator |
aberration corrected Czerny-Turner monochromator Holographic grating (1,600 lines/mm) |
|
Wavelength range |
190-900nm Automated wavelength selection |
|
Slit |
0.2nm~0.7nm Manual setting |
|
Background Correction |
D 2 -Lamp method |
|
Lamp Tunet |
2-lamps simultaneously lit (manual turrent) |
|
Lamp Mode |
EMISSION, NON-BGC, BGC-D 2 |
|
Frequency |
100Hz |
|
Nebulizer |
Nebulizer integrated impact bead and jacket tip Pt/Ir capillary |
|
Chamber |
Polypropylene |
|
Burner |
Fixed back/forward position and burner height (Simply switching of Air/C 2 H 2 and N 2 0/C 2 H 2 burner) Titanium 10 cm slot burner (Optional high temperature burner) |
|
Gas Control |
Manual setting of flow rate Automatic Air/N 2 0 switching system |
|
Safety |
Gas pressure monitoring to prevent flashback Automatic flame monitoring Safety interlock for misuse burner Automatic flame extinghish when power failure |
|
Ignition |
Push ignite button |
|
Software |
Software based an MS-Windows TM |
|
Dimension and weight |
W690 x D425 x H370mm, 38kg |
|
Power requirements |
AC220V, 230V ± 10%, 50/60Hz, 300VA (Certification of CE marking) |
|
Ambient temperature and humidity range |
Temperature : 10~35°C Humidity : 45~80% (but less than 70% if temperature is over 30°C) |
|
Software |
|
|
Concentration conversion |
Calibration method, standard addition method, simplified standard addition method |
|
Measurement type |
AUTOZERO, BLK, STD, UNK, MSA, SMAS, WAIT, PAUSE, FILESAVE determines measurement type by designating above actions |
|
Data processing |
Linear regression analysis by least square method, Abs=f (Conc.) |
|
Calibration curve |
Dimension : first order, second order, third order, ON/OFF of zero intercept selectable calibration curve by Curve ID |
|
Repetitive analysis |
Maximum 10 analysis per sample |
|
Accuracy enhancement |
Usage of average value, cancel of abnormal data by the limitation of CV or SD values. |
|
Sensitivity enhancement |
Automatic calculation by sensitivity correction |
|
Actual concentration calcuation |
Automatic calculation of actual concentration by sample volume, constant volume, and dilution factor upon data acquisition |
|
Signal process |
|
|
Measurement time |
Pre-spray: 99 seconds, Accumulation time: 99 seconds |
|
Response |
4 response time selectable |
|
Measurement procedure setting |
Easy setting by Wizard function |
|
Data display MRT Worksheet |
Measurement procedure, analysis results, actual concentration, calcuation factor of actual concentration, combined display of data and time |
|
Signal display |
display of multiple elements by tab sheet Signal profile display area : 4 Current signal profile display area : 1 Overlay display of signal profiles by designating the row number of MRT |
|
Retrieval and storage of data |
Automatic loading of template parameter file upon software activation Automatic data storage by "FILESAVE" command Raw profile data can be stored |
|
Import/Export of data |
Import sample name from ASCII file Export data file by ASCII file |
|
Report |
Summary report : display and output of results by sample Output after analysis : Parameter, Raw data, signal, sequence Selectable format and items for out put |