Abbott Cell-Dyn 3500SL
| Objektnummer | B00004257 |
|---|---|
| ID-number | 004257 |
| Object name | Abbott Cell-Dyn 3500SL |
| Status | Stock unit |
Product group: Hematology
Status, terms of delivery and payment
Verification of devices
The second-hand devices are verified by Labexchange Service GmbH before delivery. You are receiving only fully functional devices.
Dispatch time
The stated dispatch times are the shortest possible ones for each article. The effective dispatch times can vary. The effective dispatch times will be stated in the order confirmation.
As a matter of principle, we are offering collective deliveries. The shipping time is calculated based on the position with the longest lead time. A partial delivery is possible on explicit request.
Shipping methods
Parcel services, forwarding agencies, self-pickup, delivery by Labexchange fleet.
Delivery information
Prices exclude shipping costs. Stated shipping costs are to be expected. Deviations are possible. If transport costs are not specified, please ask separately for them.
The stated transport and packing charges apply to the most favorable route if transport and are to be understood as subject to verification due to unexpected cost increases. By reason of unpredictable events, cargo rates and delivery times can change at any time and therefore have to be adapted to the recent situation. Import formalities and possible customs charges will be borne by the purchaser. Incoterm coding according to Incoterms 2010: For persons who collect the devices themselves: EXW, for dipatch by sea: CFR, by air freight: CPT, other shipments: DAP. Note for international shipments: A proof of preference/EUR1 will not be issued by us. When self-collecting/ordering EXW from countries within or outside the European Union, 16% VAT will be retained as a deposit until we have received the corresponding confirmation of arrival/bill of delivery from the buyer.
Terms of payment
We do not accept payment by letter of credit, PayPal, etc. In each case the invoice amount is payable without deduction. Discount is not granted.
|
Country |
Possible payment methods |
Comment |
|
DE, AT, CH |
Payment by invoice, payment in advance, paymet by credit card |
Payment by invoice is only possible for corporate clients. |
|
NL, BE, LU |
Payment by invoice, payment in advance, payment by credit card |
Payment by invoice is only possible for corporate clients |
|
Other countries |
Payment in advance, payment by credit card |
|
Our General Terms of Sale, Delivery and Payment are valid and are available for download here.
The goods are offered subject to prior sale.
Definition of status
All articles are used articles, except an article is listed especially as a new device.
|
Status |
Condition |
Comment |
|
Immediately available |
Used | The article is fully functional and in impeccable condition. It can be shipped immediately. |
| In stock |
Used |
The article is on stock. Our service technicians will verify the article before delivery. You are receiving only a fully functional article. |
|
Published |
Used |
The article is still with the provider. After your order the article will be purchased and verified by us before being shipped to you. A certificate of operativeness as well as a service report are included in delivery. |
|
New device |
new |
The article is brand new and unused. Regarding new equipment the guarantee/warranty conditions of the corresponding manufacturer apply. |
|
Labprocure |
Used |
Labprocure GmbH, as the advertiser, is responsible for the content of this device offer. Labprocure assumes liability for the offers advertised here and for the photos and offer texts included. Labprocure GmbH, Bruckstraße 58, 72393 Burladingen. |
Manufacturer: Abbott
Model: Cell-Dyn 3500SL
The following illustrations and descriptions are model-specific and taken from brochures.
They do not reflect the scope of delivery of the system.
The exact scope of delivery can be found in the offer text.
Specifications
|
Analyzer |
Data station |
Ticket printer |
Graphics printer |
Sampler |
|
|
height |
61 cm |
61 cm |
15 cm |
22 cm |
42 cm |
|
broad |
76 cm |
38 cm |
41 cm |
65 cm |
71 cm |
|
depth |
56 cm |
56 cm |
39 cm |
46 cm |
29 cm |
|
Weight |
86 kg |
33 kg |
7.5 kg |
18.2 kg |
24 kg |
Environmental conditions
Temperature: Patient samples: room temperature (approx. 23 ° C)
Device: 25 ± 10 ° C
Relative humidity: 45% to 85%, RHNC
1. If the temperature is too low or too high, the result is different monocyte values. At a lower temperature (<18 ° C) the monocyte concentration is approx. 1.5% lower, at higher temperatures (> 32 ° C) it is approx. 6% higher.
2. The blank values for RBC and PLT can increase at lower ambient temperatures (<18 ° C).
Readiness to measure and total duration of the analysis cycles
Autostart (from standby): approx. 3.5 minutes
Autostart (when switched on): approx. 5 minutes *
Sampling from open
Sample tube: 37 seconds (from READY to READY)
Sampling from closed
Sample tube with sampler: approx. 37 seconds
Switching to standby: 4.5 minutes
* The laser requires a warm-up time of 15 minutes. If the power supply is interrupted for more than 5 minutes, the analysis can only be resumed 10 minutes after the end of the autostart cycle.
Sample volume drawn in (whole blood, reference values)
Sampling from open
Sample tube: 130 µl
Sampling from closed
Sample tube with sampler: 355 µl
Series size
1-100 tubes can be provided, tubes can be continuously recharged during operation
Throughput
Maximum throughput = 90 samples / hour
NOTE: Maximum throughput will be achieved with normal samples if no error messages occur.
Sample tube and sample volume
13 mm diameter x 75 mm height with Hemoguard lock
• Minimum sample volume = 1 ml
• Maximum sample volume = 3 ml
NOTE: The EDTA anticoagulant should be in a dissolved form in the tubes. The tube must be filled so far that the prescribed blood / EDTA ratio is neither exceeded nor undercut. There must be an air volume of approx. 1 cm in the sample tube, otherwise the mixture is incomplete.
Specification for the barcode - CELL-DYN 3500 SL
Barcode format
The following formats can be used with or without a check digit:
Code 39
Interleaf 2 of 5
Codabar
Barcode Label Specifications
Requirements for the barcode labels:
• Print on good quality paper
• Free margin on each side: 6.4 mm
• Min. Strip width: 0.25 mm
• Ratio between wide and narrow strips: 2: 1
• Min. Strip length: 12.7 mm
• Max. Label length: 50.8 mm
• Max. Label width: 31.8 mm
Description of services and systems
The CELL-DYN 3500 SL is an analyzer for creating the large blood count for routine and emergency laboratories.
Special features
* Optical differentiation principle: Patented laser differentiation
* Double white blood cell count using two different methods (WIC / WOC, double Security)
* Analysis program with extended lysis (infant blood) and measurement time extension
* Color printer
* Extensive data memory for 10,000 complete reports as standard
* Corrected data for the "red blood count"
* Leukocyte count and differentiation also in the cytopenic area
Method program
Large blood count with 22 parameters, scattergrams, histograms and additional information.
The parameters in detail:
numeric:
Erythrocyte concentration Neutrophils *)
Leukocyte concentration Lymphocytes *)
Hemoglobin concentration Monocytes *)
Erythrocyte quotients MCV, MCH, MCHC Eosinophils *)
Erythrocyte Distribution Width (RDW) Basophils *)
Thrombocyte concentration
Mean Platelet Volume (MPV)
Platelet Distribution Width (PDW) *) each in absolute concentrations
Thrombocrit (PCT) and relative percentages
graphically:
6 Scatter cells of leukocytes
5 histograms, including the volume distribution curves of erythrocytes and platelets.